Stress and Strain Patterns of 1-Piece and 2-Piece Implant Systems in Bone: A 3-Dimensional Finite Element Analysis

被引:15
|
作者
Ormianer, Zeev [1 ]
Ben Amar, Arial [1 ]
Duda, Mariusz [2 ]
Marku-Cohen, Sharon [1 ]
Lewinstein, Israel [1 ]
机构
[1] Tel Aviv Univ, Dept Oral Rehabil, Maurice & Gabriela Goldschleger Sch Dent Med, IL-69978 Tel Aviv, Israel
[2] Silesia Med Oral Med Clin, Katowice, Poland
关键词
1-piece; 2-piece; dental implant; finite element analysis; bone; DENTAL IMPLANTS; BITE FORCE; IMMEDIATE; SURVIVAL; TISSUE;
D O I
10.1097/ID.0b013e31823fce22
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of Problem: The transition from implant to abutment is solid in 1-piece (1P) and broken in 2-piece (2P) implant designs. This difference may affect occlusal load distribution and marginal bone response. Purpose: To determine whether 1P and 2P implants with equivalent geometries exhibited stresses and strains differently under applied loading conditions. Materials and Methods: Design software simulated 1P and 2P implants restored with metal copings and embedded in 3 cylindrical bone block models that varied in dimensions, density, and percentage of bone-to-implant contact. Three-dimensional, finite element analysis simulated occlusal loading. Experiments evaluated stresses and strains relative to implant design and (1) periimplant bone thickness, (2) cortical bone thickness, (3) magnitude and direction of occlusal loading, and (4) % bone-to-implant contact. Results: Implants with equivalent dimensions exhibited comparable stresses and strains in all experimental conditions. Implant diameter and periimplant bone thickness influenced stress levels. Only small-diameter (3.0 mm) 1P implants in low-density bone exhibited stress levels that might adversely affect marginal bone stability. Conclusions: Implant diameter and periimplant bone thickness influenced load distribution in bone, but the type of implant-abutment transition had no significant effect. Small-diameter 1P implants should be limited to dense bone to minimize stress concentrations. (Implant Dent 2012;21:39-45)
引用
收藏
页码:39 / 45
页数:7
相关论文
共 50 条
  • [41] Effect of Different Implant-Abutment Mismatches on Stress Distribution: A 3-Dimensional Finite Element Study
    Esfahanian, Vahid
    Khosravi, Armin
    Ketabi, Mohammad
    Deporter, Douglas
    JOURNAL OF ORAL IMPLANTOLOGY, 2022, 48 (05) : 370 - 374
  • [42] Influence of the Implant Diameter With Different Sizes of Hexagon: Analysis by 3-Dimensional Finite Element Method
    Pellizzer, Eduardo Piza
    Verri, Fellippo Ramos
    Dantas de Moraes, Sandra Lucia
    Falcon-Antenucci, Rosse Mary
    Perri de Carvalho, Paulo Sergio
    Noritomi, Pedro Yoshito
    JOURNAL OF ORAL IMPLANTOLOGY, 2013, 39 (04) : 425 - 431
  • [43] Stress and strain distribution patterns in bone around tissue-and bone-level implant–supported mandibular overdentures using three-dimensional finite-element analysis
    Hajizadeh F.
    Panahi S.
    J. Long-Term Eff. Med. Implants, 2019, 3 (175-181): : 175 - 181
  • [44] Effect of Integration Patterns Around Implant Neck on Stress Distribution in Peri-Implant Bone: A Finite Element Analysis
    Han, Jingyun
    Sun, Yuchun
    Wang, Chao
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2017, 26 (06): : 549 - 558
  • [45] Finite element analysis of bone stress and strain around a distal osseointegrated implant for prosthetic limb attachment
    Xu, W
    Crocombe, AD
    Hughes, SC
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H6) : 595 - 602
  • [46] Stress-strain distribution at bone-implant interface of two splinted overdenture systems using 3D finite element analysis
    Hussein, Mustafa Omran
    JOURNAL OF ADVANCED PROSTHODONTICS, 2013, 5 (03): : 333 - 340
  • [47] Effects of different fixture geometries on the stress distribution in mandibular peri-implant structures: a 3-dimensional finite element analysis
    Degerliyurt, Kagan
    Simsek, Baris
    Erkmen, Erkan
    Eser, Atilim
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2010, 110 (02): : E1 - E11
  • [48] Displacement and stress distribution by different bone-borne palatal expanders with facemask: A 3-dimensional finite element analysis
    Park, Jae Hyun
    Bayome, Mohamed
    Zahrowski, James J.
    Kook, Yoon-Ah
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2017, 151 (01) : 105 - 117
  • [49] 3-DIMENSIONAL FINITE-ELEMENT ANALYSIS FOR STRESS IN THE PERIODONTAL TISSUE BY ORTHODONTIC FORCES
    TANNE, K
    SAKUDA, M
    BURSTONE, CJ
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1987, 92 (06) : 499 - 505
  • [50] Stress Analysis in Platform-Switching Implants: A 3-Dimensional Finite Element Study
    Pellizzer, Eduardo Piza
    Verri, Fellippo Ramos
    Falcon-Antenucci, Rosse Mary
    Santiago Junior, Joel Ferreira
    Perri de Carvalho, Paulo Sergio
    Dantas de Moraes, Sandra Lucia
    Noritomi, Pedro Yoshito
    JOURNAL OF ORAL IMPLANTOLOGY, 2012, 38 (05) : 587 - 594