Next-generation Raman tomography instrument for non-invasive in vivo bone imaging

被引:58
作者
Demers, Jennifer-Lynn H. [1 ]
Esmonde-White, Francis W. L. [2 ]
Esmonde-White, Karen A. [3 ]
Morris, Michael D. [2 ]
Pogue, Brian W. [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[2] Univ Michigan, Dept Chem, Ann Arbor, MI 48109 USA
[3] Univ Michigan, Sch Med, Dept Internal Med, Div Rheumatol, Ann Arbor, MI 48109 USA
关键词
QUALITY-CONTROL; SPECTROSCOPY; TRANSMISSION; DIAGNOSIS; PROBES; TOOL;
D O I
10.1364/BOE.6.000793
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Combining diffuse optical tomography methods with Raman spectroscopy of tissue provides the ability for in vivo measurements of chemical and molecular characteristics, which have the potential for being useful in diagnostic imaging. In this study a system for Raman tomography was developed and tested. A third generation microCT coupled system was developed to combine 10 detection fibers and 5 excitation fibers with laser line filtering and a Cytop reference signal. Phantom measurements of hydroxyapatite concentrations from 50 to 300 mg/ml had a linear response. Fiber placement and experiment design was optimized using cadaver animals with live animal measurements acquired to validate the systems capabilities. Promising results from the initial animal experiments presented here, pave the way for a study of longitudinal measurements during fracture healing and the scaling of the Raman tomography system towards human measurements. (C) 2015 Optical Society of America
引用
收藏
页码:793 / 806
页数:14
相关论文
共 34 条
[1]   Decomposition of in vivo spatially offset Raman spectroscopy data using multivariate analysis techniques [J].
Buckley, Kevin ;
Kerns, Jemma G. ;
Parker, Anthony W. ;
Goodship, Allen E. ;
Matousek, Pavel .
JOURNAL OF RAMAN SPECTROSCOPY, 2014, 45 (02) :188-192
[2]   Near infrared optical tomography using NIRFAST: Algorithm for numerical model and image reconstruction [J].
Dehghani, Hamid ;
Eames, Matthew E. ;
Yalavarthy, Phaneendra K. ;
Davis, Scott C. ;
Srinivasan, Subhadra ;
Carpenter, Colin M. ;
Pogue, Brian W. ;
Paulsen, Keith D. .
COMMUNICATIONS IN NUMERICAL METHODS IN ENGINEERING, 2009, 25 (06) :711-732
[3]   Acquisition and Reconstruction of Raman and Fluorescence Signals for Rat Leg Imaging [J].
Demers, Jennifer-Lynn ;
Pogue, Brian ;
Leblond, Frederic ;
Esmonde-White, Francis ;
Okagbare, Paul ;
Morris, Michael .
MULTIMODAL BIOMEDICAL IMAGING VI, 2011, 7892
[4]  
Esmonde-White F. W. L., 2013, SPIE BIOS
[5]  
Esmonde-White F. W. L., 2010, BIOS
[6]   Biomedical tissue phantoms with controlled geometric and optical properties for Raman spectroscopy and tomography [J].
Esmonde-White, Francis W. L. ;
Esmonde-White, Karen A. ;
Kole, Matthew R. ;
Goldstein, Steven A. ;
Roessler, Blake J. ;
Morris, Michael D. .
ANALYST, 2011, 136 (21) :4437-4446
[7]   Minor Distortions with Major Consequences: Correcting Distortions in Imaging Spectrographs [J].
Esmonde-White, Francis W. L. ;
Esmonde-White, Karen A. ;
Morris, Michael D. .
APPLIED SPECTROSCOPY, 2011, 65 (01) :85-98
[8]   Alterations to Bone Mineral Composition as an Early Indication of Osteomyelitis in the Diabetic Foot [J].
Esmonde-White, Karen A. ;
Esmonde-White, Francis W. L. ;
Holmes, Crystal M. ;
Morris, Michael D. ;
Roessler, Blake J. .
DIABETES CARE, 2013, 36 (11) :3652-3654
[9]   Fiber-optic Raman spectroscopy of joint tissues [J].
Esmonde-White, Karen A. ;
Esmonde-White, Francis W. L. ;
Morris, Michael D. ;
Roessler, Blake J. .
ANALYST, 2011, 136 (08) :1675-1685
[10]   Measurement of Spatial Resolution and Sensitivity in Transmission and Backscattering Raman Spectroscopy of Opaque Samples: Impact on Pharmaceutical Quality Control and Raman Tomography [J].
Everall, Neil ;
Priestnall, Ian ;
Dallin, Paul ;
Andrews, John ;
Lewis, Ian ;
Davis, Kevin ;
Owen, Harry ;
George, Michael W. .
APPLIED SPECTROSCOPY, 2010, 64 (05) :476-484