Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A phantom study

被引:20
作者
Dabli, Djamel [1 ]
Frandon, Julien [1 ]
Belaouni, Asmaa [1 ]
Akessoul, Philippe [1 ]
Addala, Takieddine [1 ]
Berny, Laure [1 ]
Beregi, Jean-Paul [1 ]
Greffer, Joel [1 ]
机构
[1] Univ Montpellier, Dept Med Imaging, CHU Nimes, Med Imaging Grp Nimes,EA 2994, Nimes, France
关键词
Dual-energy CT; Iodine quantification; Monochromatic images; Iterative reconstruction; ENERGY CT; COMPUTED-TOMOGRAPHY; PERFORMANCE; DIAGNOSIS; CELL; KV;
D O I
10.1016/j.diii.2021.08.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to assess the impact of advanced modeled iterative reconstruction (ADMIRE) algorithm and dose levels on the accuracy of Hounsfield unit (HU) measurement, image noise and contrast-to-noise ratio (CNR) in virtual monochromatic images (VMIs) with low iodine concentrations, and the accuracy of iodine quantification. Materials and methods: A CT phantom was scanned with dual-source CT using abdomen-pelvis examination parameters at four dose levels: 5, 8, 11 and 20 mGy. Images were reconstructed using filtered-back projection (FBP) and ADMIRE levels 3 and 5 (A3-A5). HU accuracy was assessed calculating the root-mean-square deviation (RMSDHU). Image noise and CNR were computed on VMIs at 40/50/60/70 keV for 4 iodine inserts with 0.5, 1, 2 and 5 mg/mL concentrations. Accuracy of iodine quantification was assessed by the RMSDiodine and iodine bias (IB). Results: The RMSDHU decreased significantly as the dose levels increased compared to 5 mGy, except for 8 mGy with A3 (P = 0.380) and with A5 level (P = 0.945). Noise increased by 63.0 +/- 3.0 (standard deviation [SD])% from 20 mGy to 5 mGy. Noise decreased significantly by -53.8 +/- 0.9 (SD) % with A5 compared to FBP. The CNR decreased by -43.1 +/- 6.5 (SD)% from 20 mGy to 5 mGy. It increased using ADMIRE, and as the ADMIRE levels increased. The RMSDiodine and IB decreased as the dose level increased, and this was similar for all reconstruction types. Conclusion: ADMIRE strongly improves image quality in VMIs and slightly improves HU accuracy but does not affect the accuracy of iodine quantification. (c) 2021 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:31 / 40
页数:10
相关论文
共 50 条
[21]   Effect of low tube voltage and low iodine concentration abdominal CT on image quality and radiation dose in children: preliminary study [J].
Sun Kyoung You ;
Young Hun Choi ;
Jung-Eun Cheon ;
Woo Sun Kim ;
In-One Kim ;
So Mi Lee ;
Hyun-Hae Cho .
Abdominal Radiology, 2019, 44 :1928-1935
[22]   Effect of iterative reconstruction techniques on image quality in low radiation dose chest CT: a phantom study [J].
Xu, Yan ;
Zhang, Ting-ting ;
Hu, Zhi-hai ;
Li, Juan ;
Hou, Hong-jun ;
Xu, Zu-shan ;
He, Wen .
DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2019, 25 (06) :442-450
[23]   Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study [J].
Meinel, Felix G. ;
Canstein, Christian ;
Schoepf, U. Joseph ;
Sedlmaier, Martin ;
Schmidt, Bernhard ;
Harris, Brett S. ;
Flohr, Thomas G. ;
De Cecco, Carlo N. .
EUROPEAN RADIOLOGY, 2014, 24 (07) :1643-1650
[24]   Intraindividual Comparison of Image Quality Between Low-Dose and Ultra-Low-Dose Abdominal CT With Deep Learning Reconstruction and Standard-Dose Abdominal CT Using Dual-Split Scan [J].
Lee, Tae Young ;
Yoon, Jeong Hee ;
Park, Jin Young ;
Park, So Hyun ;
Kim, Heesoo ;
Lee, Chul-min ;
Choi, Yunhee ;
Lee, Jeong Min .
INVESTIGATIVE RADIOLOGY, 2025, 60 (07) :454-462
[25]   Effects of radiation dose levels and spectral iterative reconstruction levels on the accuracy of iodine quantification and virtual monochromatic CT numbers in dual-layer spectral detector CT: an iodine phantom study [J].
Lu, Xiaomei ;
Lu, Zaiming ;
Yin, Jiandong ;
Gao, Yuying ;
Chen, Xingbiao ;
Guo, Qiyong .
QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (02) :188-200
[26]   A feasibility study of realizing low-dose abdominal CT using deep learning image reconstruction algorithm [J].
Li, Lu-Lu ;
Wang, Huang ;
Song, Jian ;
Shang, Jin ;
Zhao, Xiao-Ying ;
Liu, Bin .
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2021, 29 (02) :361-372
[27]   Low-dose liver CT: image quality and diagnostic accuracy of deep learning image reconstruction algorithm [J].
Damiano Caruso ;
Domenico De Santis ;
Antonella Del Gaudio ;
Gisella Guido ;
Marta Zerunian ;
Michela Polici ;
Daniela Valanzuolo ;
Dominga Pugliese ;
Raffaello Persechino ;
Antonio Cremona ;
Luca Barbato ;
Andrea Caloisi ;
Elsa Iannicelli ;
Andrea Laghi .
European Radiology, 2024, 34 :2384-2393
[28]   Image quality of ultra-low-dose dual-source CT angiography using high-pitch spiral acquisition and iterative reconstruction in young children with congenital heart disease [J].
Zheng, Minwen ;
Zhao, Hongliang ;
Xu, Jian ;
Wu, Yongjie ;
Li, Jian .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2013, 7 (06) :376-382
[29]   Low-dose liver CT: image quality and diagnostic accuracy of deep learning image reconstruction algorithm [J].
Caruso, Damiano ;
De Santis, Domenico ;
Del Gaudio, Antonella ;
Guido, Gisella ;
Zerunian, Marta ;
Polici, Michela ;
Valanzuolo, Daniela ;
Pugliese, Dominga ;
Persechino, Raffaello ;
Cremona, Antonio ;
Barbato, Luca ;
Caloisi, Andrea ;
Iannicelli, Elsa ;
Laghi, Andrea .
EUROPEAN RADIOLOGY, 2024, 34 (04) :2384-2393
[30]   Comparison of two versions of a deep learning image reconstruction algorithm on CT image quality and dose reduction: A phantom study [J].
Greffier, Joel ;
Dabli, Djamel ;
Frandon, Julien ;
Hamard, Aymeric ;
Belaouni, Asmaa ;
Akessoul, Philippe ;
Fuamba, Yannick ;
Le Roy, Julien ;
Guiu, Boris ;
Beregi, Jean-Paul .
MEDICAL PHYSICS, 2021, 48 (10) :5743-5755