Commissioning of a 4D MRI phantom for use in MR-guided radiotherapy

被引:6
作者
Schneider, Sergej [1 ,2 ]
Dolde, Kai [3 ,4 ,5 ]
Engler, Johanna [2 ]
Hoffmann, Aswin [1 ,2 ,6 ]
Pfaffenberger, Asja [3 ,4 ]
机构
[1] Helmholtz Zentrum Dresden Rossendorf, Inst Radiooncol, OncoRay, Dresden, Germany
[2] Tech Univ Dresden, Helmholtz Zentrum Dresden Rossendorf, Univ Hosp Carl Gustav Carus, OncoRay,Natl Ctr Radiat Res Oncol,Fac Med, Dresden, Germany
[3] German Canc Res Ctr, Med Phys Radiotherapy, Heidelberg, Germany
[4] Heidelberg Inst Radiooncol, Natl Ctr Radiat Res Oncol, Heidelberg, Germany
[5] Heidelberg Univ, Dept Phys & Astron, Heidelberg, Germany
[6] Tech Univ Dresden, Univ Hosp Carl Gustav Carus, Fac Med, Dept Radiotherapy & Radiat Oncol, Dresden, Germany
关键词
commissioning; dynamic phantom; image-guided radiation therapy; MRI; RESPIRATORY MOTION; GEOMETRIC DISTORTION; VALIDATION; MANAGEMENT; 4D-MRI;
D O I
10.1002/mp.13261
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Systems for integrated magnetic resonance guided radiation therapy (MRgRT) provide real-time and online MRI guidance for unequaled targeting performance of moving tumors and organs at risk. The clinical introduction of such systems requires dedicated methods for commissioning and routine machine quality assurance (QA). The aim of the study was to develop a commissioning protocol and method for automatic quantification of target motion and geometric accuracy using a 4D MRI motion phantom. Materials and methods The commissioning was performed on a clinically used 3 T MR scanner. The phantom was positioned on a flat tabletop overlay using an in-house constructed base plate for a quick and reproducible setup. The torso-shaped phantom body, which was filled with mineral oil as signal generating medium, included a 3D grid structure for image distortion analysis and a cylindrical thru-hole in which a software-controlled moving rod with a hypo-intense background gel and a decentralized hyper-intense target simulated 3D organ motion patterns. To allow for sequence optimization, MR relaxometry was performed to determine the longitudinal T-1 and transverse T-2 relaxation times of both target and background gel in the movable cylinder. The geometric image distortion was determined as the mean and maximum 3D Euclidean distance (Delta(mean), Delta(max)) of grid points determined by nonrigid registration of a 3D spoiled gradient echo MRI scan and a CT scan. Sinusoidal 1D/2D/3D motion trajectories, varying in amplitude and frequency, as well as an exemplary 1D MR navigator diaphragm motion pattern extracted from a healthy volunteer scan, were scanned by means of 2D cine MRI and 4D MRI. Target positions were automatically extracted from 2D cine MRI using an in-house developed software tool. Results The base plate enabled a reproducible setup with a deviation of <1 mm in all directions. Relaxometry yielded T-1/T-2 values for target and background gel of 208.1 +/- 2.8/30.5 +/- 4.7 ms and 871 +/- 36/13.4 +/- 1.3 ms, respectively. The 3D geometric image distortion increased with distance from the magnetic isocenter, with Delta(mean) = 0.58 +/- 0.30 mm and Delta(max) = 1.31 mm. The frequencies of the reconstructed motion patterns agreed with the preset values within 0.5%, whereas the reconstructed amplitudes showed a maximum deviation to the preset amplitudes of Conclusion A method and protocol for commissioning of a 4D MRI motion phantom on a 3 T MR scanner for MRgRT was developed. High-contrast and geometrically reliable 2D cine MR images of the phantom's moving target could be obtained. The preset motion parameters could be extracted with sufficient spatio-temporal accuracy from 2D cine MRI in all motion directions. The overall 3D geometric image distortion of
引用
收藏
页码:25 / 33
页数:9
相关论文
共 50 条
  • [21] Offline and online LSTM networks for respiratory motion prediction in MR-guided radiotherapy
    Lombardo, Elia
    Rabe, Moritz
    Xiong, Yuqing
    Nierer, Lukas
    Cusumano, Davide
    Placidi, Lorenzo
    Boldrini, Luca
    Corradini, Stefanie
    Niyazi, Maximilian
    Belka, Claus
    Riboldi, Marco
    Kurz, Christopher
    Landry, Guillaume
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (09)
  • [22] Online MR-guided proton and ion beam radiotherapy: investigation of image quality
    Paul, K.
