Stress and Strain Patterns of 1-Piece and 2-Piece Implant Systems in Bone: A 3-Dimensional Finite Element Analysis

被引:15
|
作者
Ormianer, Zeev [1 ]
Ben Amar, Arial [1 ]
Duda, Mariusz [2 ]
Marku-Cohen, Sharon [1 ]
Lewinstein, Israel [1 ]
机构
[1] Tel Aviv Univ, Dept Oral Rehabil, Maurice & Gabriela Goldschleger Sch Dent Med, IL-69978 Tel Aviv, Israel
[2] Silesia Med Oral Med Clin, Katowice, Poland
关键词
1-piece; 2-piece; dental implant; finite element analysis; bone; DENTAL IMPLANTS; BITE FORCE; IMMEDIATE; SURVIVAL; TISSUE;
D O I
10.1097/ID.0b013e31823fce22
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of Problem: The transition from implant to abutment is solid in 1-piece (1P) and broken in 2-piece (2P) implant designs. This difference may affect occlusal load distribution and marginal bone response. Purpose: To determine whether 1P and 2P implants with equivalent geometries exhibited stresses and strains differently under applied loading conditions. Materials and Methods: Design software simulated 1P and 2P implants restored with metal copings and embedded in 3 cylindrical bone block models that varied in dimensions, density, and percentage of bone-to-implant contact. Three-dimensional, finite element analysis simulated occlusal loading. Experiments evaluated stresses and strains relative to implant design and (1) periimplant bone thickness, (2) cortical bone thickness, (3) magnitude and direction of occlusal loading, and (4) % bone-to-implant contact. Results: Implants with equivalent dimensions exhibited comparable stresses and strains in all experimental conditions. Implant diameter and periimplant bone thickness influenced stress levels. Only small-diameter (3.0 mm) 1P implants in low-density bone exhibited stress levels that might adversely affect marginal bone stability. Conclusions: Implant diameter and periimplant bone thickness influenced load distribution in bone, but the type of implant-abutment transition had no significant effect. Small-diameter 1P implants should be limited to dense bone to minimize stress concentrations. (Implant Dent 2012;21:39-45)
引用
收藏
页码:39 / 45
页数:7
相关论文
共 50 条
  • [1] Bone Microstrain Values of 1-Piece and 2-Piece Implants Subjected to Mechanical Loading
    Harel, Noga
    Eshkol-Yogev, Inbar
    Piek, Dana
    Livne, Shiri
    Lavi, David
    Ormianer, Zeev
    IMPLANT DENTISTRY, 2013, 22 (03) : 277 - 281
  • [2] An in vitro evaluation of the strength of a 1-piece and 2-piece conical abutment joint in implant design
    Norton, MR
    CLINICAL ORAL IMPLANTS RESEARCH, 2000, 11 (05) : 458 - 464
  • [3] A clinical comparison of 1-piece versus 2-piece implants: A systematic review and meta-analysis
    Justino de Oliveira Limirio, Joao Pedro
    Araujo Lemos, Cleidiel Aparecido
    de Luna Gomes, Jessica Marcela
    Minatel, Lurian
    Rosifini Alves Rezende, Maria Cristina
    Pellizzer, Eduardo Piza
    JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (04): : 439 - 445
  • [4] Effect of gingival height of a titanium base on the biomechanical behavior of 2-piece custom implant abutments: A 3-dimensional nonlinear finite element study
    Poovarodom, Pongsakorn
    Rungsiyakull, Chaiy
    Suriyawanakul, Jarupol
    Li, Qing
    Sasaki, Keiichi
    Yoda, Nobuhiro
    Rungsiyakull, Pimduen
    JOURNAL OF PROSTHETIC DENTISTRY, 2023, 130 (03): : 380.e1 - 380.e9
  • [5] Three-Dimensional Finite Element Analysis of the Effect of 1-Piece Superstructure on Mandibular Flexure
    Naini, Roshanak Baghai
    Nokar, Saied
    IMPLANT DENTISTRY, 2009, 18 (05) : 428 - 437
  • [6] Stress analysis in edentulous mandibular bone supporting implant-retained 1-piece or multiple superstructures
    Yokoyama, S
    Wakabayashi, N
    Shiota, M
    Ohyama, T
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2005, 20 (04) : 578 - 583
  • [7] Influence of implant design and bone quality on stress/strain distribution in bone around implants: A 3-dimensional finite element analysis
    Tada, S
    Stegaroiu, R
    Kitamura, E
    Miyakawa, O
    Kusakari, H
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2003, 18 (03) : 357 - 368
  • [8] In Situ Tooth Replica Custom Implant: A 3-Dimensional Finite Element Stress and Strain Analysis
    Ghuneim, Wael Aly
    JOURNAL OF ORAL IMPLANTOLOGY, 2013, 39 (05) : 559 - 573
  • [9] Influence of prosthesis material on stress distribution in bone and implant: A 3-dimensional finite element analysis
    Stegaroiu, R
    Kusakari, H
    Nishiyama, S
    Miyakawa, O
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1998, 13 (06) : 781 - 790
  • [10] Finite element analysis applied to ergonomic design of 2-piece aluminum beverage cans and bottles
    Han, Jing
    Yamazaki, Koetsu
    Nishiyama, Sadao
    Itoh, Ryoichi
    PROCEEDINGS OF THE ASME INTERNATIONAL DESIGN ENGINEERING TECHNICAL CONFERENCES AND COMPUTERS AND INFORMATION IN ENGINEERING CONFERENCE 2007, VOL 6, PTS A AND B, 2008, : 387 - 396