Experimental Validation of an Efficient Fan-Beam Calibration Procedure for k-Nearest Neighbor Position Estimation in Monolithic Scintillator Detectors

被引:40
作者
Borghi, Giacomo [1 ]
Tabacchini, Valerio [1 ]
Seifert, Stefan [1 ]
Schaart, Dennis R. [1 ]
机构
[1] Delft Univ Technol, NL-2629 JB Delft, Netherlands
关键词
Calibration; fan beam; monolithic scintillator detector; nearest neighbor method; PET; positron emission tomography; 3D INTERACTION POSITION; PET DETECTORS; SILICON PHOTOMULTIPLIER; CONTINUOUS CRYSTALS; SPATIAL-RESOLUTION; APD ARRAYS; READ-OUT; DEPTH; PERFORMANCE; BLOCKS;
D O I
10.1109/TNS.2014.2375557
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Monolithic scintillator detectors can achieve excellent spatial resolution and coincidence resolving time. However, their practical use for positron emission tomography (PET) and other applications in the medical imaging field is still limited due to drawbacks of the different methods used to estimate the position of interaction. Common statistical methods for example require the collection of an extensive dataset of reference events with a narrow pencil beam aimed at a fine grid of reference positions. Such procedures are time consuming and not straightforwardly implemented in systems composed of many detectors. Here, we experimentally demonstrate for the first time a new calibration procedure for - nearest neighbor (k-NN) position estimation that utilizes reference data acquired with a fan beam. The procedure is tested on two detectors consisting of 16 mm x 16 mm x 10 mm and 16 mm x 16 mm x 20 mm monolithic, Ca-codoped LSO:Ce crystals and digital photon counter (DPC) arrays. For both detectors, the spatial resolution and the bias obtained with the new method are found to be practically the same as those obtained with the previously used method based on pencil-beam irradiation, while the calibration time is reduced by a factor of similar to 20 Specifically, a FWHM of similar to 1.1 mm and a FWTM of similar to 2.7 mm were obtained using the fan-beam method with the 10 mm crystal, whereas a FWHM of similar to 1.5 mm and a FWTM of similar to 6 mm were achieved with the 20 mm crystal. Using a fan beam made with a similar to 4.5 MBq Na-22 point-source and a tungsten slit collimator with 0.5 mm aperture, the total measurement time needed to acquire the reference dataset was similar to 3 hours for the thinner crystal and similar to 2 hours for the thicker one.
引用
收藏
页码:57 / 67
页数:11
相关论文
共 30 条
[1]   Initial characterization of a nonpixelated scintillator detector in a PET prototype demonstrator [J].
Bruyndonckx, P. ;
Lemaitre, Cedric ;
Leonard, Sophie ;
Schaart, Dennis R. ;
van der Laan, D. J. ;
Maas, Marnix C. ;
Bevroede, Olivier ;
Wu, Yibao ;
Krieguer, Magalie ;
Tavernier, Stefaan .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (05) :2543-2548
[2]   Neural network-based position estimators for PET detectors using monolithic LSO blocks [J].
Bruyndonckx, P ;
Léonard, S ;
Tavernier, S ;
Lemaître, C ;
Devroede, O ;
Wu, YB ;
Krieguer, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) :2520-2525
[3]   Evaluation of machine learning algorithms for localization of photons in undivided scintillator blocks for PET detectors [J].
Bruyndonckx, Peter ;
Lemaitre, Cedric ;
van der Laan, D. J. ;
Maas, Marnix ;
Schaart, Dennis ;
Wang Yonggang ;
Li, Zhi ;
Krieguer, M. ;
Tavernier, Stefaan .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) :918-924
[4]   High resolution detectors based on continuous crystals and SiPMs for small animal PET [J].
Cabello, J. ;
Barrillon, P. ;
Barrio, J. ;
Bisogni, M. G. ;
Del Guerra, A. ;
Lacasta, C. ;
Rafecas, M. ;
Saikouk, H. ;
Solaz, C. ;
Solevi, P. ;
de La Taille, C. ;
Llosa, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 718 :148-150
[5]   Performance of a DOI-encoding small animal PET system with monolithic scintillators [J].
Carles, M. ;
Lerche, Ch. W. ;
Sanchez, F. ;
Orero, A. ;
Moliner, L. ;
Soriano, A. ;
Benlloch, J. M. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 695 :317-321
[6]   Fast calibration of SPECT monolithic scintillation detectors using un-collimated sources [J].
Espana, Samuel ;
Deprez, Karel ;
Van Holen, Roel ;
Vandenberghe, Stefaan .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (14) :4807-4825
[7]  
Frach T, 2010, IEEE NUCL SCI CONF R, P1722, DOI 10.1109/NSSMIC.2010.5874069
[8]   The Digital Silicon Photomultiplier - Principle of Operation and Intrinsic Detector Performance [J].
Frach, Thomas ;
Prescher, Gordian ;
Degenhardt, Carsten ;
de Gruyter, Rik ;
Schmitz, Anja ;
Ballizany, Rob .
2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, :1959-1965
[9]   Calibration Method for ML Estimation of 3D Interaction Position in a Thick Gamma-Ray Detector [J].
Hunter, William C. J. ;
Barrett, Harrison H. ;
Furenlid, Lars R. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (01) :189-196
[10]   Combining Surface Treatments With Shallow Slots to Improve the Spatial Resolution Performance of Continuous, Thick LYSO Detectors for PET [J].
Kaul, M. ;
Surti, S. ;
Karp, J. S. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (01) :44-52