Comparison of air displacement plethysmography with dual-energy X-ray absorptiometry and 3 field methods for estimating body composition in middle-aged men

被引:111
|
作者
Sardinha, LB
Lohman, TG
Teixeira, PJ
Guedes, DP
Going, SB
机构
[1] Univ Arizona, Dept Physiol, Body Composit Res Lab, Tucson, AZ 85721 USA
[2] Univ Tecn Lisboa, Fac Human Movement, Exercise & Hlth Lab, P-1100 Lisbon, Portugal
关键词
bioelectrical impedance analysis; BIA; bioelectrical impedance spectroscopy; dual-energy X-ray absorptiometry; DXA; air displacement plethysmography; percentage body fat; predictive accuracy; reliability; Portugal; men;
D O I
10.1093/ajcn/68.4.786
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
This study was designed to compare air displacement plethysmography with dual-energy X-ray absorptiometry (DXA) and 3 other field methods for estimation of body composition. Subjects were 62 healthy, white men aged 37.6 +/- 2.9 y (weight: 81.8 +/- 11.3 kg; height: 171.5 +/- 4.9 cm). Body composition was also assessed by using body mass index, single-frequency bioelectrical impedance analysis, multifrequency bioelectrical impedance spectroscopy, and the skinfold-thickness equations of Jackson and Pollock and Durnin and Womersley. Percentage body fat (%BF) with the plethysmograph was 23.4 +/- 7.0 and with DXA was 26.0 +/- 7.4. The 2.6% mean difference was significant (P < 0.05). Total error was 3.7%BF As assessed by multiple regression analysis, %BF with the plethysmograph, age, weight, and height yielded a DXA-adjusted R-2 value of 89.5% fat and an SEE of 2.4% fat. All other models had higher SEEs and lower adjusted R-2 values: 4.3% and 66.5% for body mass index, 3.3% and 79.8% for bioelectrical impedance analysis, 3.6% and 76.2% for bioelectrical impedance spectroscopy, 3.7% and 74.55% for the equations of Jackson and Pollock, and 3.9% and 71.6% for the equations of Durnin and Womersley, respectively. The plethysmograph also predicted fat mass and fat-free mass more accurately than all other models, with a lower SEE and higher adjusted R-2 value. In conclusion, although %BF was systematically underestimated, body composition was closely estimated with air displacement plethysmography in middle-aged men.
引用
收藏
页码:786 / 793
页数:8
相关论文
共 50 条
  • [21] VALIDITY AND RELIABILITY OF BIOELECTRICAL IMPEDANCE ANALYSIS TO ESTIMATE BODY FAT COMPARED TO AIR DISPLACEMENT PLETHYSMOGRAPHY AND DUAL-ENERGY X-RAY ABSORPTIOMETRY
    Conlon, C.
    Ingram, M.
    Walsh, D.
    von Hurst, P.
    Kruger, R.
    Stonehouse, W.
    ANNALS OF NUTRITION AND METABOLISM, 2013, 63 : 1478 - 1478
  • [22] Body composition by dual-energy x-ray absorptiometry in women with fibromyalgia
    Marques Teles Lobo, Marcia Maria
    Paiva, Eduardo dos Santos
    Andretta, Aline
    Madalozzo Schieferdecker, Maria Eliana
    REVISTA BRASILEIRA DE REUMATOLOGIA, 2014, 54 (04) : 273 - 278
  • [23] Pediatric body composition analysis with dual-energy X-ray absorptiometry
    Maura Helba
    Larry A. Binkovitz
    Pediatric Radiology, 2009, 39 : 647 - 656
  • [24] Body Composition in Karate: A Dual-Energy X-ray Absorptiometry Study
    Cavedon, Valentina
    Milanese, Chiara
    Sacristani, Franco
    Zancanaro, Carlo
    APPLIED SCIENCES-BASEL, 2023, 13 (01):
  • [25] Pediatric body composition analysis with dual-energy X-ray absorptiometry
    Helba, Maura
    Binkovitz, Larry A.
    PEDIATRIC RADIOLOGY, 2009, 39 (07) : 647 - 656
  • [26] Body Composition Comparison: Bioelectric Impedance Analysis with Dual-Energy X-Ray Absorptiometry in Adult Athletes
    Company, Joe
    Ball, Stephen
    MEASUREMENT IN PHYSICAL EDUCATION AND EXERCISE SCIENCE, 2010, 14 (03) : 186 - 201
  • [27] The influence of magnetic resonance imaging findings of degenerative disease on dual-energy X-ray absorptiometry measurements in middle-aged men
    Donescu, O. S.
    Battie, M. C.
    Videman, T.
    ACTA RADIOLOGICA, 2007, 48 (02) : 193 - 199
  • [28] ACCURACY OF DUAL-ENERGY X-RAY ABSORPTIOMETRY FOR BODY-COMPOSITION MEASUREMENTS IN CHILDREN
    ELLIS, KJ
    SHYPAILO, RJ
    PRATT, JA
    POND, WG
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1994, 60 (05) : 660 - 665
  • [29] Dual-energy X-ray absorptiometry for the measurement of gross body composition in rats
    Jebb, SA
    Garland, SW
    Jennings, G
    Elia, M
    BRITISH JOURNAL OF NUTRITION, 1996, 75 (06) : 803 - 809
  • [30] Dual-energy X-ray absorptiometry for body composition estimation in Chinese women
    M He
    ETS Li
    AWC Kung
    European Journal of Clinical Nutrition, 1999, 53 : 933 - 937