Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications

被引:23
作者
Heine, John J. [1 ]
Thomas, Jerry A. [2 ]
机构
[1] Univ S Florida, H Lee Moffitt Canc Ctr, Canc Prevent & Control Dept, Tampa, FL 33682 USA
[2] Dept Radiol, Wichita, KS USA
关键词
D O I
10.1186/1475-925X-7-13
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Breast density is a significant breast cancer risk factor. Currently, there is no standard method for measuring this important factor. Work presented here represents an essential component of an ongoing project that seeks to determine the appropriate method for calibrating (standardizing) mammography image data to account for the x-ray image acquisition influences. Longer term goals of this project are to make accurate breast density measurements in support of risk studies. Methods: Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM) systems with breast tissue equivalent phantom imaging and compared. Normalization methods were studied to assess the possibility of reducing the amount of calibration data collection. The percent glandular calibration map functional form was investigated. Spatial variations in the calibration data were used to assess the uncertainty in the calibration application by applying error propagation analyses. Results: Logarithmic response curves are well approximated as linear. Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically. Calibration data collection can be reduced by applying a simple normalization technique. The calibration map is well approximated as linear. Intrasystem calibration variation was on the order of four percent, which was approximately half of the intersystem variation. Conclusion: FFDM systems provide a quantitative output, and the calibration quantities presented here may be used for data acquired on similar FFDM systems.
引用
收藏
页数:12
相关论文
共 16 条
[1]   Mammographic breast density as an intermediate phenotype for breast cancer [J].
Boyd, NF ;
Rommens, JM ;
Vogt, K ;
Lee, V ;
Hopper, JL ;
Yaffe, MJ ;
Paterson, AD .
LANCET ONCOLOGY, 2005, 6 (10) :798-808
[2]   Evaluation of detector dynamic range in the x-ray exposure domain in mammography: A comparison between film-screen and flat panel detector systems [J].
Cooper, VN ;
Oshiro, T ;
Cagnon, CH ;
Bassett, LW ;
McLeod-Stockmann, TM ;
Bezrukiy, NV .
MEDICAL PHYSICS, 2003, 30 (10) :2614-2621
[3]   Dissecting a hidden breast cancer risk [J].
Couzin, J .
SCIENCE, 2005, 309 (5741) :1664-1666
[4]  
Dalton William S, 2005, Cancer Control, V12, P140
[5]   Gail model for prediction of absolute risk of invasive breast cancer: Independent evaluation in the Florence-European Prospective Investigation Into Cancer and Nutrition Cohort [J].
Decarli, Adriano ;
Calza, Stefano ;
Masala, Giovanna ;
Specchia, Claudia ;
Palli, Domenico ;
Gail, Mitchell H. .
JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2006, 98 (23) :1686-1693
[6]  
Duffy Stephen W, 2005, Surg Oncol Clin N Am, V14, P671, DOI 10.1016/j.soc.2005.06.001
[7]   Cancer risk prediction models: A workshop on development, evaluation, and application [J].
Freedman, AN ;
Seminara, D ;
Gail, MH ;
Hartge, P ;
Colditz, GA ;
Ballard-Barbash, R ;
Pfeiffer, RM .
JNCI-JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2005, 97 (10) :715-723
[8]   Effective x-ray attenuation measurements with full field digital mammography [J].
Heine, John J. ;
Behera, Madhusmita .
MEDICAL PHYSICS, 2006, 33 (11) :4350-4366
[9]  
Highnam R., 1999, Mammographic Image Analysis
[10]   A calibration approach to glandular tissue composition estimation in digital mammography [J].
Kaufhold, J ;
Thomas, JA ;
Eberhard, JW ;
Galbo, CE ;
Trotter, DEG .
MEDICAL PHYSICS, 2002, 29 (08) :1867-1880