Selecting an accessory device with a metered-dose inhaler: Variable influence of accessory devices on fine particle dose, throat deposition, and drug delivery with asynchronous actuation from a metered-dose inhaler

被引:37
作者
Wilkes, W [1 ]
Fink, J [1 ]
Dhand, R [1 ]
机构
[1] US Dept Vet Affairs, Vet Affairs Edward Hines Jr Hosp, Div Pulm & Crit Care Med, Hines, IL 60141 USA
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 2001年 / 14卷 / 03期
关键词
aerosol; albuterol; bronchodilator; cascade impactor; inhalation therapy;
D O I
10.1089/089426801316970312
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Accessory devices reduce common problems with metered-dose inhalers (MDIs), namely high oropharyngeal deposition of aerosol and incoordination between actuation and inhalation by the patient. The objective of this study was to systematically compare the performance of various accessory devices in vitro. MDIs were tested alone or in combination with four spacers (Toilet paper roll, Ellipse, Optihaler, Myst Assist) and five holding chambers (Aerochamber, Optichamber, Aerosol Cloud Enhancer, Medispacer, and Inspirease). An Anderson cascade impactor was used to measure aerosol mass median aerodynamic diameter (MMAD) and fine particle dose (MMAD < 4.7 mum). In separate experiments, the influence of asynchronous MDI actuation on drug delivery was determined with a simulated spontaneous breathing model. Compared with the MDI alone, all of the accessory devices reduced aerosol MMAD and increased lung-throat ratio (fine particle dose/throat impaction; p < 0.05 for both parameters). The fine particle dose of albuterol was 40% higher with the Ellipse (p < 0.01), was equivalent with the Toilet Paper Roll, Aerochamber, Optichamber, and Medispacer, and was 33-56% lower with the Optihaler, Myst Assist, Aerosol Cloud Enhancer, and Inspirease (p < 0.03). MDI actuation in synchrony with inspiration produced highest drug delivery; when MDI actuation occurred I-sec before inspiration or during exhalation, decrease in drug delivery with holding chambers (10-40% reduction) was less than that with spacers (40-90% reduction). Accessory device selection is complicated by variability in performance between devices, and in the performance of each device in different clinical settings. In vitro characterization of a MDI and accessory device could guide appropriate device selection in various clinical settings.
引用
收藏
页码:351 / 360
页数:10
相关论文
共 29 条
  • [1] CHOOSING THE METERED-DOSE INHALER SPACER OR HOLDING CHAMBER THAT MATCHES THE PATIENTS NEED - EVIDENCE THAT THE SPECIFIC DRUG BEING DELIVERED IS AN IMPORTANT CONSIDERATION
    AHRENS, R
    LUX, C
    BAHL, T
    HAN, SH
    [J]. JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1995, 96 (02) : 288 - 294
  • [2] Metered-dose inhaler accessory devices in acute asthma - Efficacy and comparison with nebulizers: A literature review
    Amirav, I
    Newhouse, MT
    [J]. ARCHIVES OF PEDIATRICS & ADOLESCENT MEDICINE, 1997, 151 (09): : 876 - 882
  • [3] Inhalational drug delivery from seven different spacer devices
    Barry, PW
    OCallaghan, C
    [J]. THORAX, 1996, 51 (08) : 835 - 840
  • [4] THE EFFECT OF DELAY, MULTIPLE ACTUATIONS AND SPACER STATIC CHARGE ON THE IN-VITRO DELIVERY OF BUDESONIDE FROM THE NEBUHALER
    BARRY, PW
    OCALLAGHAN, C
    [J]. BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1995, 40 (01) : 76 - 78
  • [5] Comparative study of dose delivery and cascade impaction performance of four metered-dose inhaler and spacer combinations
    Childers, AG
    Cummings, RH
    Kaufman, BD
    Glennon, JJ
    Kelly, EL
    Lim, JC
    McNally, C
    [J]. CURRENT THERAPEUTIC RESEARCH-CLINICAL AND EXPERIMENTAL, 1996, 57 (02): : 75 - 87
  • [6] Clark AR, 1998, J AEROSOL MED, V11, pS1
  • [7] COADY TJ, 1976, PRACTITIONER, V217, P273
  • [8] DESIGN AND CHARACTERISTICS OF A PORTABLE BREATH ACTUATED, PARTICLE-SIZE SELECTIVE MEDICAL AEROSOL INHALER
    CORR, D
    DOLOVICH, M
    MCCORMACK, D
    RUFFIN, R
    OBMINSKI, G
    NEWHOUSE, M
    [J]. JOURNAL OF AEROSOL SCIENCE, 1982, 13 (01) : 1 - 7
  • [9] CROMPTON GK, 1982, EUR J RESPIR DIS, V63, P101
  • [10] DAVIES DS, 1975, POSTGRAD MED J, V51, P69