Does Pelvic Orientation Influence Wear Measurement of the Acetabular Cup in Total Hip Arthroplasty-An Experimental Study

被引:4
作者
Wu, Junzhe [1 ,2 ]
Taylor, Dominic [3 ]
Forst, Raimund [1 ]
Seehaus, Frank [1 ,4 ]
机构
[1] Friedrich Alexander Univ Erlangen Nurnberg FAU, Fac Med, Dept Orthopaed Surg, D-91054 Erlangen, Germany
[2] Fujian Med Univ, Affiliated Hosp 2, Dept Orthopaed Surg, Quanzhou 362000, Peoples R China
[3] Hosp Coburg, Dept Orthopaed & Trauma Surg, D-96450 Coburg, Germany
[4] Friedrich Alexander Univ Erlangen Nurnberg FAU, Univ Hosp Erlangen, Dept Trauma & Orthopaed Surg, D-91054 Erlangen, Germany
来源
APPLIED SCIENCES-BASEL | 2021年 / 11卷 / 21期
关键词
polyethylene wear; roentgen stereophotogrammetric analysis; total hip arthroplasty; precision; accuracy; pelvic orientation; POLYETHYLENE WEAR; RADIOGRAPHIC METHODS; FOLLOW-UP; ACCURACY; VALIDATION; LONG;
D O I
10.3390/app112110014
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Roentgen stereophotogrammetric analysis (RSA) is the gold standard to detect in vivo material wear of the bearing couples in hip arthroplasty. Some surgical planning tools offer the opportunity to detect wear by using standard a.p. radiographs (2D(wear)), whilst RSA (3D(wear)) needs a special radiological setup. The aims of this study are to prove the interchangeable applicability of a 2D(wear) approach next to RSA and to assess the influence of different pelvic positions on measurement outcomes. An implant-bone model was used to mimic three different wear scenarios in seven pelvic-femur alignment positions. RSA and a.p. radiographs of the reference and a follow-up (simulated wear) pose were acquired. Accuracy and precision were worse for the 2D(wear) approach (0.206 mm; 0.159 mm) in comparison to the 3D(wear) approach (0.043 mm; 0.017 mm). Changing the pelvic position significantly influenced the 2D(wear) results (4 of 7, p < 0.05), whilst 3D(wear) results showed almost no change. The 3D(wear) is superior to the 2D(wear) approach, as it is less susceptible to changes in pelvic position. However, the results suggest that a 2D(wear) approach may be an alternative method if the wear present is in the range of 100-500 mu m and a.p. radiographs are available with the pelvis projected in a neutral position.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] ACETABULAR COMPONENT ORIENTATION IN TOTAL HIP ARTHROPLASTY: THE ROLE OF ACETABULAR TRANSVERSE LIGAMENT
    El Idrissi, Mohammed
    Elibrahimi, Abdelhalim
    Shimi, Mohammed
    Elmrini, Abdelmajid
    Acta Ortopedica Brasileira, 2016, 24 (05): : 267 - 269
  • [22] Pelvic tilt measurement before and after total hip arthroplasty
    Blondel, B.
    Parratte, S.
    Tropiano, P.
    Pauly, V.
    Aubaniac, J. -M.
    Argenson, J. -N.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2009, 95 (08) : 568 - 572
  • [23] Novel Measure of Acetabular Cup Inclination and Anteversion During Total Hip Arthroplasty
    Walter, William L.
    Baker, Neri A.
    Marsden-Jones, Daniel
    Sadeghpour, Ameneh
    MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2022, 15 : 1 - 14
  • [24] Catastrophic Wear of an Acetabular Component Misdiagnosed as Total Hip Arthroplasty Dislocation
    Papaioannou, Ioannis
    Pantazidou, Georgia
    Chatziperi, Stamatia
    Baikousis, Andreas
    Korovessis, Panagiotis
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2020, 12 (11)
  • [25] Is the acetabular cup orientation after total hip arthroplasty on a two dimension or three dimension model accurate?
    Benjamin Craiovan
    Tobias Renkawitz
    Markus Weber
    Joachim Grifka
    Lutz Nolte
    Guoyan Zheng
    International Orthopaedics, 2014, 38 : 2009 - 2015
  • [26] Pelvic positioning in the supine position leads to more consistent orientation of the acetabular component after total hip arthroplasty
    Grammatopoulos, G.
    Gofton, W.
    Cochran, M.
    Dobransky, J.
    Carli, A.
    Abdelbary, H.
    Gill, H. S.
    Beaule, P. E.
    BONE & JOINT JOURNAL, 2018, 100B (10) : 1280 - 1288
  • [27] The effect of total hip arthroplasty cup design on polyethylene wear rate
    Hamilton, WG
    Hopper, RH
    Ginn, SD
    Hammell, NP
    Engh, CA
    Engh, CA
    JOURNAL OF ARTHROPLASTY, 2005, 20 (07) : 63 - 72
  • [28] Effect of Cementless Acetabular Component Orientation, Position, and Containment in Total Hip Arthroplasty for Congenital Hip Disease
    Georgiades, George
    Babis, George C.
    Kourlaba, Georgia
    Hartofilakidis, George
    JOURNAL OF ARTHROPLASTY, 2010, 25 (07) : 1143 - 1150
  • [29] Accuracy of acetabular cup placement during anterolateral supine total hip arthroplasty using intraoperative fluoroscopy: a retrospective study
    Cho, Eiji
    Hisatome, Takashi
    Oda, Shuhei
    Fujimaki, Hirohisa
    Nakanishi, Kazuyoshi
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
  • [30] Accuracy of acetabular cup placement during anterolateral supine total hip arthroplasty using intraoperative fluoroscopy: a retrospective study
    Eiji Cho
    Takashi Hisatome
    Shuhei Oda
    Hirohisa Fujimaki
    Kazuyoshi Nakanishi
    Journal of Orthopaedic Surgery and Research, 17