Isobaric Spike Method for Absolute Isotopic Ratio Determination by MC-ICP-MS

被引:5
|
作者
Shuai, Kang [1 ,2 ]
Li, Weiqiang [1 ,2 ,3 ]
Hui, Hejiu [1 ,2 ,3 ]
机构
[1] Nanjing Univ, Sch Earth Sci & Engn, State Key Lab Mineral Deposits Res, Nanjing 210023, Peoples R China
[2] Nanjing Univ, Sch Earth Sci & Engn, Lunar & Planetary Sci, Nanjing 210023, Peoples R China
[3] CAS Ctr Excellence Comparat Planetol, Hefei 230026, Peoples R China
关键词
PLASMA-MASS SPECTROMETRY; FRACTIONATION; CALCIUM; PRECISE; ANOMALIES; SAMPLE; EARTH; ZN; CU;
D O I
10.1021/acs.analchem.9b04160
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Absolute isotopic ratios are required for isobaric interference corrections, spike calibrations, and isotopic analysis by external normalization methods. However, high-precision natural isotopic abundance data are lacking for many elements, particularly those with less than four isotopes or having isobaric isotopes with other elements. In this study, we developed a method for absolute isotope ratio analysis, which integrates the concept of the double-spike method with isotopic analysis of element pairs that have isobaric isotopes. Using multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS), the isotopic composition of a sample can be derived by measuring a series of mixtures of the sample and a spike element that has an isobaric isotope with the element being analyzed. We applied this method to five pairs of elements (Ca-Ti, V-Ti, Cr-Ti, Ni-Zn, and In-Sn) and obtained the absolute isotopic ratios for Ca, V, Cr, Ni, and In, as well as the relative Ca isotopic composition. By simultaneous measurement of Ti and Ca isotopes, a quantitative relationship between the instrumental mass fractionation factors and element masses was developed. After correcting for the difference in instrumental mass fractionation factors, the obtained absolute ratios agree well with the literature data and have per mil level accuracy. This method has considerable potential in measuring the absolute isotopic ratios of elements that have isobaric isotope with other elements. Such precisely determined absolute isotopic ratios and the relationship between the instrumental mass fractionation factors and elemental masses will improve isobaric interference corrections, particularly when chemical purification is imperfect or during laser ablation analysis.
引用
收藏
页码:4820 / 4828
页数:9
相关论文
共 50 条
  • [1] Determination of the Isotopic Composition of an Enriched Hafnium Spike by MC-ICP-MS Using a Regression Model
    Lin, Ran
    Lin, Jie
    Zong, Keqing
    Chen, Kang
    Tong, Shuoyun
    Feng, Lanping
    Zhang, Wen
    Li, Ming
    Liu, Yongsheng
    Hu, Zhaochu
    Zhou, Lian
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2020, 44 (04) : 753 - 762
  • [2] Determination of the isotopic composition of tungsten using MC-ICP-MS
    Zhang, Rouxi
    Meija, Juris
    Huang, Yi
    Pei, Xiangjun
    Mester, Zoltan
    Yang, Lu
    ANALYTICA CHIMICA ACTA, 2019, 1089 : 19 - 24
  • [3] Precise Determination of Stable Strontium Isotopic Compositions by MC-ICP-MS
    Xu, Juan
    Yang, Shouye
    Yang, Yueheng
    Liu, Yongsheng
    Xie, Xiaolei
    ATOMIC SPECTROSCOPY, 2020, 41 (02) : 64 - 73
  • [4] Copper Isotopic Analysis in Geological and Biological Reference Materials by MC-ICP-MS
    Sullivan, Kaj
    Layton-Matthews, Daniel
    Leybourne, Matthew
    Kidder, James
    Mester, Zoltan
    Yang, Lu
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2020, 44 (02) : 349 - 362
  • [5] In situ Ti isotopic analysis by femtosecond laser ablation MC-ICP-MS
    Huang, Chao
    Wang, Hao
    Xie, Liewen
    Nie, Nicole X.
    Yang, Yueheng
    Zhao, Xinmiao
    Li, Jin
    Tian, Heng-Ci
    Wu, Shitou
    Xu, Lei
    Yang, Jinhui
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (10) : 2165 - 2175
  • [6] A single-stage purification method for the precise determination of zirconium isotopic composition in geological samples by double spike MC-ICP-MS
    Feng, Lanping
    Wang, Chen
    Han, Yan
    Zhang, Wen
    Guo, Jing-Liang
    Zhou, Lian
    Hu, Zhaochu
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (12) : 3035 - 3047
  • [7] A Rapid and Simple Method for Lithium Purification and Isotopic Analysis of Geological Reference Materials by MC-ICP-MS
    Zhu, Guanhong
    Ma, Jinlong
    Wei, Gangjian
    Zhang, Le
    FRONTIERS IN CHEMISTRY, 2020, 8
  • [8] A Highly Efficient Method for the Accurate and Precise Determination of Zinc Isotopic Ratios in Zinc-rich Minerals Using MC-ICP-MS
    Nie, Xiaojuan
    Bao, Zhian
    Liang, Peng
    Chen, Kaiyun
    Zong, Chunlei
    Yuan, Honglin
    ATOMIC SPECTROSCOPY, 2022, 43 (02) : 117 - 125
  • [9] A method for determining the isotopic composition of dissolved zinc in seawater by MC-ICP-MS with a 67Zn-68Zn double spike
    Samanta, Moneesha
    Ellwood, Michael J.
    Mortimer, Graham E.
    MICROCHEMICAL JOURNAL, 2016, 126 : 530 - 537
  • [10] Accurate and precise determination of stable Cr isotope compositions in carbonates by double spike MC-ICP-MS
    Bonnand, Pierre
    Parkinson, Ian J.
    James, Rachael H.
    Karjalainen, Anne-Mari
    Fehr, Manuela A.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2011, 26 (03) : 528 - 535