Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A phantom study

被引:18
作者
Dabli, Djamel [1 ]
Frandon, Julien [1 ]
Belaouni, Asmaa [1 ]
Akessoul, Philippe [1 ]
Addala, Takieddine [1 ]
Berny, Laure [1 ]
Beregi, Jean-Paul [1 ]
Greffer, Joel [1 ]
机构
[1] Univ Montpellier, Dept Med Imaging, CHU Nimes, Med Imaging Grp Nimes,EA 2994, Nimes, France
关键词
Dual-energy CT; Iodine quantification; Monochromatic images; Iterative reconstruction; ENERGY CT; COMPUTED-TOMOGRAPHY; PERFORMANCE; DIAGNOSIS; CELL; KV;
D O I
10.1016/j.diii.2021.08.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to assess the impact of advanced modeled iterative reconstruction (ADMIRE) algorithm and dose levels on the accuracy of Hounsfield unit (HU) measurement, image noise and contrast-to-noise ratio (CNR) in virtual monochromatic images (VMIs) with low iodine concentrations, and the accuracy of iodine quantification. Materials and methods: A CT phantom was scanned with dual-source CT using abdomen-pelvis examination parameters at four dose levels: 5, 8, 11 and 20 mGy. Images were reconstructed using filtered-back projection (FBP) and ADMIRE levels 3 and 5 (A3-A5). HU accuracy was assessed calculating the root-mean-square deviation (RMSDHU). Image noise and CNR were computed on VMIs at 40/50/60/70 keV for 4 iodine inserts with 0.5, 1, 2 and 5 mg/mL concentrations. Accuracy of iodine quantification was assessed by the RMSDiodine and iodine bias (IB). Results: The RMSDHU decreased significantly as the dose levels increased compared to 5 mGy, except for 8 mGy with A3 (P = 0.380) and with A5 level (P = 0.945). Noise increased by 63.0 +/- 3.0 (standard deviation [SD])% from 20 mGy to 5 mGy. Noise decreased significantly by -53.8 +/- 0.9 (SD) % with A5 compared to FBP. The CNR decreased by -43.1 +/- 6.5 (SD)% from 20 mGy to 5 mGy. It increased using ADMIRE, and as the ADMIRE levels increased. The RMSDiodine and IB decreased as the dose level increased, and this was similar for all reconstruction types. Conclusion: ADMIRE strongly improves image quality in VMIs and slightly improves HU accuracy but does not affect the accuracy of iodine quantification. (c) 2021 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:31 / 40
页数:10
相关论文
共 50 条
  • [21] Effect of low tube voltage and low iodine concentration abdominal CT on image quality and radiation dose in children: preliminary study
    Sun Kyoung You
    Young Hun Choi
    Jung-Eun Cheon
    Woo Sun Kim
    In-One Kim
    So Mi Lee
    Hyun-Hae Cho
    Abdominal Radiology, 2019, 44 : 1928 - 1935
  • [22] Effect of low tube voltage and low iodine concentration abdominal CT on image quality and radiation dose in children: preliminary study
    You, Sun Kyoung
    Choi, Young Hun
    Cheon, Jung-Eun
    Kim, Woo Sun
    Kim, In-One
    Lee, So Mi
    Cho, Hyun-Hae
    ABDOMINAL RADIOLOGY, 2019, 44 (05) : 1928 - 1935
  • [23] Effect of iterative reconstruction techniques on image quality in low radiation dose chest CT: a phantom study
    Xu, Yan
    Zhang, Ting-ting
    Hu, Zhi-hai
    Li, Juan
    Hou, Hong-jun
    Xu, Zu-shan
    He, Wen
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2019, 25 (06) : 442 - 450
  • [24] Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study
    Meinel, Felix G.
    Canstein, Christian
    Schoepf, U. Joseph
    Sedlmaier, Martin
    Schmidt, Bernhard
    Harris, Brett S.
    Flohr, Thomas G.
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2014, 24 (07) : 1643 - 1650
  • [25] Effects of radiation dose levels and spectral iterative reconstruction levels on the accuracy of iodine quantification and virtual monochromatic CT numbers in dual-layer spectral detector CT: an iodine phantom study
    Lu, Xiaomei
    Lu, Zaiming
    Yin, Jiandong
    Gao, Yuying
    Chen, Xingbiao
    Guo, Qiyong
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (02) : 188 - 200
  • [26] A feasibility study of realizing low-dose abdominal CT using deep learning image reconstruction algorithm
    Li, Lu-Lu
    Wang, Huang
    Song, Jian
    Shang, Jin
    Zhao, Xiao-Ying
    Liu, Bin
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2021, 29 (02) : 361 - 372
  • [27] Image quality of ultra-low-dose dual-source CT angiography using high-pitch spiral acquisition and iterative reconstruction in young children with congenital heart disease
    Zheng, Minwen
    Zhao, Hongliang
    Xu, Jian
    Wu, Yongjie
    Li, Jian
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2013, 7 (06) : 376 - 382
  • [28] Low-dose liver CT: image quality and diagnostic accuracy of deep learning image reconstruction algorithm
    Damiano Caruso
    Domenico De Santis
    Antonella Del Gaudio
    Gisella Guido
    Marta Zerunian
    Michela Polici
    Daniela Valanzuolo
    Dominga Pugliese
    Raffaello Persechino
    Antonio Cremona
    Luca Barbato
    Andrea Caloisi
    Elsa Iannicelli
    Andrea Laghi
    European Radiology, 2024, 34 : 2384 - 2393
  • [29] Comparison of two versions of a deep learning image reconstruction algorithm on CT image quality and dose reduction: A phantom study
    Greffier, Joel
    Dabli, Djamel
    Frandon, Julien
    Hamard, Aymeric
    Belaouni, Asmaa
    Akessoul, Philippe
    Fuamba, Yannick
    Le Roy, Julien
    Guiu, Boris
    Beregi, Jean-Paul
    MEDICAL PHYSICS, 2021, 48 (10) : 5743 - 5755
  • [30] Radiation dose and physical image quality in 128-section dual-source computed tomographic coronary angiography: a phantom study
    Matsubara, Kosuke
    Koshida, Haruka
    Sakuta, Keita
    Takata, Tadanori
    Horii, Junsei
    Iida, Hiroji
    Koshida, Kichiro
    Ichikawa, Katsuhiro
    Matsui, Osamu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (05): : 252 - 261