Dialysate partitioning in the Genius® batch hemodialysis system:: Effect of temperature and solute concentration

被引:14
作者
Dhondt, A
Eloot, S
De Wachter, D
De Smet, R
Waterloos, MA
Glorieux, G
Lameire, N
Verdonck, P
Vanholder, R
机构
[1] State Univ Ghent Hosp, Div Renal, Dept Internal Med, B-9000 Ghent, Belgium
[2] State Univ Ghent, Inst Biomed Technol, Hydraul Lab, B-9000 Ghent, Belgium
关键词
Genius (R); hemodialysis; urea; uremic solutes; temperature; density;
D O I
10.1111/j.1523-1755.2005.00356.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background. The Genius(R) batch system contains a 75-L closed reservoir from which fresh dialysate is extracted at the top, and to which spent dialysate is returned at the bottom. In vivo studies have demonstrated that almost the entire amount of dialysate can be used before contamination of fresh with spent dialysate occurs. The question is raised whether density differences cause this separation, and what the relative contributions of temperature and solute content are. Methods. As patient substitute, a container filled with dialysate was loaded with various amounts of urea. Temperature differences between spent and fresh dialysate were imposed by not heating the dialysate at the outlet line from the dialyzer (A), heating the outlet to obtain continuously equal temperatures at inlet and outlet (B), or to temperatures as in vivo (C). With a dialysate flow set at 300 mL/min, urea is not expected at the inlet before 250 minutes. Results. With a urea concentration of 33 mg/dL, urea contamination at the dialysate inlet line occurred after 185 +/- 20 (A), 122 +/- 11 (B), and 175 +/- 12 minutes (C) of dialysis, whereas with 67 mg/dL, this happened at 219 +/- 5 (A), 162 +/- 11 (B), and 202 +/- 8 minutes (C). With 100 and 150 mg/dL, urea contamination appeared at 224 +/- 2 (A) and 204 +/- 14 minutes (B), and 227 +/- 5 (A) and 232 +/- 13 minutes (B), respectively. Conclusion. Both temperature differences between spent and fresh dialysate and solute content of spent dialysate contribute to dialysate partitioning in the Genius' dialysis system.
引用
收藏
页码:2470 / 2476
页数:7
相关论文
共 13 条
  • [1] Citrate anticoagulation does not correct cuprophane bioincompatibility as evaluated by the expression of leukocyte surface molecules
    Dhondt, A
    Vanholder, R
    Waterloos, MA
    Glorieux, G
    De Smet, R
    Lameire, N
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 1998, 13 (07) : 1752 - 1758
  • [2] Studies on dialysate mixing in the Genius® single-pass batch system for hemodialysis therapy
    Dhondt, AW
    Vanholder, RC
    De Smet, RV
    Claus, SA
    Waterloos, MA
    Glorieux, GL
    Delanghe, JR
    Lameire, NH
    [J]. KIDNEY INTERNATIONAL, 2003, 63 (04) : 1540 - 1547
  • [3] Regional citrate anticoagulation for hemodialysis using a conventional calcium-containing dialysate
    Evenepoel, P
    Maes, B
    Vanwalleghem, J
    [J]. AMERICAN JOURNAL OF KIDNEY DISEASES, 2002, 39 (02) : 315 - 323
  • [4] Renaissance of the batch method?
    Fassbinder, W
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 1998, 13 (12) : 3010 - 3012
  • [5] One for all - A multi-use dialysis system for effective treatment of severe thallium intoxication
    Kielstein, JT
    Linnenweber, S
    Schoepke, T
    Fliser, D
    [J]. KIDNEY & BLOOD PRESSURE RESEARCH, 2004, 27 (03) : 197 - 199
  • [6] Efficacy and cardiovascular tolerability of extended dialysis in critically ill patients: A randomized controlled study
    Kielstein, JT
    Kretschmer, U
    Ernst, T
    Hafer, C
    Bahr, MJ
    Haller, H
    Fliser, D
    [J]. AMERICAN JOURNAL OF KIDNEY DISEASES, 2004, 43 (02) : 342 - 349
  • [7] Efficiency of high-flux hemodialysis in the treatment of valproic acid intoxication
    Kielstein, JT
    Woywodt, A
    Schumann, G
    Haller, H
    Fliser, D
    [J]. JOURNAL OF TOXICOLOGY-CLINICAL TOXICOLOGY, 2003, 41 (06): : 873 - 876
  • [8] Kielstein JT, 2002, CLIN NEPHROL, V57, P484
  • [9] A simplified method for adequate hemodialysis
    Kleophas, W
    Backus, G
    [J]. BLOOD PURIFICATION, 2001, 19 (02) : 189 - 194
  • [10] LENER C, 1981, PHYS CONSTANTS GEIGY, V1, P50