Set-up improvement in head and neck radiotherapy using a 3D off-line EPID-based correction protocol and a customised head and neck support

被引:62
|
作者
van Lin, ENJT [1 ]
van der Vight, L [1 ]
Huizenga, H [1 ]
Kaanders, JHAM [1 ]
Visser, AG [1 ]
机构
[1] Univ Nijmegen, Med Ctr, Dept Radiat Oncol, NL-6500 HB Nijmegen, Netherlands
关键词
portal imaging; immobilisation; set-up correction; head and neck; radiotherapy;
D O I
10.1016/S0167-8140(03)00134-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: First, to investigate the set-up improvements resulting from the introduction of a customised head and neck (HN) support system in combination with a technologist-driven off-line correction protocol in HN radiotherapy. Second. to define margins for planning target volume definition, accounting for systematic and random set-up uncertainties. Methods and materials: In 63 patients 498 treatment fractions were evaluated to develop and implement a 3D shrinking action level correction protocol. In the comparative Study two different HN-supports were compared: a flexible 'standard HN-support" and a 'customised HN-support'. For all three directions (x, y and z) random and systematic set-up deviations (1 S.D.) were measured. Results: The customised HN-support improves the patient positioning compared to the standard HN-support. The 1D systematic errors in the x, y and z directions were reduced from 2.2-2.3 mm to 1.2-2.0 mm (1 S.D.). The 1D random errors for the y and z directions were reduced from 1.6 and 1.6 mm to 1.1 and 1.0 mm (1 S.D.). The correction protocol reduced the 1D systematic errors further to 0.8-1.1 mm (1 S.D.) and all deviations in any direction were within 5 mm. Treatment time per measured fraction was increased from 10 to 13 min. The total time required per patient, for the complete correction procedure, was approximately 40 min. Conclusions: Portal imaging is a powerful tool in the evaluation of the department specific patient positioning procedures. The introduction of a comfortable customised HN-support, in combination with an electronic portal imaging device-based correction protocol, executed by technologists, led to an improvement of overall patient set-up. As a result, application of proposed recipes for CTV-PTV margins indicates that these can be reduced to 3-4 mm. (C) 2003 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:137 / 148
页数:12
相关论文
共 5 条
  • [1] Evaluation of set-up errors in head and neck radiotherapy using electronic portal imaging
    Strbac, Bojan
    Jokic, Vesna Spasic
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (05): : 531 - 536
  • [2] Evaluation of Patient Set Up Errors in Head and Neck Three-Dimensional (3D) Conformal and Intensity-Modulated Radiotherapy Using Electronic Portal Imaging Device
    Farajollahi A.
    Mohammadzadeh N.
    Momennezhad M.
    Naseri S.
    Mohebbi S.
    Shahedi F.
    Mohebbi S.
    Iranian Journal of Medical Physics, 2022, 19 (05) : 270 - 274
  • [3] Optimisation of cone beam CT radiotherapy imaging protocols using a novel 3D printed head and neck anthropomorphic phantom
    Alzahrani, Meshal
    O'Hara, Christopher
    Bird, David
    Baldwin, Jack P. C.
    Naisbit, Mitchell
    Teh, Irvin
    Broadbent, David A.
    Al-Qaisieh, Bashar
    Johnstone, Emily
    Speight, Richard
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (21):
  • [4] Organ at Risk Segmentation in Head and Neck CT Images Using a Two-Stage Segmentation Framework Based on 3D U-Net
    Wang, Yueyue
    Zhao, Liang
    Wang, Manning
    Song, Zhijian
    IEEE ACCESS, 2019, 7 : 144591 - 144602
  • [5] Evaluation of kV-CBCT based 3D dose calculation accuracy and its validation using delivery fluence derived dose metrics in Head and Neck Cancer
    Shinde, Prashantkumar
    Jadhav, Anand
    Shankar, V.
    Gupta, Karan Kumar
    Dhoble, Nirupama S.
    Dhoble, Sanjay
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 96 : 32 - 45