This paper presents a method for providing volitional control of a powered knee prosthesis during nonweight-bearing activity such as sitting. The method utilizes an impedance framework, such that the joint can be programmed with a given stiffness and damping that reflects the nominal impedance properties of an intact joint. Volitional movement of the knee joint is commanded via the stiffness set-point angle of the joint impedance, which is commanded by the user as a function of the measured surface electromyogram (EMG) from the hamstring and quadriceps muscles of the residual limb. Rather than using the respective EMG measurements from these muscles to directly command the flexion or extension set point of the knee, the presented approach utilizes a combination of quadratic discriminant analysis and principal component analysis to align the user's intent to flex or extend the knee joint with the pattern of measured EMG. The approach was implemented on three transfemoral amputees, and their ability to control knee movement was characterized by a set of knee joint trajectory tracking tasks. Each amputee subject also performed the same set of trajectory tracking tasks with his sound side (intact) knee joint. The average root mean square trajectory tracking errors of the prosthetic knee employing the EMG-based volitional control and the intact knee of the three subjects were 6.2 degrees and 5.2 degrees, respectively.
引用
收藏
页码:144 / 151
页数:8
相关论文
共 24 条
  • [1] DEVELOPMENT OF AN ABOVE-KNEE PROSTHESIS EQUIPPED WITH A MICROCOMPUTER-CONTROLLED KNEE-JOINT - 1ST TEST-RESULTS
    AEYELS, B
    PEERAER, L
    VANDERSLOTEN, J
    VANDERPERRE, G
    [J]. JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (03): : 199 - 202
  • [2] Aeyels B., 1995, IEEE 17th Annual Conference of Engineering in Medicine and Biology Society, V2, P1315
  • [3] [Anonymous], 2002, Principal components analysis
  • [4] [Anonymous], 2004, WILEY SER PROB STAT
  • [5] Powered ankle-foot prosthesis to assist level-ground and stair-descent gaits
    Au, Samuel
    Berniker, Max
    Herr, Hugh
    [J]. NEURAL NETWORKS, 2008, 21 (04) : 654 - 666
  • [6] Au SK, 2005, INT C REHAB ROBOT, P375
  • [7] Locomotor adaptation to a powered ankle-foot orthosis depends on control method
    Cain, Stephen M.
    Gordon, Keith E.
    Ferris, Daniel P.
    [J]. JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2007, 4 (1)
  • [8] DONATH M, 1974, THESIS MIT CAMBRIDGE
  • [9] An improved powered ankle-foot orthosis using proportional myoelectric control
    Ferris, Daniel P.
    Gordon, Keith E.
    Sawicki, Gregory S.
    Peethambaran, Arnmanath
    [J]. GAIT & POSTURE, 2006, 23 (04) : 425 - 428
  • [10] An ankle-foot orthosis powered by artificial pneumatic muscles
    Ferris, DP
    Czerniecki, JM
    Hannaford, B
    [J]. JOURNAL OF APPLIED BIOMECHANICS, 2005, 21 (02) : 189 - 197