On absorbed dose in narrow 60Co gamma-ray beams and dosimetry of the Gamma Knife

被引:18
|
作者
Nizin, PS
机构
[1] Baylor Coll Med, Dept Radiol, Houston, TX 77030 USA
[2] Methodist Hosp, Dept Radiat Therapy, Houston, TX 77030 USA
关键词
Co-60 gamma rays; narrow beams; electronic disequilibrium; absorbed dose; Gamma Knife;
D O I
10.1118/1.598444
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Using separate analytical functions describing primary dose, P-0(d(m),r), collimator scatter, S-c(r), and phantom scatter, TAR(d,r), an expression for absorbed dose in narrow Co-60 gamma-ray beams is developed and each function is quantified: D(d, r)= P-0(d(m),r) S-c(r) TAR(d, r). The absorbed dose is calculated in beams as narrow as 0.2 cm in radius. Analytical and experimental results are compared using measured dose data for the Gamma Knife. Close agreement with experimental data is observed. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)02612-1].
引用
收藏
页码:2347 / 2351
页数:5
相关论文
共 50 条
  • [1] Tissue-air ratios for narrow 60Co gamma-ray beams
    Nizin, PS
    Mooij, RB
    MEDICAL PHYSICS, 1998, 25 (08) : 1458 - 1463
  • [2] EXPERIMENTAL STUDIES ON GAMMA-RAY DOSE RATES FROM A 60CO CYLINDRICAL SOURCE
    FURUTA, Y
    KANEMORI, Y
    NUCLEAR SCIENCE AND ENGINEERING, 1967, 30 (02) : 261 - &
  • [3] COMPUTED VALUES OF SPECIFIC GAMMA-RAY CONSTANT GAMMA FOR 137 CS AND 60CO
    ATTIX, FH
    PHYSICS IN MEDICINE AND BIOLOGY, 1968, 13 (01): : 119 - &
  • [4] ABSORBED GAMMA-RAY DOSE IN OXIDE DIELECTRICS
    DZHAFAROV, YD
    GARIBOV, AA
    ALIEV, KA
    ISKENDEROV, SM
    KRASNOSHTANOV, VF
    SOVIET ATOMIC ENERGY, 1987, 63 (04): : 788 - 790
  • [5] ELECTRON CONTAMINATION IN CO-60 GAMMA-RAY BEAMS
    ATTIX, FH
    LOPEZ, F
    OWOLABI, S
    PALIWAL, BR
    MEDICAL PHYSICS, 1983, 10 (03) : 301 - 306
  • [6] Total Ionizing Dose Effects of SiGe HBTs Induced by 60Co Gamma-Ray Irradiation
    Liu, Shu-Huan
    Hussain, Aqil
    Li, Da
    Guo, Xiaoqiang
    Li, Zhuo-Qi
    Lawal, Olarewaju Mubashiru
    Yang, Jiangkun
    Chen, Wei
    NUCLEAR SCIENCE AND ENGINEERING, 2018, 191 (01) : 98 - 103
  • [7] Precision dosimetry in narrow collimated radiation beams of the Leksell Gamma Knife
    Duftschmid, KE
    Kind, P
    Obenaus, B
    Strachotinsky, C
    Winker, N
    RADIATION PROTECTION DOSIMETRY, 1996, 66 (1-4) : 295 - 298
  • [8] 60CO GAMMA-RAY RADIOLYSIS OF TRIACETONEAMINOXYL IN AQUEOUS-SOLUTIONS
    CHANDRINOS, J
    RAKINTZI.NT
    ZEITSCHRIFT FUR PHYSIKALISCHE CHEMIE-FRANKFURT, 1974, 90 (3-4): : 221 - 224
  • [9] USE OF GLUCOSE SOLUTIONS IN CO-60 GAMMA-RAY DOSIMETRY
    RUSSELL, RD
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1970, 21 (03): : 143 - &
  • [10] Energy deposition by electron, bremsstrahlung, and 60Co gamma-ray beams in multi-layer media
    Seltzer, S.M.
    Berger, M.J.
    International journal of radiation applications and instrumentation. Part A, Applied radiation and isotopes, 1987, 38 (05): : 349 - 364