RF rectifiers for EM power harvesting in a Deep Brain Stimulating device

被引:12
作者
Hosain, Md Kamal [1 ]
Kouzani, Abbas Z. [1 ]
Tye, Susannah [2 ]
Kaynak, Akif [1 ]
Berk, Michael [3 ]
机构
[1] Deakin Univ, Sch Engn, Waurn Ponds, Geelong, Vic 3216, Australia
[2] Mayo Clin, Dept Psychiat & Psychol, Rochester, MN 55905 USA
[3] Deakin Univ, Sch Med, IMPACT Strateg Res Ctr, Waurn Ponds, Geelong, Vic 3216, Australia
关键词
RF; Rectifier; Deep brain stimulation; Energy harvesting; Passive device; DESIGN; RECTENNA; CIRCUIT; SYSTEM; FIELD;
D O I
10.1007/s13246-015-0328-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A passive deep brain stimulation (DBS) device can be equipped with a rectenna, consisting of an antenna and a rectifier, to harvest energy from electromagnetic fields for its operation. This paper presents optimization of radio frequency rectifier circuits for wireless energy harvesting in a passive head-mountable DBS device. The aim is to achieve a compact size, high conversion efficiency, and high output voltage rectifier. Four different rectifiers based on the Delon doubler, Greinacher voltage tripler, Delon voltage quadrupler, and 2-stage charge pumped architectures are designed, simulated, fabricated, and evaluated. The design and simulation are conducted using Agilent Genesys at operating frequency of 915 MHz. A dielectric substrate of FR-4 with thickness of 1.6 mm, and surface mount devices (SMD) components are used to fabricate the designed rectifiers. The performance of the fabricated rectifiers is evaluated using a 915 MHz radio frequency (RF) energy source. The maximum measured conversion efficiency of the Delon doubler, Greinacher tripler, Delon quadrupler, and 2-stage charge pumped rectifiers are 78, 75, 73, and 76 % at -5 dBm input power and for load resistances of 5-15 k Omega. The conversion efficiency of the rectifiers decreases significantly with the increase in the input power level. The Delon doubler rectifier provides the highest efficiency at both -5 and 5 dBm input power levels, whereas the Delon quadrupler rectifier gives the lowest efficiency for the same inputs. By considering both efficiency and DC output voltage, the charge pump rectifier outperforms the other three rectifiers. Accordingly, the optimised 2-stage charge pumped rectifier is used together with an antenna to harvest energy in our DBS device.
引用
收藏
页码:157 / 172
页数:16
相关论文
共 39 条
[1]   Analytical models for low-power rectenna design [J].
Akkermans, JAG ;
van Beurden, MC ;
Doodeman, GJN ;
Visser, HJ .
IEEE ANTENNAS AND WIRELESS PROPAGATION LETTERS, 2005, 4 :187-190
[2]   Deep brain stimulation for treatment-resistant depression: Efficacy, safety and mechanisms of action [J].
Anderson, Rodney J. ;
Frye, Mark A. ;
Abulseoud, Osama A. ;
Lee, Kendall H. ;
McGillivray, Jane A. ;
Berk, Michael ;
Tye, Susannah J. .
NEUROSCIENCE AND BIOBEHAVIORAL REVIEWS, 2012, 36 (08) :1920-1933
[3]  
[Anonymous], ADV COMPUTER COMMUNI
[4]  
[Anonymous], NEUROSCI BOMED ENG
[5]  
[Anonymous], P SPIE
[6]  
Bowick C., 2011, RF CIRCUIT DESIGN
[7]   Analysis of the Rectifier Circuit Valid Down to Its Low-Voltage Limit [J].
Cardoso, Adilson J. ;
de Carli, Lucas G. ;
Galup-Montoro, Carlos ;
Schneider, Marcio C. .
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2012, 59 (01) :106-112
[8]  
Dobkin D, 2012, The RF in RFID: Passive UHF RFID in Practice
[9]  
Hosain M. K., 2012, 2012 International Symposium on Communications and Information Technologies (ISCIT), P643, DOI 10.1109/ISCIT.2012.6380979
[10]  
Hosain Md K., 2012, Proceedings of the 2012 ICME International Conference on Complex Medical Engineering (CME), P485, DOI 10.1109/ICCME.2012.6275746