Trace elemental analysis of commercial bottled drinking water by a portable total reflection X-ray fluorescence spectrometer

被引:32
|
作者
Kunimura, Shinsuke [1 ]
Kawai, Jun [1 ]
机构
[1] Kyoto Univ, Dept Mat Sci & Engn, Sakyo Ku, Kyoto 6068501, Japan
关键词
17;
D O I
10.2116/analsci.23.1185
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A portable total reflection X-ray fluorescence spectrometer was applied to commercial bottled drinking water containing 62 ppb (mu g/L) V, 7.5 ppm (mg/L) Ca, and I ppm K. A dry residue, prepared by repeated pipetting and drying a 20 mu L portion of drinking water on a sample holder six times, and another dry residue, prepared by pipetting and drying a 120 mu L portion of the drinking water, were measured. The size of the dry residue of the drinking water was reduced by repeating the pipetting and drying. Calcium and potassium were easily detected in these two dry residues. Vanadium was detected in the dry residue prepared by repeated pipetting and drying. Vanadium, calcium, and potassium were quantified by using an internal standard, and the average quantified concentrations of V, Ca, and K were 91 ppb, 8.2 ppm, and 1.5 ppm, respectively. Although these quantified concentrations deviated from the certified concentrations (62 ppb of V, 7.5 ppm of Ca, and I ppm of K), it was possible for this portable spectrometer to estimate approximate concentrations.
引用
收藏
页码:1185 / 1188
页数:4
相关论文
共 50 条
  • [31] Determination of the trace element content in Bulgarian bottled potable waters by total reflection X-ray fluorescence analysis
    Georgieva, R. H.
    Detcheva, A. K.
    Karadjov, M. G.
    Mitsiev, S. E.
    Jordanov, J. H.
    Ivanova, E. H.
    BULGARIAN CHEMICAL COMMUNICATIONS, 2014, 46 (04): : 840 - 844
  • [32] Analysis of Metals Eluted from Stainless Steel in Ultrapure Water Using a Portable Total Reflection X-ray Fluorescence Spectrometer
    Kako, Arisa
    Kunimura, Shinsuke
    BUNSEKI KAGAKU, 2022, 71 (7-8) : 431 - 435
  • [33] Environmental Trace-Element Analysis Using a Benchtop Total Reflection X-Ray Fluorescence Spectrometer
    Hagen Stosnach
    Analytical Sciences, 2005, 21 : 873 - 876
  • [34] Environmental trace-element analysis using a benchtop total reflection X-ray, fluorescence spectrometer
    Stosnach, H
    ANALYTICAL SCIENCES, 2005, 21 (07) : 873 - 876
  • [35] A depth profile fitting model for a commercial total reflection X-ray fluorescence spectrometer
    Mori, Y
    Uemura, K
    Shimanoe, K
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (07) : 823 - 828
  • [36] Total Reflection X-Ray Fluorescence Analysis of Natural and Drinking Waters
    Oskolok, K. V.
    Monogarova, O. V.
    Alov, N. V.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2018, 73 (11) : 1093 - 1097
  • [37] Total Reflection X-Ray Fluorescence Analysis of Natural and Drinking Waters
    K. V. Oskolok
    O. V. Monogarova
    N. V. Alov
    Journal of Analytical Chemistry, 2018, 73 : 1093 - 1097
  • [38] A portable total reflection X-ray fluorescence spectrometer with a diamond-like carbon coated X-ray reflector
    Kunimura, Shinsuke
    Ohmori, Hitoshi
    ANALYST, 2012, 137 (02) : 312 - 314
  • [39] Elemental analysis of environmental samples by total reflection X-ray fluorescence: A review
    Klockenkamper, R
    vonBohlen, A
    X-RAY SPECTROMETRY, 1996, 25 (04) : 156 - 162
  • [40] Development and Characterization of a Portable Total Reflection X-ray Fluorescence System Using a Waveguide for Trace Elements Analysis
    da Costa, Ana Cristina M.
    de Araujo, Ubiratan B.
    de Jesus, Edgar F. O.
    Anjos, Marcelino J.
    Lopes, Ricardo T.
    ANALYTICAL SCIENCES, 2014, 30 (10) : 955 - 960