Bioresorbable polylactide interbody implants in an ovine anterior cervical discectomy and fusion model

被引:39
作者
Thomas, Kevin A. [1 ]
Toth, Jeffrey M. [2 ]
Crawford, Neil R. [3 ]
Seim, Howard B., III [4 ]
Shi, Lewis L. [5 ]
Harris, Mitchel B. [5 ]
Turner, A. Simon [3 ]
机构
[1] Cytori Therapeut Co, MacroPore Biosurg, San Diego, CA 92121 USA
[2] Med Coll Wisconsin, Dept Orthopaed Surg, Milwaukee, WI 53226 USA
[3] Barrow Neurol Inst, Spinal Biomech Res Lab, Phoenix, AZ 85013 USA
[4] Colorado State Univ, Coll Vet Med, Ft Collins, CO 80523 USA
[5] Brigham & Womens Hosp, Dept Orthopaed Surg, Boston, MA 02115 USA
关键词
bioresorbable interbody implant; polylactide; 70; 30; poly(L-lactide-co-D; L-lactide); PLDLLa; ACDF; animal model;
D O I
10.1097/BRS.0b013e3181695716
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. In vivo study of anterior discectomy and fusion using a bioresorbable 70: 30 poly(L-lactide-co-D, L-lactide) interbody implant in an ovine model. Objective. To evaluate the efficacy of the polylactide implant to function as an interbody fusion device, and to assess the tissue reaction to the material during the resorption process. Summary of Background Data. The use of polylactide as a cervical interbody implant has several potential advantages when compared with traditional materials. Having an elastic modulus very similar to bone minimizes the potential for stress shielding, and as the material resorbs additional loading is transferred to the developing fusion mass. Although preclinical and clinical studies have demonstrated the suitability of polylactide implants for lumbar interbody fusion, detailed information on cervical anterior cervical discectomy and fusion (ACDF) with polylactide devices is desirable. Methods. Single level ACDF was performed in 8 skeletally mature ewes. Bioresorbable 70: 30 poly (L-lactide-co-D, L-lactide) interbody implants packed with autograft were used with single-level metallic plates. Radiographs were made every 3 months up to 1 year, and yearly thereafter. The animals were killed at 6 months (3 animals), 12 months (3 animals), and 36 months (2 animals). In addition to the serial plain radiographs, the specimens were evaluated by nondestructive biomechanical testing and undecalcified histologic analysis. Results. The bioresorbable polylactide implants were effective in achieving interbody fusion. The 6-month animals appeared fused radiographically and biomechanically, whereas histologic sections demonstrated partial fusion (in 3 of 3 animals). Radiographic fusion was confirmed histologically and biomechanically at 12 months (3 of 3 animals) and 36 months (2 of 2 animals). A mild chronic inflammatory response to the resorbing polylactide implant was observed at both 6 months and 12 months. At 36 months, the operative levels were solidly fused and the implants were completely resorbed. No adverse tissue response was observed in any animal at any time period. Conclusion. Interbody fusion was achieved using bioresorbable polylactide implants, with no evidence of implant collapse, extrusion, or adverse tissue response to the material. The use of polylactide as a cervical interbody device appears both safe and effective based on these ACDF animal model results.
引用
收藏
页码:734 / 742
页数:9
相关论文
共 41 条
[1]   Biomechanical analysis of a resorbable anterior cervical graft containment plate [J].
Ames, CP ;
Crawford, NR ;
Chamberlain, RH ;
Deshmukh, V ;
Sadikovic, B ;
Sonntag, VKH .
SPINE, 2005, 30 (09) :1031-1038
[2]   Pre-clinical in vivo evaluation of orthopaedic bioabsorbable devices [J].
An, YH ;
Woolf, SK ;
Friedman, RJ .
BIOMATERIALS, 2000, 21 (24) :2635-2652
[3]  
*ASTM, 2006, F98104 ASTM
[4]   Sterilization, toxicity, biocompatibility and clinical applications of polylactic acid polyglycolic acid copolymers [J].
Athanasiou, KA ;
Niederauer, GG ;
Agrawal, CM .
BIOMATERIALS, 1996, 17 (02) :93-102
[5]   Clinical biocompatibility of biodegradable orthopaedic implants for internal fixation:: a review [J].
Böstman, O ;
Pihlajamäki, H .
BIOMATERIALS, 2000, 21 (24) :2615-2621
[6]   ABSORBABLE IMPLANTS FOR THE FIXATION OF FRACTURES [J].
BOSTMAN, OM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (01) :148-153
[7]  
Cherry C., 2002, AORN J, V76, P996
[8]  
Ciccone W J 2nd, 2001, J Am Acad Orthop Surg, V9, P280
[9]  
CORNWALL GB, 2003, T 29 ANN M SOC BIOM, V26, P393
[10]   A new technique for determining 3-D joint angles: the tilt/twist method [J].
Crawford, NR ;
Yamaguchi, GT ;
Dickman, CA .
CLINICAL BIOMECHANICS, 1999, 14 (03) :153-165