Image quality and dose in spiral computed tomography

被引:5
作者
Verdun, FR [1 ]
Meuli, RA [1 ]
Bochud, FO [1 ]
Imsand, C [1 ]
Raimondi, S [1 ]
Schnyder, P [1 ]
Valley, JF [1 ]
机构
[1] CHU VAUDOIS,SERV RADIODIAGNOST,CH-1011 LAUSANNE,SWITZERLAND
关键词
computed tomography; helical technology; image quality; radiation exposure; dosimetry;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Image quality and dose produced by a spiral CT with various pitch values were investigated on the basis of test-object images and measurements of CT dose indexes. No major difference in image quality or dose was observed when comparing sequential and spiral mode acquisition with a pitch of one. Increase in pitch value produces a loss of contrast which leads to a loss of detectability. Nevertheless, in realistic protocols the image quality in the transverse plane remains acceptable up to pitch values of 1.3-1.6.