Human-device interface pressure measurement in prosthetic, orthotic and exoskeleton applications: A systematic review

被引:11
作者
Armitage, Lucy [1 ]
Turner, Shruti [2 ]
Sreenivasa, Manish [1 ]
机构
[1] Univ Wollongong, Sch Mech Mat Mechatron & Biomed Engn, Northfields Ave, Wollongong, NSW 2522, Australia
[2] Imperial Coll London, Sackler MSk Lab, Dept Surg & Canc, Sir Michael Uren Hub, 86 Wood Ln, London W12 0BZ, England
关键词
BRAGG GRATING SENSORS; BIOMECHANICAL EVALUATION; RESIDUAL LIMB; SKIN PROBLEMS; SOCKET; STRESS; ORTHOSIS; AMPUTEES; FORCES; BODY;
D O I
10.1016/j.medengphy.2021.09.008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aimed to investigate normal and shear load sensor technology that has been characterised and used at the human-device interface in prosthetic, orthotic and exoskeleton applications. In addition to taking a cross-disciplinary view, this study expands on previous reviews by considering recently published papers, clinical translation of sensors, and development of the sensor technology itself. A search of MEDLINE, INSPEC, SCOPUS and Web of Science was performed up to 26 January 2021. A total of 33 studies were assessed for quality and their data extracted. The review found variable quality of published papers, with normal load being most commonly measured, and resistive sensor technology most commonly used. The translation to clinical environments was indicated in most studies, though the study population was not always made up of the target users. Studies could benefit from more direct comparison with clinically relevant load thresholds and by ensuring clinical testing is performed in the most realistic and representative way possible. Additionally, more focus on developing sensors that measure shear loads would enable further insights into conditions at the human-device interface. Finally, all researchers would benefit from better and more widespread anonymous data sharing practices to facilitate further experimentation.
引用
收藏
页码:56 / 69
页数:14
相关论文
共 69 条
[31]   Wireless sensors for continuous, multimodal measurements at the skin interface with lower limb prostheses [J].
Kwak, Jean Won ;
Han, Mengdi ;
Xie, Zhaoqian ;
Chung, Ha Uk ;
Lee, Jong Yoon ;
Avila, Raudel ;
Yohay, Jessica ;
Chen, Xuexian ;
Liang, Cunman ;
Patel, Manish ;
Jung, Inhwa ;
Kim, Jongwon ;
Namkoong, Myeong ;
Kwon, Kyeongha ;
Guo, Xu ;
Ogle, Christopher ;
Grande, Dominic ;
Ryu, Dennis ;
Kim, Dong Hyun ;
Madhvapathy, Surabhi ;
Liu, Claire ;
Yang, Da Som ;
Park, Yoonseok ;
Caldwell, Ryan ;
Banks, Anthony ;
Xu, Shuai ;
Huang, Yonggang ;
Fatone, Stefania ;
Rogers, John A. .
SCIENCE TRANSLATIONAL MEDICINE, 2020, 12 (574)
[32]   A pressure and shear sensor system for stress measurement at lower limb residuum/socket interface [J].
Laszczak, P. ;
McGrath, M. ;
Tang, J. ;
Gao, J. ;
Jiang, L. ;
Bader, D. L. ;
Moser, D. ;
Zahedi, S. .
MEDICAL ENGINEERING & PHYSICS, 2016, 38 (07) :695-700
[33]   Fiber Bragg Gratings in CYTOP Fibers Embedded in a 3D-Printed Flexible Support for Assessment of Human-Robot Interaction Forces [J].
Leal-Junior, Arnaldo ;
Theodosiou, Antreas ;
Diaz, Camilo ;
Marques, Carlos ;
Pontes, Maria Jose ;
Kalli, Kyriacos ;
Frizera-Neto, Anselmo .
MATERIALS, 2018, 11 (11)
[34]   Polymer optical fiber-embedded, 3D-printed instrumented support for microclimate and human-robot interaction forces assessment [J].
Leal-Junior, Arnaldo G. ;
Diaz, Camilo R. ;
Pontes, Maria Jose ;
Marques, Carlos ;
Frizera, Anselmo .
OPTICS AND LASER TECHNOLOGY, 2019, 112 :323-331
[35]   Measuring human-robot interaction on wearable robots: A distributed approach [J].
Lenzi, Tommaso ;
Vitiello, Nicola ;
De Rossi, Stefano Marco Maria ;
Persichetti, Alessandro ;
Giovacchini, Francesco ;
Roccella, Stefano ;
Vecchi, Fabrizio ;
Carrozza, Maria Chiara .
MECHATRONICS, 2011, 21 (06) :1123-1131
[36]  
LEVY SW, 1980, PROSTHET ORTHOT INT, V4, P37
[37]   Analysis of the corrective forces exerted by a dynamic derotation brace (DDB) [J].
Loukos, Ioannis ;
Zachariou, Constantinos ;
Nicolopoulos, Christos ;
Korres, Dimitrios ;
Efstathopoulos, Nicolaos .
PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2011, 35 (04) :365-372
[38]   Biomechanical evaluation of the Boston brace system for the treatment of adolescent idiopathic scoliosis [J].
Mac-Thiong, JM ;
Petit, Y ;
Aubin, CÉ ;
Delorme, S ;
Dansereau, J ;
Labelle, H .
SPINE, 2004, 29 (01) :26-32
[39]   Food incentives and completion of tuberculosis treatment [J].
Maher, Dermot .
BRITISH MEDICAL JOURNAL, 2009, 339
[40]   Biomechanics of Pressure Ulcer in Body Tissues Interacting with External Forces during Locomotion [J].
Mak, Arthur F. T. ;
Zhang, Ming ;
Tam, Eric W. C. .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, VOL 12, 2010, 12 :29-53