Mechanical failure of the Intramedullary Skeletal Kinetic Distractor in limb lengthening

被引:54
作者
Burghardt, R. D.
Herzenberg, J. E. [1 ]
Specht, S. C. [1 ]
Paley, D.
机构
[1] Sinai Hosp Baltimore, Int Ctr Limb Lengthening, Rubin Inst Adv Orthoped, Baltimore, MD 21215 USA
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2011年 / 93B卷 / 05期
关键词
NAIL;
D O I
10.1302/0301-620X.93B5.25986
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Between October 2001 and September 2009 we lengthened 242 lower-limb segments in 180 patients using the Intramedullary Skeletal Kinetic Distractor (ISKD). Mechanical failure was defined either as breakage of the ISKD or failure of the internal mechanism to activate. Retrieved nails which failed mechanically were examined by the manufacturer for defects. In all, 15 ISKDs in 12 patients (13 limbs) failed mechanically representing an overall failure rate of 6.2%, with fracture of the device occurring in ten of the 15 failures. Two nails in one patient failed to lengthen and had to be replaced. The manufacturer detected an error in the assembly of the nail, which prompted a wide recall. One nail jammed after being forcefully inserted, and two nails failed to lengthen fully. Lengthening was achieved in all 12 patients, although three required a second operation to exchange a defective nail for a new, functioning device. The ISKD is a complex mechanical device which lengthens by the oscillation of two telescopic sections connected by a threaded rod. The junction between these sections is surrounded by a keyring collar. This keyring collar is the weakest part of the device.
引用
收藏
页码:639 / 643
页数:5
相关论文
共 23 条
  • [11] Guichet JM, 1997, CLIN ORTHOP RELAT R, P281
  • [12] HANKEMEIER S, 2005, OPER ORTHOP TRAUMATO, V17, P59
  • [13] Femoral lengthening over an intramedullary nail - A matched-case comparison with Ilizarov femoral lengthening
    Paley, D
    Herzenberg, JE
    Paremain, G
    Bhave, A
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (10) : 1464 - 1480
  • [14] CURRENT TECHNIQUES OF LIMB LENGTHENING
    PALEY, D
    [J]. JOURNAL OF PEDIATRIC ORTHOPAEDICS, 1988, 8 (01) : 73 - 92
  • [15] PALEY D, 1990, CLIN ORTHOP RELAT R, P81
  • [16] PFEIL J, 1996, EXTREMITATENVERLANGE
  • [17] Rangger C, 1996, UNFALLCHIRURG, V99, P68
  • [18] Implant failure in tibial nailing
    Ruiz, AL
    Kealey, WDC
    MCCoy, GF
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 (05): : 359 - 362
  • [19] Schmidgen A, 1999, UNFALLCHIRURG, V102, P975, DOI 10.1007/s001130050513
  • [20] SCHOLLNER D, 1972, Z ORTHOP GRENZGEB, V110, P971