Evaluation of Microwave Steam Bags for the Decontamination of Filtering Facepiece Respirators

被引:64
作者
Fisher, Edward M. [1 ]
Williams, Jessica L. [1 ]
Shaffer, Ronald E. [1 ]
机构
[1] NIOSH, Natl Personal Protect Technol Lab, Pittsburgh, PA USA
关键词
D O I
10.1371/journal.pone.0018585
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Reusing filtering facepiece respirators (FFRs) has been suggested as a strategy to conserve available supplies for home and healthcare environments during an influenza pandemic. For reuse to be possible, used FFRs must be decontaminated before redonning to reduce the risk of virus transmission; however, there are no approved methods for FFR decontamination. An effective method must reduce the microbial threat, maintain the function of the FFR, and present no residual chemical hazard. The method should be readily available, inexpensive and easily implemented by healthcare workers and the general public. Many of the general decontamination protocols used in healthcare and home settings are unable to address all of the desired qualities of an efficient FFR decontamination protocol. The goal of this study is to evaluate the use of two commercially available steam bags, marketed to the public for disinfecting infant feeding equipment, for FFR decontamination. The FFRs were decontaminated with microwave generated steam following the manufacturers' instructions then evaluated for water absorption and filtration efficiency for up to three steam exposures. Water absorption of the FFR was found to be model specific as FFRs constructed with hydrophilic materials absorbed more water. The steam had little effect on FFR performance as filtration efficiency of the treated FFRs remained above 95%. The decontamination efficacy of the steam bag was assessed using bacteriophage MS2 as a surrogate for a pathogenic virus. The tested steam bags were found to be 99.9% effective for inactivating MS2 on FFRs; however, more research is required to determine the effectiveness against respiratory pathogens.
引用
收藏
页数:7
相关论文
共 26 条
[1]  
[Anonymous], 2008, Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008
[2]  
*ASTM INT, 2010, E27212010 ASTM
[3]  
Bergman MS, 2010, J ENG FIBER FABR, V5, P33
[4]  
CDC, 2010, QUEST ANSW REG RESP
[5]  
CDC, 1999, TB RESP PROT PROGR H
[6]   A method to determine the available UV-C dose for the decontamination of filtering facepiece respirators [J].
Fisher, E. M. ;
Shaffer, R. E. .
JOURNAL OF APPLIED MICROBIOLOGY, 2011, 110 (01) :287-295
[7]   Development of a Test System To Apply Virus-Containing Particles to Filtering Facepiece Respirators for the Evaluation of Decontamination Procedures [J].
Fisher, Edward ;
Rengasamy, Samy ;
Viscusi, Dennis ;
Vo, Evanly ;
Shaffer, Ronald .
APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 2009, 75 (06) :1500-1507
[8]  
Fisher EW., 2010, J INT SOC RESP PROTE, V27, P16
[9]  
GRIESBACH HC, 2008, DECONTAMINATION FILT
[10]   A pandemic influenza preparedness study: Use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets [J].
Heimbuch, Brian K. ;
Wallace, William H. ;
Kinney, Kimberly ;
Lumley, April E. ;
Wu, Chang-Yu ;
Woo, Myung-Heui ;
Wander, Joseph D. .
AMERICAN JOURNAL OF INFECTION CONTROL, 2011, 39 (01) :E1-E9