Comparison between laser speckle contrast imaging and laser Doppler imaging to assess skin blood flow in humans

被引:80
作者
Millet, C.
Roustit, M.
Blaise, S.
Cracowski, J. L.
机构
[1] Grenoble Univ Hosp, Grenoble Clin Res Ctr, Inserm CIC3, Grenoble, France
[2] Univ Grenoble 1, Inserm U1042 HP2, Grenoble 1, France
关键词
Microcirculation; Laser speckle contrast imaging; Laser Doppler; Thermal hyperemia; MICROVASCULAR REACTIVITY; ENDOTHELIAL FUNCTION; IN-VIVO; PERFUSION; WOMEN; CYCLE;
D O I
10.1016/j.mvr.2011.06.006
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective: We tested the linearity between skin blood flux recorded with laser speckle contrast imaging (LSCI) and laser Doppler imaging (LDI), comparing different ways of expressing data. A secondary objective was to test within-subject variability of baseline flux with the two techniques. Methods: We performed local heating at 36, 39, 42, and 44 degrees C on the forearm of healthy volunteers, and measured cutaneous blood flux with LDI and LSCI. Biological zero (BZ) was obtained by occluding the brachial artery. We expressed data as raw arbitrary perfusion units (APUs) and as a percentage increase from baseline (%BL), with and without subtracting BZ. Inter-site variability was expressed as a within subject coefficient of variation (CV). Results: Twelve participants were enrolled. Inter-site variability at baseline was lower with LSCI (CV = 9.2%) than with LDI (CV = 20.7%). We observed an excellent correlation between both techniques when data were expressed as raw APUs or APU - BZ (R = 0.90: p<0.001). The correlation remained correct for %BL (R = 0.77, p<0.001), but decreased for %BL - BZ (R = 0.44, p = 0.003). Bland-Altman plots revealed a major proportional bias between the two techniques. Conclusion: This study suggests that skin blood flux measured with LSCI is linearly related to the LDI signal over a wide range of perfusion. Subtracting BZ does not affect this linearity but introduces variability in baseline flux, thus decreasing the correlation when data are expressed as a function of baseline. Finally, systematic bias makes it impossible to assimilate arbitrary perfusion units provided by the two systems. (C) 2011 Elsevier Inc. All rights reserved.
引用
收藏
页码:147 / 151
页数:5
相关论文
共 21 条
  • [1] Bland M., 1995, INTRO MED STAT, V2nd
  • [2] Laser Doppler, speckle and related techniques for blood perfusion mapping and imaging
    Briers, JD
    [J]. PHYSIOLOGICAL MEASUREMENT, 2001, 22 (04) : R35 - R66
  • [3] 17β-Estradiol and Progesterone Independently Augment Cutaneous Thermal Hyperemia But Not Reactive Hyperemia
    Brunt, Vienna E.
    Miner, Jennifer A.
    Meendering, Jessica R.
    Kaplan, Paul F.
    Minson, Christopher T.
    [J]. MICROCIRCULATION, 2011, 18 (05) : 347 - 355
  • [4] Linear response range characterization and in vivo application of laser speckle imaging of blood flow dynamics
    Choi, Bernard
    Ramirez-San-Juan, Julio C.
    Lotfi, Justin
    Nelson, J. Stuart
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2006, 11 (04)
  • [5] Methodological issues in the assessment of skin microvascular endothelial function in humans
    Cracowski, Jean-Luc
    Minson, Christopher T.
    Salvat-Melis, Muriel
    Halliwill, John R.
    [J]. TRENDS IN PHARMACOLOGICAL SCIENCES, 2006, 27 (09) : 503 - 508
  • [6] In vivo comparison of scanning technique and wavelength in laser Doppler perfusion imaging: measurement in knee ligaments of adult rabbits
    Forrester, K
    Doschak, M
    Bray, R
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (06) : 581 - 586
  • [7] Characteristics of laser Doppler perfusion imaging in vitro and in vivo
    Kernick, DP
    Shore, AC
    [J]. PHYSIOLOGICAL MEASUREMENT, 2000, 21 (02) : 333 - 340
  • [8] Microvascular function has no menstrual-cycle-dependent variation in healthy ovulatory women
    Ketel, Iris J. G.
    Stehouwer, Coen D. A.
    Serne, Erik H.
    Poel, Danielle M.
    Groot, Leonieke
    Kager, Cathrien
    Hompes, Peter G. A.
    Homburg, Roy
    Twisk, Jos W.
    Smulders, Yvo M.
    Lambalk, Cornelis B.
    [J]. MICROCIRCULATION, 2009, 16 (08) : 714 - 724
  • [9] Laser speckle contrast imaging accurately measures blood flow over moving skin surfaces
    Mahe, Guillaume
    Rousseau, Pascal
    Durand, Sylvain
    Bricq, Stephanie
    Leftheriotis, Georges
    Abraham, Pierre
    [J]. MICROVASCULAR RESEARCH, 2011, 81 (02) : 183 - 188
  • [10] Laser-Doppler imaging of forearm skin: Perfusion features and dependence of the biological zero on heat-induced hyperemia
    Mayrovitz, HN
    Leedham, JA
    [J]. MICROVASCULAR RESEARCH, 2001, 62 (01) : 74 - 78