Differences in effective dose estimation from dose-area product and entrance surface dose measurements in intravenous urography
被引:47
|作者:
Yakoumakis, E
Tsalafoutas, IA
Nikolaou, D
Nazos, I
Koulentianos, E
Proukakis, C
机构:
[1] Univ Athens, Sch Med, Dept Med Phys, GR-11527 Athens, Greece
[2] Konstantopoulio Agia Olga Hosp, Dept Radiol, Athens 14233, Greece
[3] Konstantopoulio Agia Olga Hosp, Med Phys Unit, Athens 14233, Greece
关键词:
D O I:
10.1259/bjr.74.884.740727
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
In this study, measurements of dose-area product (DAP) and entrance surface dose (ESD) were carried out in a sample of 25 adult patients who underwent intravenous urography (IVU). These measured quantities were used to estimate the effective dose E from the IVU examination, a quantity closely correlated to radiation risk. Estimating E involves the use of conversion coefficients that have been determined for specific X-ray views in a mathematical phantom. These are obtained under conditions which are not usually met in clinical practice. As a result, the E estimates using the two different measurable quantities can be quite different. Analysis of the calculation procedure suggests that the E estimate using the DAP measurements, in addition to being more practical, could be more accurate than using ESD measurements, as DAP is sensitive to the X-ray field size settings. Furthermore, it is shown that in the absence of the appropriate equipment, a reliable E estimate can be obtained from the ESD calculated using the exposure data for each X-ray view.