Acceptability of CoaguChek S and CoaguChek XS generated international normalised ratios against a laboratory standard in a paediatric setting

被引:28
|
作者
Williams, Vaughan K. [1 ]
Griffiths, Adrian B. M. [1 ]
机构
[1] Womens & Childrens Hosp, Dept Haematol, Adelaide, SA, Australia
关键词
CoaguChek S; CoaguChek XS; point of care testing; POC; INR;
D O I
10.1080/00313020701684797
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Aims: Point of care (POC) testing is convenient for both patients and physicians but the accuracy of the results must be guaranteed. In this study we evaluated the accuracy of the international normalised ratios (INR) generated from the CoaguChek S and the newer CoaguChek XS against our laboratory standard. Methods: The INR values obtained from 87 tests on the CoaguChek S and 97 tests on the CoaguChek XS were compared with those generated from a laboratory based coagulation analyser. The results were statistically analysed. Results: The values from the CoaguChek XS were closer to the laboratory based results than those from the CoaguChek S but were still significantly different (p<0.001). Using an arbitrary INR difference of <= 0.5 INR units as a basis for comparable results, then 43% of the CoaguChek S and 19% of the CoaguChek XS results would have been considered unacceptable. Warfarin dose changes would have been considered in 25% of the CoaguChek S cases and 13% of the CoaguChek XS cases. Despite the disparity between the INR results from each of the methods, external quality assurance programs gave excellent results for both the CoaguChek S and the laboratory based INR results. Conclusions: The CoaguChek S INR will be within 0.5 INR units of the laboratory result up to an INR of 4 in only 66% of cases compared with 96% for the CoaguChek XS. Thus this study suggests that CoaguChek S machines should be substituted with CoaguChek XS and these can be used with confidence for INR results up to 4.
引用
收藏
页码:575 / 579
页数:5
相关论文
共 12 条
  • [1] International normalised ratio (INR) measured on the CoaguChek S and XS compared with the laboratory for determination of precision and accuracy
    Christensen, Thomas D.
    Larsen, Torben B.
    Jensen, Claus
    Maegaard, Marianne
    Sorensen, Benny
    THROMBOSIS AND HAEMOSTASIS, 2009, 101 (03) : 563 - 569
  • [2] Assessment of International Normalized Ratio Using CoaguChek XS and CoaguChek S as Compared With Central Laboratory Testing
    Adkinson, Casey L.
    Pettus, Joanna D.
    Chirico, Mark J.
    Taylor, James R.
    POINT OF CARE, 2009, 8 (03): : 126 - 130
  • [3] Assessment of point-of-care measurement of international normalised ratio using the CoaguChek XS Plus system in the setting of acute ischaemic stroke
    Nusa, D.
    Harvey, I.
    Almansouri, A. Y.
    Wright, S.
    Neeman, T.
    Ahmad, O.
    Hughes, A. R.
    Lueck, C. J.
    INTERNAL MEDICINE JOURNAL, 2013, 43 (11) : 1205 - 1209
  • [4] CoaguChek (R) XS versus standard laboratory prothrombin time for anticoagulant monitoring in patients with antiphospholipid syndrome
    Fonseca, Maria Ester S.
    Balbi, Gustavo G. M.
    Signorelli, Flavio
    Gouvea, Christiane P.
    de Andrade, Danieli C. O.
    LUPUS, 2022, 31 (05) : 565 - 574
  • [5] Comparative study of accuracy and clinical agreement of the CoaguChek XS portable device versus standard laboratory practice in unexperienced patients
    Torreiro, Eduardo G.
    Gomez Fernandez, Elizabeth
    Marino Rodriguez, Rosa
    Vazquez Lopez, Carmen
    Barreal Nunez, Julia
    THROMBOSIS AND HAEMOSTASIS, 2009, 101 (05) : 969 - 974
  • [6] Comparative evaluation of Point Of Care coagulation monitoring by coaguchek XS-comparison with standard laboratory method
    Lakshmy R.
    Kumar A.S.
    Indian Journal of Thoracic and Cardiovascular Surgery, 2010, 26 (2) : 125 - 128
  • [7] Comparison of point-of-care analysis using Coaguchek XS and standard laboratory-measured prothrombin time in dogs
    Newbould, A. R.
    Norman, E. J.
    NEW ZEALAND VETERINARY JOURNAL, 2013, 61 (01) : 18 - 24
  • [8] Accuracy and clinical usefulness of the CoaguChek S and XS Point of Care devices when starting warfarin in a hospital outreach setting
    Sobieraj-Teague, Magdalena
    Daniel, Dolly
    Farrelly, Barbara
    Coghlan, Douglas
    Gallus, Alexander
    THROMBOSIS RESEARCH, 2009, 123 (06) : 909 - 913
  • [9] Point-of-care testing for coagulation function: CoaguChek® XS System versus standard laboratory testing in pediatric patients with normal and abnormal coagulation function
    Kako, Hiromi
    Raman, Vidya T.
    Tumin, Dmitry
    Rice, Julie
    Tobias, Joseph D.
    JOURNAL OF ANESTHESIA, 2017, 31 (03) : 345 - 350
  • [10] Point-of-care monitoring of vitamin K-antagonists: validation of CoaguChek XS test strips with International Standard thromboplastin
    van den Besselaar, Anton M. H. P.
    Pequeriaux, Nathalie C. V.
    Ebben, Marij
    van der Feest, Joke
    de Jong, Kerst
    Ganzeboom, Maria B. J.
    van Ooijen, Jetty
    Postema, Ferna
    Witteveen, Evelina
    van der Meer, Felix J. M.
    JOURNAL OF CLINICAL PATHOLOGY, 2012, 65 (11) : 1031 - 1035