    Dorsch, S.
    Elter, A.
    Beyer, C.
    Naumann, J.
    Hansmann, T.
    Feldmeier, E.
    Haberer, T.
    Karger, C. P.
    Debus, J.
    Klueter, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (18)
  • [23] Commissioning of MRI-guided gynaecological brachytherapy using an MR-linac
    Rojas-Lopez, Jose Alejandro
    Cabrera-Santiago, Alexis
    Adragna, Celeste
    Ibarra-Ortega, Brenda Elisabeth
    Lopez-Luna, Jose Eleazar
    Contreras-Rodriguez, Jonathan Agustin
    Martinez-Ortiz, Efrain
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (05):
  • [24] Gaussian Processes for real-time 3D motion and uncertainty estimation during MR-guided radiotherapy
    Huttinga, Niek R. F.
    Bruijnen, Tom
    van den Berg, Cornelis A. T.
    Sbrizzi, Alessandro
    MEDICAL IMAGE ANALYSIS, 2023, 88
  • [25] Proton vs. photon radiotherapy for MR-guided dose escalation of intraprostatic lesions
    Moteabbed, Maryam
    Harisinghani, Mukesh
    Paganetti, Harald
    Trofimov, Alexei
    Lu, Hsiao-Ming
    Efstathiou, Jason A.
    ACTA ONCOLOGICA, 2021, 60 (10) : 1283 - 1290
  • [26] Clinical assessment of geometric distortion for a 0.35T MR-guided radiotherapy system
    Neylon, John
    Cook, Kiri A.
    Yang, Yingli
    Du, Dongsu
    Sheng, Ke
    Chin, Robert K.
    Kishan, Amar U.
    Lamb, James M.
    Low, Daniel A.
    Cao, Minsong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (08): : 303 - 309
  • [27] Analysis of Magnetic Resonance Image Signal Fluctuations Acquired During MR-Guided Radiotherapy
    Breto, Adrian L.
    Padgett, Kyle R.
    Ford, John C.
    Kwon, Deukwoo
    Chang, Channing
    Fuss, Martin
    Stoyanova, Radka
    Mellon, Eric A.
    CUREUS, 2018, 10 (03):
  • [28] Prospective feasibility analysis of a novel off-line approach for MR-guided radiotherapy
    Bostel, Tilman
    Pfaffenberger, Asja
    Delorme, Stefan
    Dreher, Constantin
    Echner, Gernot
    Haering, Peter
    Lang, Clemens
    Splinter, Mona
    Laun, Frederik
    Mueller, Marco
    Jaekel, Oliver
    Debus, Juergen
    Huber, Peter E.
    Sterzing, Florian
    Nicolay, Nils H.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2018, 194 (05) : 425 - 434
  • [29] Image-driven, model-based 3D abdominal motion estimation for MR-guided radiotherapy
    Stemkens, Bjorn
    Tijssen, Rob H. N.
    de Senneville, Baudouin Denis
    Lagendijk, Jan J. W.
    van den Berg, Cornelis A. T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (14) : 5335 - 5355
  • [30] Feasibility of delivered dose reconstruction for MR-guided SBRT of pancreatic tumors with fast, real-time 3D cine MRI
    Grimbergen, Guus
    Pötgens, Giulia G.
    Eijkelenkamp, Hidde
    Raaymakers, Bas W.
    Intven, Martijn P. W.
    Meijer, Gert J.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182