Porcine liver: Morphologic characteristics and cell viability at experimental radiofrequency ablation with internally cooled electrodes

被引:50
|
作者
Ng, KK
Lam, CM
Poon, RT
Shek, TW
Yu, WC
To, JY
Wo, YH
Lau, CP
Tang, TC
Ho, DW
Fan, ST
机构
[1] Univ Hong Kong, Queen Mary Hosp, Dept Surg, Hong Kong, Hong Kong, Peoples R China
[2] Univ Hong Kong, Queen Mary Hosp, Dept Pathol, Hong Kong, Hong Kong, Peoples R China
[3] Univ Hong Kong, Queen Mary Hosp, Ctr Study Liver Dis, Hong Kong, Hong Kong, Peoples R China
关键词
D O I
10.1148/radiol.2352040425
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate morphologic characteristics and cell viability of radiofrequency ablation zones in porcine liver. MATERIALS AND METHODS: Approval of the study protocol was obtained from the Ethics Committee on Use of Live Animals for Teaching and Research at University of Hong Kong. Internally cooled electrodes were used to produce 120 ablated zones ex vivo and 60 ablated zones in vivo with single electrodes (1-, 2-, and 3-cm exposed lengths) or clustered electrodes (1.0-, 2.0-, and 2.5-cm exposed lengths) at 4, 8, 12, and 16 minutes of ablation (ex vivo) and 8 and 12 minuted of ablation (in vivo). Morphologic measurements of each ablated zone were preformed. Cell viability in each ablated zone was assessed qualitatively with histochemical staining and quantitatively with measurement of intracellular adenosine 5'-triphosphate (ATP) concentration. RESULTS: Exposed length of electrode (coefficient = 0.79, standard error = 0.04, P < .001), duration of ablation (coefficient = 0.14, standard error = 0.01, P < .001), and clustered electrode design (coefficient = 1.21, standard error = 0.005, P < .001) were independent factors that affected minimal transverse diameter and volume of ablated zone in ex vivo study. Similar morphologic characteristics existed among ablated in in vivo study. Mean distance of ablation beyond the electrode tip remained constant (ex vivo, 1.0 cm +/- 0.08 [standard deviation]; in vivo, 0.5 cm +/- 0.05) regardless of different ablation conditions. Histochemical staining revealed no viable hepatocytes from center to margins of white zone (9.5 x 10(-12) mol/mu g DNA +/- 1.43) was lower than that in red zone (4088 x 10(-12) mol/mu g DNA +/- 65.97, P < .001) and in adjacent normal liver (4528 x 10(-12) mol/mu g DNA +/- 52.74, P < .001). CONCLUSION: Distance of ablation beyond the tip of the electrode remained constant (ex vivo, 1.0 cm; in vivo, 0.5 cm) with different conditions of ablation. complete and uniform cellular destruction was achieved in the white zone of ablated area. (C) RSNA, 2005.
引用
收藏
页码:478 / 486
页数:9
相关论文
共 50 条
  • [1] Radiofrequency Ablation Zones in Ex Vivo Bovine and In Vivo Porcine Livers: Comparison of the Use of Internally Cooled Electrodes and Internally Cooled Wet Electrodes
    Cha, Jihoon
    Choi, Dongil
    Lee, Min Woo
    Rhim, Hyunchul
    Kim, Young-sun
    Lim, Hyo K.
    Yoon, Jung Hwan
    Park, Cheol Keun
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 32 (06) : 1235 - 1240
  • [2] Radiofrequency Ablation Zones in Ex Vivo Bovine and In Vivo Porcine Livers: Comparison of the Use of Internally Cooled Electrodes and Internally Cooled Wet Electrodes
    Jihoon Cha
    Dongil Choi
    Min Woo Lee
    Hyunchul Rhim
    Young-sun Kim
    Hyo K. Lim
    Jung Hwan Yoon
    Cheol Keun Park
    CardioVascular and Interventional Radiology, 2009, 32 : 1235 - 1240
  • [3] Radiofrequency ablation of the porcine liver in vivo: Increased coagulation with an internally cooled perfusion electrode
    Lee, JM
    Han, JK
    Chang, JM
    Chung, SY
    Kim, SH
    Lee, JY
    Lee, MW
    Choi, BI
    ACADEMIC RADIOLOGY, 2006, 13 (03) : 343 - 352
  • [4] Multipolar radiofrequency ablation with internally cooled electrodes:: Experimental study in ex vivo bovine liver with mathernatic modeling
    Clasen, S
    Schmidt, D
    Boss, A
    Dietz, K
    Kröber, SM
    Claussen, CD
    Pereira, PL
    RADIOLOGY, 2006, 238 (03) : 881 - 890
  • [5] Radiofrequency renal ablation: In vivo comparison of internally cooled, multitined expandable and internally cooled perfusion electrodes
    Lee, JM
    Han, JK
    Chang, JM
    Chung, SY
    Son, KR
    Kim, SH
    Lee, JY
    Choi, BI
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (03) : 549 - 556
  • [6] Percutaneous radiofrequency ablation using internally cooled wet electrodes for treatment of colorectal liver metastases
    Yoon, H. M.
    Kim, J. H.
    Shin, Y. M.
    Won, H. J.
    Kim, P. N.
    CLINICAL RADIOLOGY, 2012, 67 (02) : 122 - 127
  • [7] CT findings after radiofrequency ablation in rabbit livers: Comparison of internally cooled electrodes, perfusion electrodes, and internally cooled perfusion electrodes
    Kim, Seung Kwon
    Gu, Myung Seo
    Hong, Hyun Pyo
    Choi, Dongil
    Chae, Seoung Wan
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 18 (11) : 1417 - 1427
  • [8] Percutaneous Radiofrequency Ablation with Internally Cooled versus Internally Cooled Wet Electrodes for Small Subphrenic Hepatocellular Carcinomas
    Kim, Jin Hyoung
    Kim, Pyo Nyun
    Won, Hyung Jin
    Shin, Yong Moon
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (03) : 351 - 356
  • [9] Comparison of ablation performance between dual internally cooled wet tip and conventional dual internally cooled tip radiofrequency electrodes: an experimental study in ex vivo bovine liver
    Cha, Dong Ik
    Lee, Min Woo
    Jeong, Woo Kyoung
    Ha, Sang Yun
    Ahn, Soo Hyun
    Rhim, Hyunchul
    Lim, Hyo Keun
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2021, 38 (01) : 332 - 340
  • [10] Radiofrequency hepatic ablation with internally cooled electrodes and hybrid applicators with distant saline infusion using an in vivo porcine model
    Burdio, F.
    Navarro, A.
    Berjano, E. J.
    Burdio, J. M.
    Gonzalez, A.
    Guemes, A.
    Sousa, R.
    Rufas, M.
    Cruz, I.
    Castiella, T.
    Lozano, R.
    Lequerica, J. L.
    Grande, L.
    EJSO, 2008, 34 (07): : 822 - 830