Off-axis x-ray spectra: A comparison of Monte Carlo simulated and computed x-ray spectra with measured spectra

被引:41
作者
Bhat, M
Pattison, J
Bibbo, G
Caon, M
机构
[1] Univ S Australia, Sch Phys & Elect Syst Engn, Mawsan Lakes, SA 5095, Australia
[2] Womens & Childrens Hosp, Div Med Imaging, N Adelaide, SA 5006, Australia
[3] Flinders Univ S Australia, Sch Nursing, Bedford Pk, SA 5042, Australia
关键词
off-axis x-ray spectra; Monte Carlo; EGS4; heel effect;
D O I
10.1118/1.598517
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The off-axis x-ray spectra from a constant potential x-ray generator were measured with a high purity germanium spectrometer cooled to liquid nitrogen temperature. The measured spectra were compared with off-axis x-ray spectra calculated using a code based on the semiempirical model developed by Tucker et nl, and Monte Carlo simulated x-ray spectra using the EGS4 code system. In this study, both the Tucker model, and the EGS4 code system, were found to produce off-axis bremsstrahlung x-ray spectra which agreed well with the spectra measured at three emerging angles. In the measured and the EGS4 generated spectra the total K-characteristic peaks were in increasing order, as observed in the anode to cathode direction, whereas the Tucker model produced maximum total K-characteristic peaks at the 6 degrees anode side, and lesser amounts at the central axis and the 6 degrees cathode side. Large differences in the total K-characteristic lines is seen among the three different methods. The EGS4 code system was able to produce x-ray spectra for a combination of target materials. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)02102-1].
引用
收藏
页码:303 / 309
页数:7
相关论文
共 30 条
[1]   MONTE-CARLO TECHNIQUES IN MEDICAL RADIATION PHYSICS [J].
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :861-920
[2]   MONTE-CARLO STUDIES OF X-RAY-SCATTERING IN TRANSMISSION DIAGNOSTIC-RADIOLOGY [J].
BARNEA, G ;
DICK, CE .
MEDICAL PHYSICS, 1986, 13 (04) :490-495
[3]   Diagnostic x-ray spectra: A comparison of spectra generated by different computational methods with a measured spectrum [J].
Bhat, M ;
Pattison, J ;
Bibbo, G ;
Caon, M .
MEDICAL PHYSICS, 1998, 25 (01) :114-120
[4]  
BIELAJEW AF, 1989, PIRS0203 COUNC CAN D
[5]  
BIELAJEW AF, 1986, PIRS0052 NAT RES COU
[6]   COMPUTATION OF BREMSSTRAHLUNG X-RAY-SPECTRA AND COMPARISON WITH SPECTRA MEASURED WITH A GE(LI) DETECTOR [J].
BIRCH, R ;
MARSHALL, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) :505-517
[7]  
BIRCH R, 1979, UNPUB SCI REPORT SER, V30
[8]   THE 3 PARAMETER EQUIVALENT SPECTRA AS AN INDEX OF BEAM QUALITY [J].
BOONE, JM .
MEDICAL PHYSICS, 1988, 15 (03) :304-310
[9]   The effect on dose to computed tomography phantoms of varying the theoretical x-ray spectrum: A comparison of four diagnostic x-ray spectrum calculating codes [J].
Caon, M ;
Bibbo, G ;
Pattison, J ;
Bhat, M .
MEDICAL PHYSICS, 1998, 25 (06) :1021-1027
[10]  
Caon M., 1996, Australasian Physical and Engineering Sciences in Medicine, V19, P201