Does Pelvic Orientation Influence Wear Measurement of the Acetabular Cup in Total Hip Arthroplasty-An Experimental Study

被引:4
|
作者
Wu, Junzhe [1 ,2 ]
Taylor, Dominic [3 ]
Forst, Raimund [1 ]
Seehaus, Frank [1 ,4 ]
机构
[1] Friedrich Alexander Univ Erlangen Nurnberg FAU, Fac Med, Dept Orthopaed Surg, D-91054 Erlangen, Germany
[2] Fujian Med Univ, Affiliated Hosp 2, Dept Orthopaed Surg, Quanzhou 362000, Peoples R China
[3] Hosp Coburg, Dept Orthopaed & Trauma Surg, D-96450 Coburg, Germany
[4] Friedrich Alexander Univ Erlangen Nurnberg FAU, Univ Hosp Erlangen, Dept Trauma & Orthopaed Surg, D-91054 Erlangen, Germany
来源
APPLIED SCIENCES-BASEL | 2021年 / 11卷 / 21期
关键词
polyethylene wear; roentgen stereophotogrammetric analysis; total hip arthroplasty; precision; accuracy; pelvic orientation; POLYETHYLENE WEAR; RADIOGRAPHIC METHODS; FOLLOW-UP; ACCURACY; VALIDATION; LONG;
D O I
10.3390/app112110014
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Roentgen stereophotogrammetric analysis (RSA) is the gold standard to detect in vivo material wear of the bearing couples in hip arthroplasty. Some surgical planning tools offer the opportunity to detect wear by using standard a.p. radiographs (2D(wear)), whilst RSA (3D(wear)) needs a special radiological setup. The aims of this study are to prove the interchangeable applicability of a 2D(wear) approach next to RSA and to assess the influence of different pelvic positions on measurement outcomes. An implant-bone model was used to mimic three different wear scenarios in seven pelvic-femur alignment positions. RSA and a.p. radiographs of the reference and a follow-up (simulated wear) pose were acquired. Accuracy and precision were worse for the 2D(wear) approach (0.206 mm; 0.159 mm) in comparison to the 3D(wear) approach (0.043 mm; 0.017 mm). Changing the pelvic position significantly influenced the 2D(wear) results (4 of 7, p < 0.05), whilst 3D(wear) results showed almost no change. The 3D(wear) is superior to the 2D(wear) approach, as it is less susceptible to changes in pelvic position. However, the results suggest that a 2D(wear) approach may be an alternative method if the wear present is in the range of 100-500 mu m and a.p. radiographs are available with the pelvis projected in a neutral position.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Improving Acetabular Cup Orientation in Total Hip Arthroplasty by Using Smartphone Technology
    Peters, Frank M.
    Greeff, Richard
    Goldstein, Neal
    Frey, Chris T.
    JOURNAL OF ARTHROPLASTY, 2012, 27 (07) : 1324 - 1330
  • [2] The effect of joint loading on acetabular wear measurement in total hip arthroplasty
    Martell, JM
    Leopold, SS
    Liu, XL
    JOURNAL OF ARTHROPLASTY, 2000, 15 (04) : 512 - 518
  • [3] The Effect of Functional Pelvic Tilt on the Three-Dimensional Acetabular Cup Orientation in Total Hip Arthroplasty Dislocations
    Snijders, Thom E.
    Schlosser, Tom P. C.
    Heckmann, Nathanael D.
    Tezuka, Taro
    Castelein, Rene M.
    Stevenson, Rob P.
    Weinans, Harry
    de Gast, Arthur
    Dorr, Lawrence D.
    JOURNAL OF ARTHROPLASTY, 2021, 36 (06) : 2184 - +
  • [4] Measurement of acetabular cup inclination in anteroposterior pelvic radiogram. An indicator of quality after primary total hip arthroplasty?
    Loitsch, T.
    Freitag, T.
    Leucht, F.
    Reichel, H.
    Bieger, R.
    ORTHOPADE, 2018, 47 (12): : 1003 - 1007
  • [5] Effect of Pelvic Obliquity on the Orientation of the Acetabular Component in Total Hip Arthroplasty
    Zhou, Xiaoxiao
    Wang, Qi
    Zhang, Xianlong
    Chen, Yunshu
    Shen, Hao
    Jiang, Yao
    JOURNAL OF ARTHROPLASTY, 2012, 27 (02) : 299 - 304
  • [6] MDCT Image Based Assessment of Acetabular Cup Orientation in Total Hip Arthroplasty
    Kobashi, Syoji
    Iwasa, Kenjiro
    Fujishiro, Takaaki
    Hayashi, Shiya
    Hashimoto, Shingo
    Kuroda, Ryosuke
    Kurosaka, Masahiro
    Kamiura, Naotake
    2014 IEEE INTERNATIONAL CONFERENCE ON SYSTEMS, MAN AND CYBERNETICS (SMC), 2014, : 2557 - 2562
  • [7] Acetabular component orientation in total hip arthroplasty: the impact of obesity
    McArthur, Benjamin A.
    Vulcano, Ettore
    Cross, Michael
    Nguyen, Joseph
    Della Valle, Alejandro G.
    Salvati, Eduardo
    HIP INTERNATIONAL, 2014, 24 (03) : 263 - 269
  • [8] Patient positioning and cup orientation during total hip arthroplasty: assessment of current UK practice
    Rutherford, Megan
    O'Connor, John D.
    Hill, Janet C.
    Beverland, David E.
    Lennon, Alex B.
    Dunne, Nicholas J.
    HIP INTERNATIONAL, 2019, 29 (01) : 89 - 95
  • [9] Radiographic assessment of the cup orientation after total hip arthroplasty: a literature review
    Zhao, Jing-Xin
    Su, Xiu-Yun
    Zhao, Zhe
    Xiao, Ruo-Xiu
    Zhang, Li-Cheng
    Tang, Pei-Fu
    ANNALS OF TRANSLATIONAL MEDICINE, 2020, 8 (04)
  • [10] Validation of a Computer-Assisted Method for Measurement of Radiographic Wear in Total Hip Arthroplasty Using All Polyethylene Cemented Acetabular Components
    Langlois, Jean
    Zaoui, Amine
    Scemama, Caroline
    Martell, John
    Bragdon, Charles
    Hamadouche, Moussa
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2015, 33 (03) : 417 - 420