Magnetic resonance imaging compatible remote catheter navigation system with 3 degrees of freedom

被引:19
|
作者
Tavallaei, M. A. [1 ]
Lavdas, M. K. [1 ]
Gelman, D. [1 ]
Drangova, M. [1 ]
机构
[1] Univ Western Ontario, Robarts Res Inst, Imaging Res Labs, London, ON N6A 5B9, Canada
基金
加拿大自然科学与工程研究理事会; 加拿大健康研究院;
关键词
Magnetic resonance imaging; Master-slave; MRI-compatible; Real-time systems; Surgical robotics; Telerobotics; Catheterization; ATRIAL-FIBRILLATION; ABLATION; TIME; PATIENT; SAFETY; MRI; ELECTROPHYSIOLOGY; VISUALIZATION; EXPERIENCE; EFFICACY;
D O I
10.1007/s11548-015-1337-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To facilitate MRI-guided catheterization procedures, we present an MRI-compatible remote catheter navigation system that allows remote navigation of steerable catheters with 3 degrees of freedom. The system consists of a user interface (master), a robot (slave), and an ultrasonic motor control servomechanism. The interventionalist applies conventional motions (axial, radial and plunger manipulations) on an input catheter in the master unit; this user input is measured and used by the servomechanism to control a compact catheter manipulating robot, such that it replicates the interventionalist's input motion on the patient catheter. The performance of the system was evaluated in terms of MRI compatibility (SNR and artifact), feasibility of remote navigation under real-time MRI guidance, and motion replication accuracy. Real-time MRI experiments demonstrated that catheter was successfully navigated remotely to desired target references in all 3 degrees of freedom. The system had an absolute value error of 1 mm in axial catheter motion replication over 30 mm of travel and for radial catheter motion replication over . The worst case SNR drop was observed to be 3 %; the robot did not introduce any artifacts in the MR images. An MRI-compatible compact remote catheter navigation system has been developed that allows remote navigation of steerable catheters with 3 degrees of freedom. The proposed system allows for safe and accurate remote catheter navigation, within conventional closed-bore scanners, without degrading MR image quality.
引用
收藏
页码:1537 / 1545
页数:9
相关论文
共 50 条
  • [21] Effect of Magnetic Resonance Imaging on the Motion Accuracy of Magnetic Resonance Imaging-compatible Focused Ultrasound Robotic System
    Antoniou, Anastasia
    Evripidou, Nikolas
    Chrysanthou, Antreas
    Georgiou, Leonidas
    Ioannides, Cleanthis
    Spanoudes, Kyriakos
    Damianou, Christakis
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (02) : 203 - 212
  • [22] Safety evaluation of magnetic catheter steering with upgraded magnetic resonance imaging system
    Bringout, G.
    Lalande, V.
    Gosselin, F. P.
    Martel, S.
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 6702 - 6705
  • [23] Performance of Ventilators Compatible With Magnetic Resonance Imaging: A Bench Study
    Chikata, Yusuke
    Okuda, Nao
    Izawa, Masayo
    Onodera, Mutsuo
    Nishimura, Masaji
    RESPIRATORY CARE, 2015, 60 (03) : 341 - 346
  • [24] Symptomatic and asymptomatic intracerebral hemorrhages detected by magnetic resonance imaging after catheter ablation of atrial fibrillation
    Nakamura, Kohki
    Sasaki, Takehito
    Minami, Kentaro
    Take, Yutaka
    Kimura, Koki
    Haraguchi, Yumiko
    Sasaki, Wataru
    Kishi, Shohei
    Yoshimura, Shingo
    Nakatani, Yosuke
    Miki, Yuko
    Goto, Koji
    Kaseno, Kenichi
    Yamashita, Eiji
    Koyama, Keiko
    Funabashi, Nobusada
    Naito, Shigeto
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2022, 64 (02) : 443 - 454
  • [25] The Pulsiocath® catheter and magnetic resonance imaging
    Greco, F.
    Vendrell, J. -F.
    Deras, P.
    Boularan, A.
    Perrigault, P. -F.
    ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2011, 30 (09): : 697 - 697
  • [26] Remote magnetic navigation to guide endocardial and epicardial catheter mapping of scar-related ventricular tachycardia
    Aryana, Arash
    d'Avila, Andre
    Heist, E. Kevin
    Mela, Theofanie
    Singh, Jagmeet P.
    Ruskin, Jeremy N.
    Reddy, Vivek Y.
    CIRCULATION, 2007, 115 (10) : 1191 - 1200
  • [27] Remote magnetic versus manual catheter navigation for circumferential pulmonary vein ablation in patients with atrial fibrillation
    Lars Lüthje
    Dirk Vollmann
    Joachim Seegers
    Marc Dorenkamp
    Christian Sohns
    Gerd Hasenfuss
    Markus Zabel
    Clinical Research in Cardiology, 2011, 100 : 1003 - 1011
  • [28] Developing a Magnetic Resonance-Compatible Catheter for Cardiac Catheterization
    Yao, Wei
    Schaeffter, Tobias
    Seneviratne, Lakmal
    Althoefer, Kaspar
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2012, 6 (04):
  • [29] Cavo-tricuspid isthmus radiofrequency ablation using a novel remote navigation catheter system in patients with typical atrial flutter
    Lopez-Gil, Maria
    Salgado, Ricardo
    Luis Merino, Jose
    Datino, Tomas
    Figueroa, Jorge
    Arenal, Angel
    Mejia, Elena
    Salguero, Rafael
    Fontenla, Adolfo
    Arribas, Fernando
    EUROPACE, 2014, 16 (04): : 558 - 562
  • [30] Characterization of the deflections of a catheter steered using a magnetic resonance imaging system
    Gosselin, Frederick P.
    Lalande, Viviane
    Martel, Sylvain
    MEDICAL PHYSICS, 2011, 38 (09) : 4994 - 5002