Spectral discrimination of Cerenkov radiation in scintillating dosimeters

被引:168
作者
Frelin, AM
Fontbonne, JM
Ban, G
Colin, J
Labalme, M
Batalla, A
Isambert, A
Vela, A
Leroux, T
机构
[1] Phys Corpusculaire Lab, F-14050 Caen, France
[2] Ctr Francois Baclesse, Unite Radiophys, F-14076 Caen, France
[3] ELDIM, F-14200 Herouville St Clair, France
关键词
plastic scintillator dosimetry; CCD camera; calibration; real time;
D O I
10.1118/1.2008487
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiation therapy accelerators require highly accurate dose deposition and the output must be monitored frequently and regularly. Ionization chambers are the primary tool for this control, but their size, their high voltage needed, and the correction needed for electrons make them unsuitable for use during patient treatment. We have developed a small (1-mm-diam and I-mm-long active part), flexible, and water-equivalent dosimeter. It is suitable for photon and electron beams without corrections, and performs on line dose measurements. This detector is based on only one scintillating fiber and a CCD camera. A new signal processing is used to remove the effect of Cerenkov radiation background, which only requires a preliminary calibration. Central-axis depth-dose distribution comparisons have been achieved with standard ionization chambers, over a range from 8 to 25 MV photons and from 6 to 21 MeV electrons in order to validate this calibration. Results show a very good agreement, with less than 1% difference between the two detectors. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:3000 / 3006
页数:7
相关论文
共 14 条
[1]   Radiation-induced light in optical fibers and plastic scintillators: Application to brachytherapy dosimetry [J].
Arnfield, MR ;
Gaballa, HE ;
Zwicker, RD ;
Islam, Q ;
SchmidtUllrich, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :2077-2084
[2]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .2. PROPERTIES AND MEASUREMENTS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1901-1913
[3]   WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) :1883-1900
[4]   A miniature "scintillator-fiberoptic-PMT" detector system for the dosimetry of small fields in stereotactic radiosurgery [J].
Beddar, AS ;
Kinsella, TJ ;
Ikhlef, A ;
Sibata, CH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) :924-928
[5]   CERENKOV LIGHT GENERATED IN OPTICAL FIBERS AND OTHER LIGHT PIPES IRRADIATED BY ELECTRON-BEAMS [J].
BEDDAR, AS ;
MACKIE, TR ;
ATTIX, FH .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (04) :925-935
[6]   A temporal method of avoiding the Cerenkov radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron and photon beams [J].
Clift, MA ;
Johnston, PN ;
Webb, DV .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) :1421-1433
[7]   Dealing with Cerenkov radiation generated in organic scintillator dosimeters by bremsstrahlung beams [J].
Clift, MA ;
Sutton, RA ;
Webb, DV .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1165-1182
[8]   OPTICAL FILTERING AND SPECTRAL MEASUREMENTS OF RADIATION-INDUCED LIGHT IN PLASTIC SCINTILLATION DOSIMETRY [J].
DEBOER, SF ;
BEDDAR, AS ;
RAWLINSON, JA .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (07) :945-958
[9]  
Fluhs D, 1996, MED PHYS, V23, P427, DOI 10.1118/1.597736
[10]  
Fontbonne JM, 2002, IEEE T NUCL SCI, V49, P2223, DOI 10.1109/TNS.2002.803680