Characterization of Al2O3 optically stimulated luminescence films for 2D dosimetry using a 6 MV photon beam

被引:23
作者
Ahmed, M. F. [1 ]
Shrestha, N. [1 ]
Schnell, E. [2 ]
Ahmad, S. [2 ,3 ]
Akselrod, M. S. [3 ]
Yukihara, E. G. [1 ]
机构
[1] Oklahoma State Univ, Dept Phys, Stillwater, OK 74078 USA
[2] Univ Oklahoma, Hlth Sci Ctr, Dept Radiat Oncol, Oklahoma City, OK 73104 USA
[3] Landauer Inc, Stillwater Crystal Growth Div, Stillwater, OK 74074 USA
关键词
optically stimulated luminescence; Al2O3:C; Mg; 2D dosimetry; radiotherapy; small field dosimetry; HIGH-ENERGY PHOTON; DOSE-RESPONSE; STORAGE PHOSPHORS; RADIATION-THERAPY; PROTON; DETECTOR; DEPENDENCE; ELECTRON;
D O I
10.1088/0031-9155/61/21/7551
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work evaluates the dosimetric properties of newly developed optically stimulated luminescence (OSL) films, fabricated with either Al2O3:C or Al2O3:C,Mg, using a prototype laser scanning reader, a developed image reconstruction algorithm, and a 6 MV therapeutic photon beam. Packages containing OSL films (Al2O3:C and Al2O3:C, Mg) and a radiochromic film (Gafchromic EBT3) were irradiated using a 6 MV photon beam using different doses, field sizes, with and without wedge filter. Dependence on film orientation of the OSL system was also tested. Diode-array (MapCHECK) and ionization chamber measurements were performed for comparison. The OSLD film doses agreed with the MapCHECK and ionization chamber data within the experimental uncertainties (< 2% at 1.5 Gy). The system background and minimum detectable dose (MDD) were < 0.5 mGy, and the dose response was approximately linear from the MDD up to a few grays (the linearity correction was < 10% up to similar to 2-4 Gy), with no saturation up to 30 Gy. The dose profiles agreed with those obtained using EBT3 films (analyzed using the triple channel method) in the high dose regions of the images. In the low dose regions, the dose profiles from the OSLD films were more reproducible than those from the EBT3 films. We also demonstrated that the OSL film data are independent on scan orientation and field size over the investigated range. The results demonstrate the potential of OSLD films for 2D dosimetry, particularly for the characterization of small fields, due to their wide dynamic range, linear response, resolution and dosimetric properties. The negligible background and potential simple calibration make these OSLD films suitable for remote audits. The characterization presented here may motivate further commercial development of a 2D dosimetry system based on the OSL from Al2O3:C or Al2O3:C,Mg.
引用
收藏
页码:7551 / 7570
页数:20
相关论文
共 44 条
[1]   Development of a 2D dosimetry system based on the optically stimulated luminescence of Al2O3 [J].
Ahmed, M. F. ;
Eller, S. A. ;
Schnell, E. ;
Ahmad, S. ;
Akselrod, M. S. ;
Hanson, O. D. ;
Yukihara, E. G. .
RADIATION MEASUREMENTS, 2014, 71 :187-192
[2]  
Ahmed M F, 2016, PHYS MED BI IN PRESS
[3]   New Al2O3:C,Mg crystals for radiophotoluminescent dosimetry and optical imaging [J].
Akselrod, Mark S. ;
Akselrod, Anna E. .
RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) :218-221
[4]   A two dimensional silicon detectors array for quality assurance in stereotactic radiotherapy: MagicPlate-512 [J].
Aldosari, A. H. ;
Petasecca, M. ;
Espinoza, A. ;
Newall, M. ;
Fuduli, I. ;
Porumb, C. ;
Alshaikh, S. ;
Alrowaili, Z. A. ;
Weaver, M. ;
Metcalfe, P. ;
Carolan, M. ;
Lerch, M. L. F. ;
Perevertaylo, V. ;
Rosenfeld, A. B. .
MEDICAL PHYSICS, 2014, 41 (09)
[5]  
[Anonymous], 2016, THESIS
[6]   Silicon strip detector for a novel 2D dosimetric method for radiotherapy treatment verification [J].
Bocci, A. ;
Cortes-Giraldo, M. A. ;
Gallardo, M. I. ;
Espino, J. M. ;
Arrans, R. ;
Alvarez, M. A. G. ;
Abou-Haidar, Z. ;
Quesada, J. M. ;
Perez Vega-Leal, A. ;
Perez Nieto, F. J. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 673 :98-106
[7]   Detector dose response in megavoltage small photon beams. I. Theoretical concepts [J].
Bouchard, Hugo ;
Seuntjens, Jan ;
Duane, Simon ;
Kamio, Yuji ;
Palmans, Hugo .
MEDICAL PHYSICS, 2015, 42 (10) :6033-6047
[8]   Novel Silicon Devices for Radiation Therapy Monitoring [J].
Bruzzi, Mara .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 809 :105-112
[9]   2-D dosimetry of a proton radiotherapy beam using large-area LiF: Mg,Cu,P TL detectors [J].
Czopyk, L. ;
Cirrone, G. A. P. ;
Cuttone, G. ;
Klosowski, M. ;
Olko, P. ;
Sroka, U. ;
Waligorski, M. P. R. .
RADIATION MEASUREMENTS, 2008, 43 (2-6) :977-980
[10]   Two-dimensional dosimetry of radiotherapeutical proton beams using thermoluminescence foils [J].
Czopyk, L. ;
Klosowski, M. ;
Olko, P. ;
Swakon, J. ;
Waligorski, M. P. R. ;
Kajdrowicz, T. ;
Cuttone, G. ;
Cirrone, G. A. P. ;
Di Rosa, F. .
RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) :185-189