Microcalorimetry: a novel method for detection of microbial contamination in platelet products

被引:82
作者
Trampuz, Andrej
Salzmann, Simone
Antheaume, Jeanne
Daniels, Alma U.
机构
[1] Univ Basel Hosp, Div Infect Dis & Hosp Epidemiol, CH-4031 Basel, Switzerland
[2] Univ Basel, Lab Orthoped Biomech, Basel, Switzerland
关键词
D O I
10.1111/j.1537-2995.2007.01336.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: Measuring heat from replicating microorganisms in culture may be a rapid, accurate, and simple screening method for platelets (PLTs). Microcalorimetry for detection of microorganisms in in vitro contaminated PLT products was evaluated. STUDY DESIGN AND METHODS: Staphylococcus epidermidis, Staphylococcus aureus, Streptococcus sanguinis, Escherichia coli, Propionibacterium acnes, and Candida albicans were inoculated in single-donor apheresis PLTs to achieve target concentrations of 10(5), 10(3), 10, or 1 colony-forming units (CFU) per mL of PLTs. Contaminated PLTs in growth medium were incubated at 37 degrees C for 5 days in a calorimeter. Positivity was defined as heat flow of at least 10 mu W above the lowest value of the power-time curve. RESULTS: With microcalorimetry, inocula of 10 CFUs per mL PLTs could be detected with the following detection times: S. epidermidis (31.65 hr), S. aureus (24.24 hr), S. sanguinis (7.82 hr), E. coli (7.53 hr), P acnes (73.57 hr), and C. albicans (43.77 hr). The detection time was less than 4 hr at 101 CFUs per mL PLTs for S. aureus, S. sanguinis, and E coli. Noncontaminated PLTs remained negative. The total heat ranged from 2.8 (S. sanguinis) to 8.3 J (E coli). The shape of the power-time curve was species-specific and independent from the initial concentration of microorganisms. CONCLUSION: The detection limit of microcalorimetry was 1 to 10 CFUs per mL PLTs. Microcalorimetry is a promising novel method for detection of contaminated PLTs. Applying this method to all PLT products could reduce the frequency of transfusion-related sepsis and prolong the shelf life of PLTs.
引用
收藏
页码:1643 / 1650
页数:8
相关论文
共 48 条
[1]   Protecting the blood supply from emerging pathogens: The role of pathogen inactivation [J].
Allain, JP ;
Bianco, C ;
Blajchman, MA ;
Brecher, ME ;
Busch, M ;
Leiby, D ;
Lin, L ;
Stramer, S .
TRANSFUSION MEDICINE REVIEWS, 2005, 19 (02) :110-126
[2]   CELL-GROWTH EXPERIMENTS USING A MICROCALORIMETRIC VESSEL EQUIPPED WITH OXYGEN AND PH ELECTRODES [J].
BACKMAN, P ;
WADSO, I .
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 1991, 23 (04) :283-293
[3]   Differential effects of catecholamines on in vitro growth of pathogenic bacteria [J].
Belay, T ;
Sonnenfeld, G .
LIFE SCIENCES, 2002, 71 (04) :447-456
[4]  
Blajchman M A, 2002, Dev Biol (Basel), V108, P59
[5]   Improving the bacteriological safety of platelet transfusions [J].
Blajchman, MA ;
Goldman, M ;
Baeza, F .
TRANSFUSION MEDICINE REVIEWS, 2004, 18 (01) :11-24
[6]   The continuing risk of transfusion-transmitted infections [J].
Blajchman, Morris A. ;
Vamvakas, Eleftherios C. .
NEW ENGLAND JOURNAL OF MEDICINE, 2006, 355 (13) :1303-1305
[7]  
Brecher Mark E, 2004, Curr Hematol Rep, V3, P121
[8]   Growth of bacteria in inoculated platelets: implications for bacteria detection and the extension of platelet storage [J].
Brecher, ME ;
Holland, PV ;
Pineda, AA ;
Tegtmeier, GE ;
Yomtovian, R .
TRANSFUSION, 2000, 40 (11) :1308-1312
[9]   Bacterial contamination of blood components [J].
Brecher, ME ;
Hay, SN .
CLINICAL MICROBIOLOGY REVIEWS, 2005, 18 (01) :195-+
[10]   Evaluation of an automated culture system for detecting bacterial contamination of platelets: an analysis with 15 contaminating organisms [J].
Brecher, ME ;
Means, N ;
Jere, CS ;
Heath, D ;
Rothenberg, S ;
Stutzman, LC .
TRANSFUSION, 2001, 41 (04) :477-482