Commissioning of a 4D MRI phantom for use in MR-guided radiotherapy

被引:6
|
作者
Schneider, Sergej [1 ,2 ]
Dolde, Kai [3 ,4 ,5 ]
Engler, Johanna [2 ]
Hoffmann, Aswin [1 ,2 ,6 ]
Pfaffenberger, Asja [3 ,4 ]
机构
[1] Helmholtz Zentrum Dresden Rossendorf, Inst Radiooncol, OncoRay, Dresden, Germany
[2] Tech Univ Dresden, Helmholtz Zentrum Dresden Rossendorf, Univ Hosp Carl Gustav Carus, OncoRay,Natl Ctr Radiat Res Oncol,Fac Med, Dresden, Germany
[3] German Canc Res Ctr, Med Phys Radiotherapy, Heidelberg, Germany
[4] Heidelberg Inst Radiooncol, Natl Ctr Radiat Res Oncol, Heidelberg, Germany
[5] Heidelberg Univ, Dept Phys & Astron, Heidelberg, Germany
[6] Tech Univ Dresden, Univ Hosp Carl Gustav Carus, Fac Med, Dept Radiotherapy & Radiat Oncol, Dresden, Germany
关键词
commissioning; dynamic phantom; image-guided radiation therapy; MRI; RESPIRATORY MOTION; GEOMETRIC DISTORTION; VALIDATION; MANAGEMENT; 4D-MRI;
D O I
10.1002/mp.13261
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Systems for integrated magnetic resonance guided radiation therapy (MRgRT) provide real-time and online MRI guidance for unequaled targeting performance of moving tumors and organs at risk. The clinical introduction of such systems requires dedicated methods for commissioning and routine machine quality assurance (QA). The aim of the study was to develop a commissioning protocol and method for automatic quantification of target motion and geometric accuracy using a 4D MRI motion phantom. Materials and methods The commissioning was performed on a clinically used 3 T MR scanner. The phantom was positioned on a flat tabletop overlay using an in-house constructed base plate for a quick and reproducible setup. The torso-shaped phantom body, which was filled with mineral oil as signal generating medium, included a 3D grid structure for image distortion analysis and a cylindrical thru-hole in which a software-controlled moving rod with a hypo-intense background gel and a decentralized hyper-intense target simulated 3D organ motion patterns. To allow for sequence optimization, MR relaxometry was performed to determine the longitudinal T-1 and transverse T-2 relaxation times of both target and background gel in the movable cylinder. The geometric image distortion was determined as the mean and maximum 3D Euclidean distance (Delta(mean), Delta(max)) of grid points determined by nonrigid registration of a 3D spoiled gradient echo MRI scan and a CT scan. Sinusoidal 1D/2D/3D motion trajectories, varying in amplitude and frequency, as well as an exemplary 1D MR navigator diaphragm motion pattern extracted from a healthy volunteer scan, were scanned by means of 2D cine MRI and 4D MRI. Target positions were automatically extracted from 2D cine MRI using an in-house developed software tool. Results The base plate enabled a reproducible setup with a deviation of <1 mm in all directions. Relaxometry yielded T-1/T-2 values for target and background gel of 208.1 +/- 2.8/30.5 +/- 4.7 ms and 871 +/- 36/13.4 +/- 1.3 ms, respectively. The 3D geometric image distortion increased with distance from the magnetic isocenter, with Delta(mean) = 0.58 +/- 0.30 mm and Delta(max) = 1.31 mm. The frequencies of the reconstructed motion patterns agreed with the preset values within 0.5%, whereas the reconstructed amplitudes showed a maximum deviation to the preset amplitudes of Conclusion A method and protocol for commissioning of a 4D MRI motion phantom on a 3 T MR scanner for MRgRT was developed. High-contrast and geometrically reliable 2D cine MR images of the phantom's moving target could be obtained. The preset motion parameters could be extracted with sufficient spatio-temporal accuracy from 2D cine MRI in all motion directions. The overall 3D geometric image distortion of
引用
收藏
页码:25 / 33
页数:9
相关论文
共 50 条
  • [1] Predicting 4D liver MRI for MR-guided interventions
    Gulamhussene, Gino
    Meyer, Anneke
    Rak, Marko
    Bashkanov, Oleksii
    Omari, Jazan
    Pech, Maciej
    Hansen, Christian
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2022, 101
  • [2] Investigation of the 4D composite MR image distortion field associated with tumor motion for MR-guided radiotherapy
    Stanescu, T.
    Jaffray, D.
    MEDICAL PHYSICS, 2016, 43 (03) : 1550 - 1562
  • [3] Review of MR-Guided Radiotherapy for Esophageal Cancer
    Lee, Sangjune Laurence
    Bassetti, Michael
    Meijer, Gert J.
    Mook, Stella
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [4] MR-Guided Radiotherapy for Liver Malignancies
    Boldrini, Luca
    Corradini, Stefanie
    Gani, Cihan
    Henke, Lauren
    Hosni, Ali
    Romano, Angela
    Dawson, Laura
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [5] A pilot study of respiratory motion characterization in the abdomen using a fast volumetric 4D-MRI for MR-guided radiotherapy
    Wong, Oi Lei
    Law, Max Wai Kong
    Poon, Darren Ming Chun
    Yung, Raymond Wai Hung
    Yu, Siu ki
    Cheung, Kin yin
    Yuan, Jing
    PRECISION RADIATION ONCOLOGY, 2022, 6 (02): : 100 - 109
  • [6] Editorial: Online Adaptive MR-Guided Radiotherapy
    Kerkmeijer, Linda G. W.
    Valentini, Vincenzo
    Fuller, Clifton D.
    Slotman, Ben J. J.
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [7] MR-Guided Adaptive Radiotherapy for Bladder Cancer
    Hijab, Adham
    Tocco, Boris
    Hanson, Ian
    Meijer, Hanneke
    Nyborg, Christina Junker
    Bertelsen, Anders Smedegaard
    Smeenk, Robert Jan
    Smith, Gillian
    Michalski, Jeff
    Baumann, Brian C.
    Hafeez, Shaista
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [8] MR-Guided Radiotherapy for Brain and Spine Tumors
    Maziero, Danilo
    Straza, Michael W.
    Ford, John C.
    Bovi, Joseph A.
    Diwanji, Tejan
    Stoyanova, Radka
    Paulson, Eric S.
    Mellon, Eric A.
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [9] Synthetic 4DCT(MRI) lung phantom generation for 4D radiotherapy and image guidance investigations
    Duetschler, Alisha
    Bauman, Grzegorz
    Bieri, Oliver
    Cattin, Philippe C.
    Ehrbar, Stefanie
    Engin-Deniz, Georg
    Giger, Alina
    Josipovic, Mirjana
    Jud, Christoph
    Krieger, Miriam
    Nguyen, Damien
    Persson, Gitte F.
    Salomir, Rares
    Weber, Damien C.
    Lomax, Antony J.
    Zhang, Ye
    MEDICAL PHYSICS, 2022, 49 (05) : 2890 - 2903
  • [10] A gel phantom for the calibration of MR-guided ultrasound thermal therapy
    Lochhead, S
    Bradwell, D
    Chopra, R
    Bronskill, MJ
    2004 IEEE Ultrasonics Symposium, Vols 1-3, 2004, : 1481 - 1483