Modification of the sensitivity of glucose sensor implanted into subcutaneous tissue

被引:0
作者
ThomeDuret, V
Gangnerau, MN
Zhang, Y
Wilson, GS
Reach, G
机构
[1] HOP HOTEL DIEU,SERV DIABETOL,INSERM,U341,F-75004 PARIS,FRANCE
[2] UNIV KANSAS,DEPT CHEM,LAWRENCE,KS 66045
来源
DIABETES & METABOLISM | 1996年 / 22卷 / 03期
关键词
glucose sensor; sensitivity; subcutaneous tissue; rat;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The mechanism of reducing the glucose sensitivity of sensors implanted into the subcutaneous tissue of the normal rat was evaluated (n=10) by comparing sensitivities observed in vitro and in vivo. In vivo sensitivity was significantly lower than that observed in vitro before implantation (p < 0.005). Most interestingly, in vitro sensitivity immediately after explantation did not differ from that in vivo and increased progressively during rinsing (p < 0.02 after 30 min). These results demonstrate that the reduction of in vivo sensitivity was not due to a local factor or factors but to a reversible alteration of the glucose sensor characteristics induced in vivo by some local factor(s). This suggests that modifications of the outer sensor membrane, the nature of which remains to be determined, may prevent this effect and resolve the problem.
引用
收藏
页码:174 / 178
页数:5
相关论文
共 14 条
  • [11] POITOUT V, 1991, T AM SOC ART INT ORG, V37, pM298
  • [12] REACH G, 1993, DIABETES ANN, P332
  • [13] VELHO G, 1989, BIOMED BIOCHIM ACTA, V48, P957
  • [14] VELHO G, 1988, DIABETES NUTR METAB, V1, P227