Objective measurement of intraocular forward light scatter using Hartmann-Shack spot patterns from clinical aberrometers -: Model-eye and human-eye study

被引:11
作者
Cervino, Alejandro [1 ]
Bansal, Dheeraj [2 ]
Hosking, Sarah L. [3 ,4 ,5 ]
Montes-Mico, Robert
机构
[1] Univ Valencia, Dept Opt, Optometry Res Grp, E-46100 Valencia, Spain
[2] Optopol Technol SA, Zawiercie, Poland
[3] City Univ London, Dept Optometry & Vis Sci, London EC1V 0HB, England
[4] Aston Univ, Sch Life & Hlth Sci, Birmingham B4 7ET, W Midlands, England
[5] Univ Melbourne, Dept Ophthalmol, Melbourne, Vic, Australia
关键词
D O I
10.1016/j.jcrs.2008.03.027
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To apply software-based image-analysis tools to objectively determine intraocular scatter determined from clinically derived Hartmann-Shack patterns. SETTING: Aston Academy of Life Sciences, Aston University, Birmingham, United Kingdom, and Department of Optics, University of Valencia, Valencia, Spain. METHODS: Purpose-designed image-analysis software was used to quantify scatter from centroid patterns obtained using a clinical Hartmann-Shack analyzer (WASCA, Zeiss/Meditec). Three scatter values, as the maximum standard deviation within a lenslet for all lenslets in the pattern, were obtained in 6 model eyes and 10 human eyes. In the model-eye sample, patterns were obtained in 4 sessions: 2 without realigning between measurements, 1 with realignment, and 1 with an angular shift of 6 degrees from the instrument axis. Three measurements were made in the human eyes with the C-Quant straylight meter (Oculus) to obtain psychometric and objective measures of retinal straylight. Analysis of variance, intraclass correlation coefficients, coefficient of repeatability (CoR), and correlations were used to determine intrasession and intersession repeatability and the relationship between measures. RESULTS: No significant differences were found between the sessions in the model eye (P = .234). The mean CoR was less than 10% in all model- and human-eye sessions. After incomplete patterns were removed, good correlation was achieved between psychometric and objective scatter measurements despite the small sample size (n = 6; r = -0.831; P = .040). CONCLUSIONS: The methodology was repeatable in model and human eyes, strong against realignment and misalignment, and sensitive. Clinical application would benefit from effective use of the sensor's dynamic range.
引用
收藏
页码:1089 / 1095
页数:7
相关论文
共 14 条
[1]   Aberro-polariscope for the human eye [J].
Bueno, JM ;
Berrio, E ;
Artal, P .
OPTICS LETTERS, 2003, 28 (14) :1209-1211
[2]  
CERVINO A, 2008, BR J OPHTHA IN PRESS
[3]   Retinal straylight in patients with monofocal and multifocal intraocular lenses [J].
Cervino, Alejandro ;
Hosking, Sarah L. ;
Montes-Mico, Robert ;
Alio, Jorge L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (03) :441-446
[4]   Operator-induced errors in Hartmann-Shack wavefront sensing: Model eye study [J].
Cervino, Alejandro ;
Hosking, Sarah L. ;
Dunne, Mark C. M. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (01) :115-121
[5]   Wavefront analyzers induce instrument myopia [J].
Cervino, Alejandro ;
Hosking, Sarah L. ;
Rai, Gurjeet K. ;
Naroo, Shezhad A. ;
Gilmartin, Bernard .
JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (08) :795-803
[6]   Validation of a clinical Shack-Hartmann aberrometer [J].
Cheng, X ;
Himebaugh, NL ;
Kollbaum, PS ;
Thibos, LN ;
Bradley, A .
OPTOMETRY AND VISION SCIENCE, 2003, 80 (08) :587-595
[7]  
Donnelly W., 2004, J REFRACT SURG, V20, pS515
[8]  
Donnelly WJ, 2005, J REFRACT SURG, V21, pS547
[9]   Compensation comparison method for assessment of retinal straylight [J].
Franssen, L ;
Coppens, JE ;
van den Berg, TJTP .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2006, 47 (02) :768-776
[10]   Light scattering and optical aberrations as objective parameters to predict visual deterioration in eyes with cataracts [J].
Fujikado, T ;
Kuroda, T ;
Maeda, N ;
Ninomiya, S ;
Goto, H ;
Tano, Y ;
Oshika, T ;
Hirohara, Y ;
Mihashi, T .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (06) :1198-1208