On the use of computed radiography plates for quality assurance of intensity modulated radiation therapy dose distributions

被引:3
|
作者
Day, R. A. [1 ]
Sankar, A. P. [1 ]
Nailon, W. H. [1 ]
MacLeod, A. S. [1 ]
机构
[1] Western Gen Hosp, Dept Oncol Phys, Edinburgh Canc Ctr, Edinburgh EH4 2XU, Midlothian, Scotland
关键词
computed radiography; intensity modulated radiation therapy; IMRT; quality assurance; dosimetry; FILM DOSIMETRY; QUANTITATIVE-EVALUATION; DIODE-ARRAY; IMRT; VERIFICATION; SOFTWARE; PHANTOM;
D O I
10.1118/1.3525841
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: As traditional film is phased out in most radiotherapy centers, computed radiography (CR) systems are increasingly being purchased as a replacement. CR plates can be used for patient imaging, but may also be used for a variety of quality assurance (QA) purposes and can be calibrated in terms of dose. This study looks at their suitability for verification of intensity modulated radiation therapy (IMRT) dose distributions. Methods: A CR plate was calibrated in terms of the relative dose and the stability of response over 1 year was studied. The effect of exposing the CR plate to ambient light and of using different time delays before scanning was quantified. The CR plate was used to verify the relative dose distributions for ten IMRT patients and the results were compared to those obtained using a two dimensional (2D) diode array. Results: Exposing the CR plate to 10 s of ambient light between irradiation (174 cGy) and scanning erased approximately 80% of the signal. Changes in delay time between irradiation and scanning also affected the measurement results. The signal on the plate was found to decay at a rate of approximately 3.6 cGy/min in the first 10 min after irradiation. The use of a CR plate for IMRT patient-specific QA resulted in a significantly lower distance to agreement (DTA) and gamma pass rate than when using a 2D diode array for the measurement. This was primarily due to the over-response of the CR phosphor to low energy scattered radiation. For the IMRT QA using the CR plate, the average gamma pass rate was 97.3%. For the same IMRT QA using a diode array, the average gamma pass rate was 99.7%. The gamma criteria used were 4% dose difference and 4 mm DTA for head and neck treatments and 3% dose difference and 3 mm DTA for prostate treatments. The gamma index tolerance was 1. The lowest 10% of the dose distribution was excluded from all gamma and DTA analyses. Conclusions: Although the authors showed that CR plates can be used for patient specific IMRT QA, the practical problems such as the over-response to low energy scatter and signal fading with light exposure and time mean that alternative detectors such as radiochromic film or diode arrays will be a more sensible choice for most radiotherapy departments. (c) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3525841]
引用
收藏
页码:632 / 645
页数:14
相关论文
共 50 条
  • [1] A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy
    Son, Jaeman
    Baek, Taesung
    Lee, Boram
    Shin, Dongho
    Park, Sung Yong
    Park, Jeonghoon
    Lim, Young Kyung
    Lee, Se Byeong
    Kim, Jooyoung
    Yoon, Myonggeun
    RADIOLOGY AND ONCOLOGY, 2015, 49 (03) : 307 - 313
  • [2] Comparison of action levels for patient-specific quality assurance of intensity modulated radiation therapy and volumetric modulated arc therapy treatments
    Mancuso, Gordon M.
    Fontenot, Jonas D.
    Gibbons, John P.
    Parker, Brent C.
    MEDICAL PHYSICS, 2012, 39 (07) : 4378 - 4385
  • [3] Evaluation criteria for film based intensity modulated radiation therapy quality assurance
    Anjum, Muhammad Naeem
    Parker, William
    Ruo, Russell
    Afzal, Muhammad
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2010, 26 (01): : 38 - 43
  • [4] Quality assurance of intensity-modulated radiation therapy
    Palta, Jatinder R.
    Li, Chihray
    Li, Jonathan G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : S108 - S112
  • [5] The impacts of dose rate in sliding window intensity modulated radiation therapy quality assurance
    Gul, O., V
    Inan, G.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (03): : 625 - 632
  • [6] Design of an anthropomorphic intensity modulated radiation therapy quality assurance phantom
    Radford, DA
    Followill, DS
    Hanson, WF
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 197 - 200
  • [7] Toward automated quality assurance for intensity-modulated radiation therapy
    Low, DA
    Dempsey, JF
    Markman, J
    Mutic, S
    Klein, EE
    Sohn, JW
    Purdy, JA
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (02): : 443 - 452
  • [8] Quality assurance of helical tomotherapy intensity modulated radiation therapy
    Xu, Shouping
    Deng, Xiaowu
    Dai, Xiangkun
    Wang, Lianyuan
    Wang, Yunlai
    Xie, Chuanbin
    Ge, Ruigang
    Gong, Hanshun
    APCMBE 2008: 7TH ASIAN-PACIFIC CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING, 2008, 19 : 447 - +
  • [9] Dosimetric performance of an enhanced dose range radiographic film for intensity-modulated radiation therapy quality assurance
    Olch, AJ
    MEDICAL PHYSICS, 2002, 29 (09) : 2159 - 2168
  • [10] Strategies for quality assurance of intensity modulated radiation therapy
    Benedek, Hunor
    Isacsson, Ulf
    Olevik-Dunder, Maria
    Westermark, Mathias
    Hallstrom, Per
    Olofsson, Jorgen
    Gustafsson, Magnus
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573