Lightweight, compact, and high-performance 3T MR system for imaging the brain and extremities

被引:69
|
作者
Foo, Thomas K. F. [1 ]
Laskaris, Evangelos [1 ]
Vermilyea, Mark [1 ]
Xu, Minfeng [1 ]
Thompson, Paul [1 ]
Conte, Gene [1 ]
Van Epps, Christopher [1 ]
Immer, Christopher [1 ]
Lee, Seung-Kyun [1 ,2 ]
Tan, Ek T. [1 ]
Graziani, Dominic [1 ]
Mathieu, Jean-Baptise [3 ]
Hardy, Christopher J. [1 ]
Schenck, John F. [1 ]
Fiveland, Eric [1 ]
Stautner, Wolfgang [1 ]
Ricci, Justin [1 ]
Piel, Joseph [1 ]
Park, Keith [1 ]
Hua, Yihe [1 ]
Bai, Ye [1 ]
Kagan, Alex [1 ]
Stanley, David [5 ]
Weavers, Paul T. [4 ,5 ]
Gray, Erin [4 ]
Shu, Yunhong [4 ]
Frick, Matthew A. [4 ]
Campeau, Norbert G. [4 ]
Trzasko, Joshua [4 ]
Huston, John, III [4 ]
Bernstein, Matt A. [4 ]
机构
[1] GE Global Res, 1 Res Circle, Niskayuna, NY 12309 USA
[2] Sungkyunkwan Univ, Dept Biomed Engn, Suwon, South Korea
[3] GE Healthcare, Florence, SC USA
[4] Mayo Clin, Dept Radiol, Rochester, MN USA
[5] GE Healthcare, Waukesha, WI USA
关键词
PERIPHERAL-NERVE STIMULATION; TRANSVERSE GRADIENT COILS; CONCOMITANT FIELDS; HEAD; DISTORTION; DIFFUSION; DESIGN; LIMITS; FMRI; BODY;
D O I
10.1002/mrm.27175
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To build and evaluate a small-footprint, lightweight, high-performance 3T MRI scanner for advanced brain imaging with image quality that is equal to or better than conventional whole-body clinical 3T MRI scanners, while achieving substantial reductions in installation costs. Methods: A conduction-cooled magnet was developed that uses less than 12 liters of liquid helium in a gas-charged sealed system, and standard NbTi wire, and weighs approximately 2000 kg. A 42-cm inner-diameter gradient coil with asymmetric transverse axes was developed to provide patient access for head and extremity exams, while minimizing magnet-gradient interactions that adversely affect image quality. The gradient coil was designed to achieve simultaneous operation of 80-mT/m peak gradient amplitude at a slew rate of 700 T/m/s on each gradient axis using readily available 1-MVA gradient drivers. Results: In a comparison of anatomical imaging in 16 patients using T2-weighted 3D fluid-attenuated inversion recovery (FLAIR) between the compact 3T and whole-body 3T, image quality was assessed as equivalent to or better across several metrics. The ability to fully use a high slew rate of 700 T/m/s simultaneously with 80-mT/m maximum gradient amplitude resulted in improvements in image quality across EPI, DWI, and anatomical imaging of the brain. Conclusions: The compact 3T MRI system has been in continuous operation at the Mayo Clinic since March 2016. To date, over 200 patient studies have been completed, including 96 comparison studies with a clinical 3T whole-body MRI. The increased gradient performance has reliably resulted in consistently improved image quality.
引用
收藏
页码:2232 / 2245
页数:14
相关论文
共 50 条
  • [1] Reducing PNS with minimal performance penalties via simple pulse sequence modifications on a high-performance compact 3T scanner
    In, Myung-Ho
    Shu, Yunhong
    Trzasko, Joshua D.
    Yarach, Uten
    Kang, Daehun
    Gray, Erin M.
    Huston, John
    Bernstein, Matt A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (15)
  • [2] Simultaneous Head and Spine MR Imaging in Children Using a Dedicated Multichannel Receiver System at 3T
    Li, Nan
    Zheng, Hui
    Xu, Guangzu
    Gui, Ting
    Yin, Qiufeng
    Chen, Qiaoyan
    Lee, Jo
    Xin, Yang
    Zhang, Shuheng
    He, Qiang
    Zhang, Xiaoliang
    Liu, Xin
    Zheng, Hairong
    Wang, Dengbin
    Li, Ye
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2021, 68 (12) : 3659 - 3670
  • [3] Improved Resting-State Functional MRI Using Multi-Echo Echo-Planar Imaging on a Compact 3T MRI Scanner with High-Performance Gradients
    Kang, Daehun
    In, Myung-Ho
    Jo, Hang Joon
    Halverson, Maria A.
    Meyer, Nolan K.
    Ahmed, Zaki
    Gray, Erin M.
    Madhavan, Radhika
    Foo, Thomas K.
    Fernandez, Brice
    Black, David F.
    Welker, Kirk M.
    Trzasko, Joshua D.
    Huston III, John
    Bernstein, Matt A.
    Shu, Yunhong
    SENSORS, 2023, 23 (09)
  • [4] Full-brain coverage and high-resolution Imaging capabilities of passband b-SSFP fMRI at 3T
    Lee, Jin Hyung
    Dumoulin, Serge O.
    Saritas, Emine U.
    Glover, Gary H.
    Wandell, Brian A.
    Nishimura, Dwight G.
    Pauly, John M.
    MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (05) : 1099 - 1110
  • [5] Functional MR angiography using phase contrast imaging technique at 3T MRI
    Kang, Chang-Ki
    Kim, Sang-Hoon
    Lee, Hyon
    Park, Chan-A
    Kim, Young-Bo
    Cho, Zang-Hee
    NEUROIMAGE, 2010, 50 (03) : 1036 - 1043
  • [6] Atlas-Based Adaptive Hadamard-Encoded MR Spectroscopic Imaging at 3T
    Liu, Huawei
    Autry, Adam W.
    Larson, Peder E. Z.
    Xu, Duan
    Li, Yan
    TOMOGRAPHY, 2023, 9 (05) : 1592 - 1602
  • [7] High-resolution microscopic diffusion anisotropy imaging in the human hippocampus at 3T
    Yoo, Jiyoon
    Kerkela, Leevi
    Hales, Patrick W.
    Seunarine, Kiran K.
    Clark, Christopher A.
    MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (04) : 1903 - 1913
  • [8] What is the optimal b value in diffusion-weighted MR imaging to depict prostate cancer at 3T?
    T. Metens
    D. Miranda
    J. Absil
    C. Matos
    European Radiology, 2012, 22 : 703 - 709
  • [9] What is the optimal b value in diffusion-weighted MR imaging to depict prostate cancer at 3T?
    Metens, T.
    Miranda, D.
    Absil, J.
    Matos, C.
    EUROPEAN RADIOLOGY, 2012, 22 (03) : 703 - 709
  • [10] Flexible, 31-Channel Breast Coil for Enhanced Parallel Imaging Performance at 3T
    Hancu, Ileana
    Fiveland, Eric
    Park, Keith
    Giaquinto, Randy O.
    Rohling, Kenneth
    Wiesinger, Florian
    MAGNETIC RESONANCE IN MEDICINE, 2016, 75 (02) : 897 - 905