Transit Time Difference Flowmeter for Intravenous Flow Rate Measurement Using 1-3 Piezoelectric Composite Transducers

被引:21
作者
Kim, Taeyang [1 ]
Kim, Jinwook [1 ]
Jiang, Xiaoning [1 ,2 ]
机构
[1] North Carolina State Univ, Dept Mech & Aerosp Engn, Raleigh, NC 27695 USA
[2] City Univ Hong Kong, Dept Mech & Biomed Engn, Hong Kong, Hong Kong, Peoples R China
关键词
Intravenous (IV) flow rate; transit time difference; ultrasonic flowmeter; 1-3 composite transducer; MEDICATION ERRORS; PEDIATRIC-PATIENTS; THICKNESS-MODE; LIQUIDS;
D O I
10.1109/JSEN.2017.2727340
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The flow rate of injected medication implemented by intravenous (IV) systems must be accurately monitored and meticulously controlled to prevent medical accidents. In this paper, an ultrasonic flowmeter (UF) with 1-3 piezoelectric composite transducers was designed, fabricated, and tested on a variety of flow rates of mimic medical injections. The transducer wedge for the angled beam propagation and an acoustic impedance matching layer were included in the design for transmission enhancement. To ensure an accurate measurement of flow rate, the effect of the flow distributions inside the IV tube was taken into account. The developed UF exhibited the capability of detecting low flow rates (<0.005 m/s), with 1%-2% discrepancy compared with the reference rate of infusion.
引用
收藏
页码:5741 / 5748
页数:8
相关论文
共 34 条
[1]   Thermal flow sensor for liquids and gases based on combinations of two principles [J].
Ashauer, M ;
Glosch, H ;
Hedrich, F ;
Hey, N ;
Sandmaier, H ;
Lang, W .
SENSORS AND ACTUATORS A-PHYSICAL, 1999, 73 (1-2) :7-13
[2]   Medication errors observed in 36 health care facilities [J].
Barker, KN ;
Flynn, EA ;
Pepper, GA ;
Bates, DW ;
Mikeal, RL .
ARCHIVES OF INTERNAL MEDICINE, 2002, 162 (16) :1897-1903
[3]   EFFECT OF VOLUME AND RATE OF CONTRAST-MEDIUM INJECTION ON INTRAVENOUS DIGITAL SUBTRACTION ANGIOGRAPHIC CONTRAST-MEDIUM CURVES [J].
BURBANK, FH ;
BRODY, WR ;
BRADLEY, BR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1984, 4 (02) :308-315
[4]  
De Oliveira Buscarini I., 2015, J PHYS C SER, V648, P12005, DOI [10.1088/1742-6596/648/1/012005, DOI 10.1088/1742-6596/648/1/012005]
[5]  
Ding S., 2011, P EL SIGN PROC, P477
[6]   SIGNAL-TO-NOISE RATIO ENHANCEMENT BY LEAST-SQUARES POLYNOMIAL SMOOTHING [J].
ENKE, CG ;
NIEMAN, TA .
ANALYTICAL CHEMISTRY, 1976, 48 (08) :A705-&
[7]  
Fyrippi I, 2004, FLOW MEAS INSTRUM, V15, P131, DOI 10.1016/j.flowmeasinst.2003.12.2002
[8]  
Guyton AC, 2006, TXB MED PHYSL
[9]  
Hooker M.W., 1998, NASACR208708 NASA
[10]  
*IEEE, 1987, 1761987 ANSIIEEE