High-throughput polymerase chain reaction in parallel circular loops using magnetic actuation

被引:35
作者
Sun, Yi [1 ]
Nguyen, Nam-Trung [2 ]
Kwok, Yien Chian [1 ]
机构
[1] Nanyang Technol Univ, Natl Inst Educ, Singapore 637616, Singapore
[2] Nanyang Technol Univ, Sch Mech & Aerosp Engn, Singapore 639798, Singapore
关键词
D O I
10.1021/ac800787g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We report here a novel multichannel closed-loop magnetically actuated microchip for high-throughput polymerase chain reaction (PCR). This is achieved by designing a series of concentric circular channels on one microchip and exploiting a magnetic force to drive DNA samples flowing continuously through the closed loops. The magnetic force arises from an external permanent magnet through ferrofluid plugs inside the microchannels. The magnet enables simultaneous actuation of DNA samples in all the channels. As the samples go around the loops, they pass through three preset temperature zones. Parameters of PCR, such as incubation time, temperatures, and number of cycles, can be fully controlled and adjusted. High reproducibility was achieved for different channels in the same run and for the same channels in consecutive runs. Genetically modified organisms (GMOs) were amplified simultaneously using the developed device. This simple, reliable, and high-throughput PCR microchip would find wide applications in forensic, clinical, and biological fields.
引用
收藏
页码:6127 / 6130
页数:4
相关论文
共 14 条
[1]   Microfluidic handling of PCR solution and DNA amplification on a reaction chamber array biochip [J].
Gong, HQ ;
Ramalingam, N ;
Chen, LQ ;
Che, J ;
Wang, QH ;
Wang, YM ;
Yang, XH ;
Yap, PHE ;
Neo, CH .
BIOMEDICAL MICRODEVICES, 2006, 8 (02) :167-176
[2]   The official method for the detection of genetically modified soybeans (German Food Act LMBG 35): a semi-quantitative study of sensitivity limits with glyphosate-tolerant soybeans (Roundup Ready) and insect-resistant Bt maize (Maximizer) [J].
Jankiewicz, A ;
Broll, H ;
Zagon, J .
EUROPEAN FOOD RESEARCH AND TECHNOLOGY, 1999, 209 (02) :77-82
[3]   Chemical amplification: Continuous-flow PCR on a chip [J].
Kopp, MU ;
de Mello, AJ ;
Manz, A .
SCIENCE, 1998, 280 (5366) :1046-1048
[4]   Bulk-micromachined submicroliter-volume PCR chip with very rapid thermal response and low power consumption [J].
Lee, DS ;
Park, SH ;
Yang, HS ;
Chung, KH ;
Yoon, TH ;
Kim, SJ ;
Kim, K ;
Kim, YT .
LAB ON A CHIP, 2004, 4 (04) :401-407
[5]   Microchamber array based DNA quantification and specific sequence detection from a single copy via PCR in nanoliter volumes [J].
Matsubara, Y ;
Kerman, K ;
Kobayashi, M ;
Yamamura, S ;
Morita, Y ;
Tamiya, E .
BIOSENSORS & BIOELECTRONICS, 2005, 20 (08) :1482-1490
[6]  
NGUYEN NT, 2006, FUNDAMENTALS APPL MI, P457
[7]   Infrared-mediated thermocycling for ultrafast polymerase chain reaction amplification of DNA [J].
Oda, RP ;
Strausbauch, MA ;
Huhmer, AFR ;
Borson, N ;
Jurrens, SR ;
Craighead, J ;
Wettstein, PJ ;
Eckloff, B ;
Kline, B ;
Landers, JP .
ANALYTICAL CHEMISTRY, 1998, 70 (20) :4361-4368
[8]   Miniaturized flow-through PCR with different template types in a silicon chip thermocycler [J].
Schneegass, I ;
Bräutigam, R ;
Köhler, JM .
LAB ON A CHIP, 2001, 1 (01) :42-49
[9]  
Sun Y, 2007, LAB CHIP, V7, P1012, DOI 10.1039/b700575j
[10]  
Sun Y, 2006, ANAL CHIM ACTA, V556, P80, DOI 10.1016/j.aca.2005.09.035