Screening of Keratoconus Using Autokeratometer and Keratometer Keratoconus Index

被引:1
|
作者
Kojima, Takashi [1 ,2 ]
Isogai, Naoki [2 ]
Nishida, Tomoya [2 ]
Nakamura, Tomoaki [2 ]
Ichikawa, Kazuo [3 ]
机构
[1] Keio Univ, Dept Ophthalmol, Sch Med, Shinjuku Ku, Shinanomachi 35, Tokyo 1608582, Japan
[2] Nagoya Eye Clin, Nagoya, Aichi 4560003, Japan
[3] Chukyo Eye Clin, Nagoya, Aichi 4560032, Japan
关键词
keratoconus; diagnosis; autokeratometer; astigmatism; keratometer keratoconus index (KKI); COLLAGEN CROSS-LINKING; ULTRAVIOLET-A LIGHT; RISK-FACTORS; CORNEAL ASTIGMATISM; ALLERGIC DISEASES; RIBOFLAVIN; POSTERIOR; ASSOCIATION; CHILDREN; ANTERIOR;
D O I
10.3390/diagnostics11112120
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The keratometer keratoconus index (KKI) is a diagnostic index for the risk of keratoconus calculated from autokeratometer test values. We partially modified the KKI equation and assessed it without limiting the target age and severity of keratoconus. This retrospective study included 179 eyes of 99 patients with keratoconus and 468 eyes from 235 normal controls. In the modified KKI, oblique astigmatism or against-the-rule astigmatism was defined as & GE;1D astigmatism. KKI diagnostic power was analyzed in subgroups of < 50 and & GE;50-year-old patients, and at different keratoconus stages. Although the sensitivity of modified KKI was comparable with that of original KKI (92.7% vs. 95.5%), modified KKI specificity was significantly higher (79.7% vs. 68.6%) (p = 0.0001). Using the modified KKI, sensitivity reached 100% (4/4) and specificity, 63.5% (33/52), in & GE;50-year-old patients, while overall sensitivity in keratoconus & GE;stage 2 was 100% (30/30). In conclusion, the modified KKI proved to be effective in keratoconus screening at all stages. However, it should be noted that false-positive frequency is higher in & GE;50-year-old patients.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Keratoconus Screening Using Values Derived From Auto-Keratometer Measurements: A Multicenter Study
    Kojima, Takashi
    Nishida, Tomoya
    Nakamura, Tomoaki
    Tamaoki, Akeno
    Hasegawa, Asato
    Takagi, Yuki
    Sato, Hiroyuki
    Ichikawa, Kazuo
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2020, 215 : 127 - 134
  • [2] Biomechanical diagnosis of keratoconus: evaluation of the keratoconus match index and the keratoconus match probability
    Labiris, Georgios
    Gatzioufas, Zisis
    Sideroudi, Haris
    Giarmoukakis, Athanassios
    Kozobolis, Vassilios
    Seitz, Berthold
    ACTA OPHTHALMOLOGICA, 2013, 91 (04) : E258 - E262
  • [3] Screening of Keratoconus and Keratoconus Suspects Using Rotating Scheimpflug Corneal Topography
    Shimabukuro, M.
    Higashiura, R.
    Fuchihata, M.
    Nakagawa, T.
    Koh, S.
    Inoue, T.
    Maeda, N.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [4] Diagnostic capacity of the keratoconus match index and keratoconus match probability in subclinical keratoconus
    Labiris, Georgios
    Giarmoukakis, Athanassios
    Gatzioufas, Zisis
    Sideroudi, Hans
    Kozobolis, Vassilios
    Seitz, Berthold
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (06) : 999 - 1005
  • [5] Misclassification of Eyes With Progressive Keratoconus Using the KISA% Index
    Hammoud, Bassel
    Dupps Jr, William J.
    Scarcelli, Giuliano
    Randleman, J. Bradley
    JOURNAL OF REFRACTIVE SURGERY, 2024, 40 (09) : e614 - e624
  • [6] Vernal keratoconjunctivitis and keratoconus
    Wajnsztajn, Denise
    Solomon, Abraham
    CURRENT OPINION IN ALLERGY AND CLINICAL IMMUNOLOGY, 2021, 21 (05) : 507 - 514
  • [7] Review: Keratoconus in Asia
    Kok, Yee Onn
    Tan, Grace Feng Ling
    Loon, Seng Chee
    CORNEA, 2012, 31 (05) : 581 - 593
  • [8] Correlation of the KISA% index and Scheimpflug tomography in 'normal', 'subclinical', 'keratoconus-suspect' and 'clinically manifest' keratoconus eyes
    Steinberg, Johannes
    Aubke-Schultz, Silke
    Frings, Andreas
    Huelle, Jan
    Druchkiv, Vasyl
    Richard, Gisbert
    Katz, Toam
    Linke, Stephan J.
    ACTA OPHTHALMOLOGICA, 2015, 93 (03) : e199 - e207
  • [9] Logistic index for keratoconus detection and severity scoring (Logik)
    Issarti, Ikram
    Consejo, Alejandra
    Jimenez-Garcia, Marta
    Kreps, Elke O.
    Koppen, Carina
    Rozema, Jos J.
    COMPUTERS IN BIOLOGY AND MEDICINE, 2020, 122 (122)
  • [10] A Potential Screening Index of Corneal Biomechanics in Healthy Subjects, Forme Fruste Keratoconus Patients and Clinical Keratoconus Patients
    Tian, Lei
    Qin, Xiao
    Zhang, Hui
    Zhang, Di
    Guo, Li-Li
    Zhang, Hai-Xia
    Wu, Ying
    Jie, Ying
    Li, Lin
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2021, 9