Glaucoma Diagnostic Ability of the New Circumpapillary Retinal Nerve Fiber Layer Thickness Analysis Based on Bruch's Membrane Opening

被引:21
作者
Lee, Eun Ji [1 ]
Lee, Kyoung Min [1 ]
Kim, Hyunjoong [2 ]
Kim, Tae-Woo [1 ]
机构
[1] Seoul Natl Univ, Bundang Hosp, Dept Ophthalmol, 300 Gumi Dong, Songnam 463707, Gyeonggi Do, South Korea
[2] Yonsei Univ, Dept Appl Stat, Seoul, South Korea
基金
新加坡国家研究基金会;
关键词
Bruch's membrane opening; glaucoma diagnosis; optical coherence tomography; retinal nerve fiber layer thickness; OPTICAL COHERENCE TOMOGRAPHY; DISC MARGIN ANATOMY; GENERALIZED ESTIMATING EQUATIONS; OPERATING CHARACTERISTIC CURVES; OPEN-ANGLE GLAUCOMA; SPECTRAL-DOMAIN; PARAPAPILLARY ATROPHY; REGRESSION-ANALYSIS; CIRCLE LOCATION; CORRELATED DATA;
D O I
10.1167/iovs.16-19578
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To compare the diagnostic ability of the new spectral-domain optical coherence tomography (SD-OCT) algorithm for measuring circumpapillary retinal nerve fiber layer (RNFL) thickness centered on Bruch's membrane opening (BMO), with the conventional circumpapillary RNFL thickness measurement centered on the optic disc. METHODS. In 75 eyes with primary open-angle glaucoma (POAG) and 71 healthy control eyes, circumpapillary RNFL thickness was measured with SD-OCT, first by centering the scan circle on the optic disc (RNFLDi), and then on the BMO (RNFLBMO). Difference between the topographic profiles of RNFLDi and RNFLBMO was compared and factors influencing any discrepancies between methods were investigated. Glaucoma diagnostic abilities of each method were assessed using the areas under receiver operating characteristic curve (AUCs). RESULTS. Axial length did not differ between POAG and healthy eyes. A longer axial length and larger width of externally oblique border tissue (BT) associated with tilted optic disc were the two major factors influencing discrepancies between RNFLBMO and RNFLDi (both P < 0.001). Compared with RNFLBMO, RNFLDi tended to result in a thinner nasal RNFL in eyes with externally oblique BT, while RNFLBMO and RNFLDi were comparable in eyes without externally oblique BT. The glaucoma diagnostic capabilities were generally comparable, but RNFLBMO was superior to RNFLDi in eyes having a larger width (> 250 mu m) of externally oblique BT (AUC = 0.933 vs. 0.843, respectively, P = 0.027). CONCLUSIONS. The new circumpapillary RNFL scanning algorithm centered on BMO may provide a more reliable RNFL profile in eyes with tilted optic discs, with a largely comparable glaucoma diagnostic ability to the conventional algorithm.
引用
收藏
页码:4194 / 4204
页数:11
相关论文
共 42 条
[1]   Analysis of Normal Retinal Nerve Fiber Layer Thickness by Age, Sex, and Race Using Spectral Domain Optical Coherence Tomography [J].
Alasil, Tarek ;
Wang, Kaidi ;
Keane, Pearse A. ;
Lee, Hang ;
Baniasadi, Neda ;
de Boer, Johannes F. ;
Chen, Teresa C. .
JOURNAL OF GLAUCOMA, 2013, 22 (07) :532-541
[2]   Reproducibility of nerve fiber layer thickness measurements by use of optical coherence tomography [J].
Blumenthal, EZ ;
Williams, JM ;
Weinreb, RN ;
Girkin, CA ;
Berry, CC ;
Zangwill, LM .
OPHTHALMOLOGY, 2000, 107 (12) :2278-2282
[3]   From Clinical Examination of the Optic Disc to Clinical Assessment of the Optic Nerve Head: A Paradigm Change [J].
Chauhan, Balwantray C. ;
Burgoyne, Claude F. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2013, 156 (02) :218-227
[4]   Enhanced Detection of Open-angle Glaucoma with an Anatomically Accurate Optical Coherence Tomography-Derived Neuroretinal Rim Parameter [J].
Chauhan, Balwantray C. ;
O'Leary, Neil ;
AlMobarak, Faisal A. ;
Reis, Alexandre S. C. ;
Yang, Hongli ;
Sharpe, Glen P. ;
Hutchison, Donna M. ;
Nicolela, Marcelo T. ;
Burgoyne, Claude F. .
OPHTHALMOLOGY, 2013, 120 (03) :535-543
[5]   Effects of scan circle displacement in optical coherence tomography retinal nerve fibre layer thickness measurement: a RNFL modelling study [J].
Cheung, C. Y. L. ;
Yiu, C. K. F. ;
Weinreb, R. N. ;
Lin, D. ;
Li, H. ;
Yung, A. Y. ;
Pang, C. P. ;
Lam, D. S. C. ;
Leung, C. K. S. .
EYE, 2009, 23 (06) :1436-1441
[6]   Correct Calculation Circle Location of Optical Coherence Tomography in Measuring Retinal Nerve Fiber Layer Thickness in Eyes with Myopic Tilted Discs [J].
Chung, Jae Keun ;
Yoo, Young Cheol .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (11) :7894-7900
[7]   Enhanced Structure-Function Relationship in Glaucoma With an Anatomically and Geometrically Accurate Neuroretinal Rim Measurement [J].
Danthurebandara, Vishva M. ;
Sharpe, Glen P. ;
Hutchison, Donna M. ;
Denniss, Jonathan ;
Nicolela, Marcelo T. ;
McKendrick, Allison M. ;
Turpin, Andrew ;
Chauhan, Balwantray C. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (01) :98-105
[8]   Retinal nerve fiber layer thickness measured with optical coherence tomography is related to visual function in glaucomatous eyes [J].
El Beltagi, TA ;
Bowd, C ;
Boden, C ;
Amini, P ;
Sample, PA ;
Zangwill, LM ;
Weinreb, RN .
OPHTHALMOLOGY, 2003, 110 (11) :2185-2191
[9]   Optical coherence tomography scan circle location and mean retinal nerve fiber layer measurement variability [J].
Gabriele, Michelle L. ;
Ishikawa, Hiroshi ;
Wollstein, Gadi ;
Bilonick, Richard A. ;
Townsend, Kelly A. ;
Kagemann, Larry ;
Wojtkowski, Maciej ;
Srinivasan, Vivek J. ;
Fujimoto, James G. ;
Duker, Jay S. ;
Schuman, Joel S. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (06) :2315-2321