Feasibility study on an integrated AEC-grid device for the optimization of image quality and exposure dose in mammography

被引:0
作者
Kim, Kyo-Tae [1 ]
Yun, Ryang-Young [1 ]
Han, Moo-Jae [1 ]
Heo, Ye-Ji [1 ]
Song, Yong-Keun [2 ]
Heo, Sung-Wook [2 ]
Oh, Kyeong-Min [3 ]
Park, Sung-Kwang [4 ]
机构
[1] Inje Univ, Dept Biomed Engn, Gimhae 50834, South Korea
[2] Korea Inst Radiol & Med Sci, Div Heavy Ion Clin Res, Med Phys Res Team, Seoul 01812, South Korea
[3] Korea Atom Energy Res Inst, Radiat Equipment Res Div, Jeongeup 56212, South Korea
[4] Inje Univ, Busan Paik Hosp, Dept Radiat Oncol, Busan 47392, South Korea
关键词
Mammography; FLUKA; Automatic exposure control; Anti-scattering grid;
D O I
10.3938/jkps.71.510
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Currently, in the radiation diagnosis field, mammography is used for the early detection of breast cancer. In addition, studies are being conducted on a grid to produce high-quality images. Although the grid ratio of the grid, which affects the scattering removal rate, must be increased to improve image quality, it increases the total exposure dose. While the use of automatic exposure control is recommended to minimize this problem, existing mammography equipment, unlike general radiography equipment, is mounted on the back of a detector. Therefore, the device is greatly affected by the detector and supporting device, and it is difficult to control the exposure dose. Accordingly, in this research, an integrated AEC-grid device that simultaneously performs AEC and grid functions was used to minimize the unnecessary exposure dose while removing scattering, thereby realizing superior image quality.
引用
收藏
页码:510 / 515
页数:6
相关论文
共 10 条
  • [1] Bushberg J. T., 2003, EUR J NUCL MED MOL I, V30, P12
  • [2] Comparison of air kerma measurements between the PTB and the IAEA for x-radiation qualities used in general diagnostic radiology and mammography
    Csete, Istvan
    Bueermann, Ludwig
    Gomola, Igor
    Girzikowsky, Reinhard
    [J]. METROLOGIA, 2013, 50
  • [3] Fahrig R., 1994, MED PHYS, V21
  • [4] Lee H-M, 2014, KOREAN SOC RADIOL SC, V37, P3
  • [5] Effect of deposition and annealing temperature on mechanical properties of TaN film
    Liu, X.
    Ma, G. J.
    Sun, G.
    Duan, Y. P.
    Liu, S. H.
    [J]. APPLIED SURFACE SCIENCE, 2011, 258 (03) : 1033 - 1037
  • [6] Effect of texture on the high temperature mechanical properties of Nb-1%Zr alloy
    Nieh, T. G.
    Zhang, M.
    Zinkle, S. J.
    McGreevy, T. E.
    Hoelzer, D. T.
    Speakman, S. A.
    [J]. SCRIPTA MATERIALIA, 2006, 55 (08) : 719 - 722
  • [7] Quality assurance and quality control in mammography: A review of available guidance worldwide
    Reis C.
    Pascoal A.
    Sakellaris T.
    Koutalonis M.
    [J]. Insights into Imaging, 2013, 4 (5) : 539 - 553
  • [8] Tailoring automatic exposure control toward constant detectability in digital mammography
    Salvagnini, Elena
    Bosmans, Hilde
    Struelens, Lara
    Marshall, Nicholas W.
    [J]. MEDICAL PHYSICS, 2015, 42 (07) : 3834 - 3847
  • [9] A Comparison Study of the Ridge Filter Parameter by Using FLUKA and GEANT4 Simulation Codes
    Song, Yongkeun
    Shin, Jaeik
    Cho, Sungho
    Yoo, Seunghoon
    Cho, Ilsung
    Kim, Eunho
    Jung, Wongyun
    Choi, Sanghyoun
    Oh, Kyungmin
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (01) : 96 - 102
  • [10] Evaluation of automatic exposure control options in digital mammography
    Zhou, Yifang
    Scott, Alexander, II
    Allahverdian, Janet
    Frankel, Steve
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2014, 22 (03) : 377 - 394