Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A phantom study

被引:20
作者
Dabli, Djamel [1 ]
Frandon, Julien [1 ]
Belaouni, Asmaa [1 ]
Akessoul, Philippe [1 ]
Addala, Takieddine [1 ]
Berny, Laure [1 ]
Beregi, Jean-Paul [1 ]
Greffer, Joel [1 ]
机构
[1] Univ Montpellier, Dept Med Imaging, CHU Nimes, Med Imaging Grp Nimes,EA 2994, Nimes, France
关键词
Dual-energy CT; Iodine quantification; Monochromatic images; Iterative reconstruction; ENERGY CT; COMPUTED-TOMOGRAPHY; PERFORMANCE; DIAGNOSIS; CELL; KV;
D O I
10.1016/j.diii.2021.08.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to assess the impact of advanced modeled iterative reconstruction (ADMIRE) algorithm and dose levels on the accuracy of Hounsfield unit (HU) measurement, image noise and contrast-to-noise ratio (CNR) in virtual monochromatic images (VMIs) with low iodine concentrations, and the accuracy of iodine quantification. Materials and methods: A CT phantom was scanned with dual-source CT using abdomen-pelvis examination parameters at four dose levels: 5, 8, 11 and 20 mGy. Images were reconstructed using filtered-back projection (FBP) and ADMIRE levels 3 and 5 (A3-A5). HU accuracy was assessed calculating the root-mean-square deviation (RMSDHU). Image noise and CNR were computed on VMIs at 40/50/60/70 keV for 4 iodine inserts with 0.5, 1, 2 and 5 mg/mL concentrations. Accuracy of iodine quantification was assessed by the RMSDiodine and iodine bias (IB). Results: The RMSDHU decreased significantly as the dose levels increased compared to 5 mGy, except for 8 mGy with A3 (P = 0.380) and with A5 level (P = 0.945). Noise increased by 63.0 +/- 3.0 (standard deviation [SD])% from 20 mGy to 5 mGy. Noise decreased significantly by -53.8 +/- 0.9 (SD) % with A5 compared to FBP. The CNR decreased by -43.1 +/- 6.5 (SD)% from 20 mGy to 5 mGy. It increased using ADMIRE, and as the ADMIRE levels increased. The RMSDiodine and IB decreased as the dose level increased, and this was similar for all reconstruction types. Conclusion: ADMIRE strongly improves image quality in VMIs and slightly improves HU accuracy but does not affect the accuracy of iodine quantification. (c) 2021 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:31 / 40
页数:10
相关论文
共 50 条
[31]   Image Quality Improvement of Low-dose Abdominal CT using Deep Learning Image Reconstruction Compared with the Second Generation Iterative Reconstruction [J].
Kang, Hyo-Jin ;
Lee, Jeong Min ;
Park, Sae Jin ;
Lee, Sang Min ;
Joo, Ijin ;
Yoon, Jeong Hee .
CURRENT MEDICAL IMAGING, 2024, 20
[32]   Radiation dose and physical image quality in 128-section dual-source computed tomographic coronary angiography: a phantom study [J].
Matsubara, Kosuke ;
Koshida, Haruka ;
Sakuta, Keita ;
Takata, Tadanori ;
Horii, Junsei ;
Iida, Hiroji ;
Koshida, Kichiro ;
Ichikawa, Katsuhiro ;
Matsui, Osamu .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (05) :252-261
[33]   Metal artifact reduction combined with deep learning image reconstruction algorithm for CT image quality optimization: a phantom study [J].
Zou, Huachun ;
Wang, Zonghuo ;
Guo, Mengya ;
Peng, Kun ;
Zhou, Jian ;
Zhou, Lili ;
Fan, Bing .
PEERJ, 2025, 13
[34]   Acquisition time, radiation dose, subjective and objective image quality of dual-source CT scanners in acute pulmonary embolism: a comparative study [J].
Abdellatif, Waleed ;
Esslinger, Eric ;
Kobes, Kevin ;
Wong, Amanda ;
Powell, Jennifer ;
Ali, Ismail Tawakol ;
Andrews, Gordon ;
Nicolaou, Savvas .
EUROPEAN RADIOLOGY, 2020, 30 (05) :2712-2721
[35]   Accuracy and precision of volumetric bone mineral density assessment using dual-source dual-energy versus quantitative CT: a phantom study [J].
Koch, Vitali ;
Hokamp, Nils Grosse ;
Albrecht, Moritz H. ;
Gruenewald, Leon D. ;
Yel, Ibrahim ;
Borggrefe, Jan ;
Wesarg, Stefan ;
Eichler, Katrin ;
Burck, Iris ;
Gruber-Rouh, Tatjana ;
Lenga, Lukas ;
Vogl, Thomas J. ;
Martin, Simon S. ;
Wichmann, Julian L. ;
Hammerstingl, Renate M. ;
Alizadeh, Leona S. ;
Mader, Christoph ;
Huizinga, Nicole A. ;
D'Angelo, Tommaso ;
Ascenti, Giorgio ;
Mazziotti, Silvio ;
Booz, Christian .
EUROPEAN RADIOLOGY EXPERIMENTAL, 2021, 5 (01)
[36]   Impact of dose reduction and the use of an advanced model-based iterative reconstruction algorithm on spectral performance of a dual-source CT system: A task-based image quality assessment [J].
Greffier, Joel ;
Dabli, Djamel ;
Hamard, Aymeric ;
Akessoul, Philippe ;
Belaouni, Asmaa ;
Beregi, Jean-Paul ;
Frandon, Julien .
DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2021, 102 (7-8) :405-412
[37]   Impact of Patient Size and Radiation Dose on Accuracy and Precision of Iodine Quantification and Virtual Noncontrast Values in Dual-layer Detector CT-A Phantom Study [J].
Van Hedent, Steven ;
Tatsuoka, Curtis ;
Carr, Sarah ;
Laukamp, Kai Roman ;
Eck, Brendan ;
Hokamp, Nils Grosse ;
Kessner, Rivka ;
Ros, Pablo ;
Jordan, David .
ACADEMIC RADIOLOGY, 2020, 27 (03) :409-420
[38]   Optimization of dose and image quality using self-produced phantom with various diameters in pediatric abdominal CT scan [J].
Choi, Hye Ran ;
Kim, Ryeong Eon ;
Heo, Chang Won ;
Kim, Chae Won ;
Yoo, Myung Seok ;
Lee, Youngjin .
OPTIK, 2018, 168 :54-60
[39]   Third-generation dual-source CT of the neck using automated tube voltage adaptation in combination with advanced modeled iterative reconstruction: evaluation of image quality and radiation dose [J].
Scholtz, Jan-Erik ;
Wichmann, Julian L. ;
Huesers, Kristina ;
Albrecht, Moritz H. ;
Beeres, Martin ;
Bauer, Ralf W. ;
Vogl, Thomas J. ;
Bodelle, Boris .
EUROPEAN RADIOLOGY, 2016, 26 (08) :2623-2631
[40]   Radiation exposure and image quality in staged low-dose protocols for coronary dual-source CT angiography: a randomized comparison [J].
Tobias Pflederer ;
Josy Jakstat ;
Mohamed Marwan ;
Tiziano Schepis ;
Sven Bachmann ;
Axel Kuettner ;
Katharina Anders ;
Michael Lell ;
Gerd Muschiol ;
Dieter Ropers ;
Werner G. Daniel ;
Stephan Achenbach .
European Radiology, 2010, 20 :1197-1206