Ideal response function of a 3He proportional counter to thermal neutrons determined by different length counters

被引:6
|
作者
Kudo, K
Takeda, N
Uritani, A
Koshikawa, S
Shibata, Y
Kobayashi, K
Yoshimoto, T
机构
[1] Natl Inst AIST, Quantum Radiat Div, NMIJ, Tsukuba, Ibaraki 3058568, Japan
[2] Kyoto Univ, Inst Res Reactor, Osaka 5900494, Japan
关键词
thermal neutrons; response function; He-3 proportional counter; Monte Carlo code;
D O I
10.1016/S0168-583X(03)01633-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The gas multiplication distributions along the anode wires of He-3 proportional counters with different sensitive lengths of 10, 15 and 20 cm were measured using a collimated thermal neutron beam at the Kyoto University Reactor facility. The ideal response function excluding the effect of electric field distortion near the boundaries of the anode wire was derived by taking into account the difference between pulse height spectra measured by two different length counters. The pulse height spectrum calculated by the NRESPG Monte Carlo code assuming the ideal gas multiplication region showed relatively good agreement with the response function derived from the measurements. Both the energy resolution of the full energy peak at 764 keV and the peak ratio to the entire pulse height spectrum were improved markedly. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:305 / 309
页数:5
相关论文
共 2 条
  • [1] Event identification in 3He proportional counters using risetime discrimination
    Langford, T. J.
    Bass, C. D.
    Beise, E. J.
    Breuer, H.
    Erwin, D. K.
    Heimbach, C. R.
    Nico, J. S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 717 : 51 - 57
  • [2] DEVELOPMENT OF POSITION-SENSITIVE HE-3 PROPORTIONAL COUNTER FOR THERMAL-NEUTRONS WITH NOVEL SIGNAL READOUT
    URITANI, A
    MORI, C
    WATANABE, T
    MIYAHARA, A
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 1990, 27 (08) : 712 - 717