Validation of the MediByte® type 3 portable monitor compared with polysomnography for screening of obstructive sleep apnea

被引:43
作者
Driver, Helen S. [1 ]
Pereira, Effie J.
Bjerring, Kathryn
Toop, Fern
Stewart, Steven C.
Munt, Peter W.
Fitzpatrick, Michael F.
机构
[1] Queens Univ, Dept Med, Kingston, ON K7L 2V6, Canada
关键词
Diagnostic screening; Home sleep testing; Obstructive sleep apnea; OSA screener; PM studies; Portable monitor; DIAGNOSIS; PRESSURE;
D O I
10.1155/2011/760958
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
BACKGROUND: Portable monitors are increasingly being used as a diagnostic screening tool for obstructive sleep apnea (OSA), and in-laboratory validation of these devices with polysomnography (PSG) is required. OBJECTIVE: To assess the reliability of the MediByte (Braebon Medical Corporation, Canada) type 3 screening device compared with overnight PSG. METHODS: To cover a range of OSA severity, a consecutive series of patients wore the screening device while simultaneously undergoing PSG. Data acquired from the screener and PSG were blinded and scored separately. The number of apneas and hypopneas per hour were calculated using recording time (respiratory disturbance index [RDI]) for the MediByte device, and sleep time (apnea-hypopnea index [AHI]) for PSG. RESULTS: Data from 73 patients with a mean age of 53 years and body mass index of 32.2 kg/m(2) showed high measurement association between the RDI and AHI, with a Pearson correlation of 0.92, accounting for 85% of the variance. Based on Bland-Altman measurement agreement, the mean difference between the RDI and AHI (-5.9 +/- 11.2 events/h) indicated screener under-reporting. For an AHI of greater than 15 events/h, the sensitivity and specificity of the screener was 80% and 97%, respectively; for an AHI of greater than 30 events/h, the positive predictive value was 100%, while the negative predictive value was 88%. CONCLUSION: The MediByte device accurately identified patients without OSA and had a high sensitivity for moderate-to-severe OSA.
引用
收藏
页码:137 / 143
页数:7
相关论文
共 32 条
  • [11] Erman MK, 2007, J CLIN SLEEP MED, V3, P387
  • [12] PORTABLE RECORDING IN THE ASSESSMENT OF OBSTRUCTIVE SLEEP-APNEA
    FERBER, R
    MILLMAN, R
    COPPOLA, M
    FLEETHAM, J
    MURRAY, CF
    IBER, C
    MCCALL, V
    NINOMURCIA, G
    PRESSMAN, M
    SANDERS, M
    STROHL, K
    VOTTERI, B
    WILLIAMS, A
    [J]. SLEEP, 1994, 17 (04) : 378 - 392
  • [13] Home diagnosis of sleep apnea: A systematic review of the literature - An evidence review cosponsored by the American academy of sleep medicine, the American College of Chest Physicians, and the American Thoracic Society
    Flemons, WW
    Littner, MR
    Rowley, JA
    Gay, P
    Anderson, WM
    Hudgel, DW
    McEvoy, RD
    Loube, DI
    [J]. CHEST, 2003, 124 (04) : 1543 - 1579
  • [14] Measuring agreement between diagnostic devices
    Flemons, WW
    Littner, MR
    [J]. CHEST, 2003, 124 (04) : 1535 - 1542
  • [15] Sleep-related breathing disorders in adults: Recommendations for syndrome definition and measurement techniques in clinical research
    Flemons, WW
    Buysse, D
    Redline, S
    Pack, A
    Strohl, K
    Wheatley, J
    Young, T
    Douglas, N
    Levy, P
    McNicholas, W
    Fleetham, J
    White, D
    Schmidt-Nowarra, W
    Carley, D
    Romaniuk, J
    [J]. SLEEP, 1999, 22 (05) : 667 - 689
  • [16] Iber C., 2007, AASM MANUAL SCORING
  • [17] Positional treatment vs continuous positive airway pressure in patients with positional obstructive sleep apnea syndrome
    Jokic, R
    Klimaszewski, A
    Crossley, M
    Sridhar, G
    Fitzpatrick, MF
    [J]. CHEST, 1999, 115 (03) : 771 - 781
  • [18] Lee Won, 2008, Expert Rev Respir Med, V2, P349, DOI 10.1586/17476348.2.3.349
  • [19] Evaluation of the accuracy of SNAP technology sleep Sonography in detecting obstructive sleep apnea in adults compared to standard polysomnography
    Liesching, TN
    Carlisle, C
    Marte, A
    Bonitati, A
    Millman, RP
    [J]. CHEST, 2004, 125 (03) : 886 - 891
  • [20] VALIDATION OF A PORTABLE SLEEP-APNEA MONITORING DEVICE
    MAN, GCW