Sensitivity and directivity measurement of ultrasonic transducer with polymer-powder matching layer

被引:3
|
作者
Fazlyyyakhmatov, Marsel [1 ]
机构
[1] Kazan Volga Reg Fed Univ, 18 Kremlyovskaya Str, Kazan 420008, Russia
来源
关键词
PHASE CALIBRATION;
D O I
10.1088/1742-6596/1038/1/012109
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The results of an experimental study of the sensitivity and directivity pattern of the ultrasonic transducers are presented. Data was obtained for transducers with central frequency at 2.5 and 5 MHz with protective matching layers, in comparison with transducers without matching layers. Increase of transducer sensitivity is shown. In addition, it is shown that the directivity patterns for transducers with the matching layer and without it are practically identical. Also was observed that the directivity pattern gets sharper as the frequency of the transducer increases.
引用
收藏
页数:8
相关论文
共 45 条
  • [1] Radiation Pattern of Ultrasonic Transducer with Polymer-Powder Matching Layer
    Fazlyyyakhmatov, Marsel
    2018 19TH INTERNATIONAL CONFERENCE OF YOUNG SPECIALISTS ON MICRO/NANOTECHNOLOGIES AND ELECTRON DEVICES (EDM 2018), 2018, : 363 - 366
  • [2] Plasma application of protective polymer-powder coatings to ultrasonic sensors
    Gavrilova V.A.
    Kashapov N.F.
    Kashapov R.N.
    Biomedical Engineering, 2011, 45 (5) : 198 - 200
  • [3] Measurement methods of ultrasonic transducer sensitivity
    Xiao, Dingguo
    Fan, Qiong
    Xu, Chunguang
    Zhang, Xiuhua
    ULTRASONICS, 2016, 68 : 150 - 154
  • [4] DEVELOPMENT OF DOUBLE MATCHING LAYER FOR ULTRASONIC POWER TRANSDUCER
    Hwang, Gunn
    Youm, WooSub
    Lee, Sung Q.
    PROCEEDINGS OF THE 23RD INTERNATIONAL CONGRESS ON SOUND AND VIBRATION: FROM ANCIENT TO MODERN ACOUSTICS, 2016,
  • [5] Effect of wavefront distortion by propagation through inhomogeneous layer to the directivity of ultrasonic transducer
    Sugimoto, Kazuko
    Tabei, Makoto
    Ueda, Mitsuhiro
    Japanese Journal of Applied Physics, Part 1: Regular Papers & Short Notes & Review Papers, 1995, 34 (5 B): : 2802 - 2807
  • [6] Effect of composition of the matching layer, directivity, temperature and pressure on ultrasonic signals in the tank
    Choi, Yonghyun
    Kim, Hongju
    Choe, Munseok
    JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, 2025, 39 (02) : 949 - 956
  • [7] Study of polymer ultrasonic transducer and complex ultrasonic transducer with PZT PVDF multi-layer structure
    Liu, XZ
    Ye, SG
    Gong, XF
    Zhang, WY
    Lu, RR
    ACUSTICA, 1999, 85 (03): : 420 - 426
  • [8] Enhancement of Ultrasonic Transducer Bandwidth by Acoustic Impedance Gradient Matching Layer
    Zhu, Ke
    Ma, Jinpeng
    Qi, Xudong
    Shen, Bingzhong
    Liu, Yang
    Sun, Enwei
    Zhang, Rui
    SENSORS, 2022, 22 (20)
  • [9] Multiple acoustical matching layer design of ultrasonic transducer for medical application
    Chen, YC
    Wu, S
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 2002, 41 (10): : 6098 - 6107
  • [10] Multiple acoustical matching layer design of ultrasonic transducer for medical application
    Chen, Yeong-Chin
    Wu, Sean
    Chen, Y.-C., 1600, Japan Society of Applied Physics (41): : 6098 - 6107