Neutron diagnostics using Compton suppression gamma-ray spectrometer for tritium breeding blanket experiments

被引:2
作者
Kim, C. S. [1 ]
Hong, S. P. [2 ]
Cheon, M. S. [1 ]
Kang, B. S. [2 ]
Cho, S. [1 ]
机构
[1] Natl Fus Res Inst, Daejeon, South Korea
[2] Konyang Univ, Lab Radiat Convergence Sci, Dept Radiol Sci, Coll Med Sci, Daejeon, South Korea
关键词
Neutron diagnostics; Compton suppression gamma spectrometry; KSTAR; ITER TBM; ACTIVATION ANALYSIS; SYSTEM;
D O I
10.1016/j.fusengdes.2016.03.048
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
To obtain tritium breeding ratio in tritium breeding blanket experiments, incident neutron flux measurement is essential. To develop high tritium breeding performance, the more precise neutron flux measurement is required. In this work, the Compton suppression gamma ray spectrometer, which can suppress the expected background noise signals from Compton scatterings, is used to detect the gamma rays from neutron activated indium foils for the purpose of reducing the measurement errors in the neutron activation system (NAS). The statistical error of the measured sample activity in the Compton suppressed system is estimated to be about 2.8% from the counting error (1.4%) and the detector efficiency error (2.4%), and the statistical error of the measured sample activity in the non-suppressed system was estimated to be about 4.7% (from the counting error 2.0% and the detector efficiency error 4.3%) in KSTAR experiment. This system can be applied to ITER TBM and future nuclear fusion devices. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:88 / 92
页数:5
相关论文
共 9 条
[1]   Monte-Carlo optimisation of a Compton suppression system for use with a broad-energy HPGe detector [J].
Britton, R. ;
Burnett, J. L. ;
Davies, A. V. ;
Regan, P. H. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 762 :42-53
[2]   Improving the effectiveness of a low-energy Compton suppression system [J].
Britton, R. ;
Burnett, J. L. ;
Davies, A. V. ;
Regan, P. H. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 729 :64-68
[4]  
Chao J. H., RAD MEAS, V45
[5]   Diagnostic neutron activation system for KSTAR [J].
Cheon, M. S. ;
Lee, Y. S. ;
England, A. C. ;
Kim, H. S. ;
Pak, S. ;
Seon, C. R. ;
Lee, H. G. .
JOURNAL OF INSTRUMENTATION, 2012, 7
[6]   Operational Radioactivity Evaluation of ITER Diagnostic Neutron Activation System [J].
Cheon, M. S. ;
Seon, C. R. ;
Pak, S. ;
Lee, H. G. ;
Bertalot, L. ;
Forrest, R. A. .
IEEE TRANSACTIONS ON PLASMA SCIENCE, 2010, 38 (03) :274-277
[7]   CHARACTERISTICS OF A NEW PNEUMATIC TRANSFER SYSTEM FOR A NEUTRON ACTIVATION ANALYSIS AT THE HANARO RESEARCH REACTOR [J].
Chung, Yong-Sam ;
Kim, Sun-Ha ;
Moon, Jong-Hwa ;
Baek, Sung-Yeol ;
Kim, Hark-Rho ;
Kim, Young-Jin .
NUCLEAR ENGINEERING AND TECHNOLOGY, 2009, 41 (06) :813-820
[8]   NEUTRON-ACTIVATION ANALYSIS OF INDIUM [J].
KARANDASHEV, VK ;
KUZNETSOV, RA ;
GRAZHULENE, SS .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1992, 158 (01) :149-161
[9]   Measurement and simulation of stilbene scintillator response for the KSTAR neutron diagnostic system [J].
Lee, Seung Kyu ;
Son, Jae Bum ;
Kang, JeongSoo ;
Seo, Hee ;
Won, Byung-Hee ;
Park, Se-Hwan ;
Kim, Ho-Dong ;
Kang, Byoung Hwi ;
Kim, Gi Dong ;
Kim, Yong Kyun .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (05) :645-650