Design Guidelines for a Double Scattering Compton Camera for Prompt-γ Imaging During Ion Beam Therapy: A Monte Carlo Simulation Study

被引:76
作者
Richard, M. -H. [1 ,2 ,3 ,7 ]
Chevallier, M. [1 ,2 ,7 ]
Dauvergne, D. [1 ,2 ,7 ]
Freud, N. [3 ]
Henriquet, P. [1 ,2 ,7 ]
Le Foulher, F. [1 ,2 ,7 ]
Letang, J. M. [3 ]
Montarou, G. [4 ,5 ]
Ray, C. [1 ,2 ,7 ]
Roellinghoff, F. [1 ,2 ,7 ]
Testa, E. [1 ,2 ,7 ]
Testa, M. [1 ,2 ,7 ]
Walenta, A. H. [6 ]
机构
[1] Univ Lyon, F-69622 Lyon, France
[2] Univ Lyon 1, F-69622 Villeurbanne, France
[3] INSA Lyon Lab Nondestruct Testing Using Ionizing, F-69621 Villeurbanne, France
[4] Clermont Univ, F-63177 Aubiere, France
[5] CNRS, IN2P3, LPC Clermont, UMR 6533, F-63177 Aubiere, France
[6] Uni Siegen, FB Phys, D-57068 Siegen, Germany
[7] CNRS, IN2P3, IPNL, UMR 5822, F-69622 Villeurbanne, France
关键词
Compton camera; Geant4; hadrontherapy; ion beam therapy; prompt gamma; X-RAY; PROTON; RESOLUTION;
D O I
10.1109/TNS.2010.2076303
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In hadrontherapy in order to fully take advantage of the assets of the ion irradiation, the position of the Bragg peak has to be monitored accurately. Here, we investigate a monitoring method relying on the detection in real time of the prompt emitted quasi instantaneously during the nuclear fragmentation processes. Our detection system combines a beam hodoscope and a double scattering Compton camera. The prompt-gamma emission points are reconstructed by intersecting the ion trajectories given by the hodoscope and the Compton cones reconstructed with the camera. We propose here to study in terms of point spread function and efficiency the theoretical feasibility of the emission points reconstruction with our set-up in the case of a photon point source in air. First we analyze the nature of all the interactions which are likely to produce an energy deposit in the three detectors of the camera. It is underlined that upper energy thresholds in both scatter detectors are required in order to select mainly Compton events (one Compton interaction in each scatter detector and one interaction in the absorber detector). Then, we study the influence of various parameters such as the photon energy and the inter-detector distances on the Compton camera response. These studies are carried out by means of Geant4 simulations. We use a source with a spectrum corresponding to the prompt-spectrum emitted during the carbon ion irradiation of a water phantom. In the current configuration, the spatial resolution of the Compton camera is about 6 mm (Full Width at Half Maximum) and the detection efficiency 10(-5). Finally, provided the detection efficiency is increased, the clinical applicability of our system is considered.
引用
收藏
页码:87 / 94
页数:8
相关论文
共 22 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]   Validation of the Geant4 electromagnetic photon cross-sections for elements and compounds [J].
Cirrone, G. A. P. ;
Cuttone, G. ;
Di Rosa, F. ;
Pandola, L. ;
Romano, F. ;
Zhang, Q. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 618 (1-3) :315-322
[3]   A SOURCE RECONSTRUCTION METHOD FOR MULTIPLE SCATTER COMPTON CAMERAS [J].
DOGAN, N ;
WEHE, DK ;
AKCASU, AZ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (05) :1427-1430
[4]   Charged hadron tumour therapy monitoring by means of PET [J].
Enghardt, W ;
Crespo, P ;
Fiedler, F ;
Hinz, R ;
Parodi, K ;
Pawelke, J ;
Pönisch, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 525 (1-2) :284-288
[5]   Semiconductor drift detectors for X- and gamma-ray spectroscopy and imaging [J].
Fiorini, C. ;
Longoni, A. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (10) :2173-2181
[6]  
HWIKANG B, 2009, IEEE T NUCL SCI, V56, P46
[7]   Measurements of deposited dose with induced β+ activity in proton and heavy-ion therapy [J].
Inaniwa, T. ;
Kohno, T. ;
Tomitani, T. ;
Urakabe, E. ;
Sato, S. ;
Kanazawa, M. ;
Kanai, T. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (02) :1140-1143
[8]   Treatment planning for heavy-ion radiotherapy:: physical beam model and dose optimization [J].
Krämer, M ;
Jäkel, O ;
Haberer, T ;
Kraft, G ;
Schardt, D ;
Weber, U .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3299-3317
[9]   Novel detector technology for clinical PET [J].
Lecomte, Roger .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 :69-85
[10]  
LEFOULHER F, IEEE T NUCL IN PRESS