Evaluation of a photon-counting breast tomosynthesis imaging system

被引:28
|
作者
Maidment, A [1 ]
Albert, M [1 ]
Thunberg, S [1 ]
Adelöw, L [1 ]
Blom, O [1 ]
Egerström, J [1 ]
Eklund, M [1 ]
Francke, T [1 ]
Jordung, U [1 ]
Kristoffersson, T [1 ]
Lindman, K [1 ]
Lindqvist, L [1 ]
Marchal, D [1 ]
Olla, H [1 ]
Penton, E [1 ]
Rantanen, J [1 ]
Solokov, S [1 ]
Ullberg, C [1 ]
Weber, N [1 ]
机构
[1] Univ Penn, Philadelphia, PA 19104 USA
关键词
digital breast tomosynthesis; synthetic tomography; digital mammography;
D O I
10.1117/12.597317
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Digital breast tomosynthesis promises solutions to many of the problems currently associated with projection mammography, including elimination of artifactual densities due to the superposition of normal tissues and increasing the conspicuity of true lesions that would otherwise be masked by superimposed normal tissue. We have investigated tomosynthesis using 45 photon counting, orientation sensitive, linear detectors which are precisely aligned with the focal spot of the x-ray source. The x-ray source and the digital detectors are scanned in a continuous motion across the object (patient); each linear detector collecting an image at a distinct angle. Simulations of the imaging system were performed to evaluate the effect of. (1) the range of angles over which projection images are acquired; and (2) the number of projection images acquired used in the tomosynthetic reconstruction. Two different simulations were evaluated; the first was a numerical simulation of a tungsten wire; the second consisted of tomosynthetic reconstructions of a cadaveric rabbit, in which the number and/or range of projection angles was varied. We have shown, analytically and through these simulations, that both the use of more projection angles and the use of a larger range of projection angles improve the image quality of tomosynthetic image reconstructions. The use of a photon-counting x-ray detector system allows us to consider image acquisition geometries with a large number of projection angles, as there is no additive detector noise to degrade the projection or reconstructed images. The maximum number of projection angles and the range of projections angles do have upper practical limits; the range of projection angles is determined predominantly by the detector element size.
引用
收藏
页码:572 / 582
页数:11
相关论文
共 50 条
  • [1] Evaluation of a photon-counting breast tomosynthesis imaging system
    Maidment, Andrew D. A.
    Ullberg, Christer
    Lindman, Karin
    Adelow, Leif
    Egerstrom, Johan
    Eklund, Mathias
    Francke, Tom
    Jordung, Ulf
    Kristoffersson, Tomas
    Lindqvist, Lars
    Marchal, Daniel
    Olla, Hans
    Penton, Erik
    Rantanen, Juha
    Solokov, Skiff
    Weber, Niclas
    Westerberg, Hans
    Medical Imaging 2006: Physics of Medical Imaging, Pts 1-3, 2006, 6142 : B1420 - B1420
  • [2] Evaluation of photon-counting spectral breast tomosynthesis
    Dahlman, Nils
    Fredenberg, Erik
    Aslund, Magnus
    Lundqvist, Mats
    Diekmann, Felix
    Danielsson, Mats
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [3] Clinical evaluation of a photon-counting tomosynthesis mammography system
    Maidment, Andrew D. A.
    Ullberg, Christer
    Francke, Tom
    Lindqvist, Lars
    Sokolov, Skiff
    Lindman, Karin
    Adelow, Leif
    Sunden, Per
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2006, 4046 : 144 - 151
  • [4] Characterization of photon-counting multislit breast tomosynthesis
    Berggren, Karl
    Cederstrom, Bjorn
    Lundqvist, Mats
    Fredenberg, Erik
    MEDICAL PHYSICS, 2018, 45 (02) : 549 - 560
  • [5] Physical characterization of photon-counting tomosynthesis
    Berggren, Karl
    Lundqvist, Mats
    Cederstrom, Bjorn
    Danielsson, Mats
    Fredenberg, Erik
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [6] Evaluation of a photon-counting X-ray imaging system
    Lundqvist, M
    Cederström, B
    Chmill, V
    Danielsson, M
    Hasegawa, B
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (04) : 1530 - 1536
  • [7] Lession characterization in spectral photon-counting tomosynthesis
    Cederstrom, Bjorn
    Fredenberg, Erik
    Berggren, Karl
    Erhard, Klaus
    Danielsson, Mats
    Wallis, Matthew
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [8] High speed imaging photon-counting system
    Ryan, O.
    Redfern, M.
    Shearier, A.
    HIGH ENERGY, OPTICAL, AND INFRARED DETECTORS FOR ASTRONOMY II, 2006, 6276
  • [10] Low dose and high image quality tomosynthesis with a photon-counting imaging technology
    Francke, T
    Maidment, A
    Thunberg, S
    Adelow, L
    Blom, O
    Egerstrom, J
    Eklund, M
    Jordung, U
    Kristoffersson, T
    Lidman, K
    Lindqvist, L
    Marchal, D
    Olla, H
    Penton, E
    Ratanen, J
    Sokolov, S
    Ullberg, C
    Weber, N
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 184 (04) : 42 - 42