Comparison of two retinal nerve fibre layer thickness measurement patterns of RTvue optical coherence tomography

被引:1
作者
Li, Shuning [1 ,2 ]
Wang, Xiaozhen [1 ]
Wu, Gewei [3 ]
Fu, Jing [1 ]
Li, Songfeng [1 ]
Wan, Xiuhua [1 ]
Wang, Ningli [1 ]
机构
[1] Capital Med Univ, Beijing Tongren Hosp, Beijing Tongren Eye Ctr, Beijing Ophthalmol & Visual Sci Lab, Nanjing, Peoples R China
[2] Nanjing Tongren Hosp, Dept Ophthalmol, Nanjing, Peoples R China
[3] Shijingshan Hosp, Dept Ophthalmol, Beijing, Peoples R China
关键词
Bland-Altman method; optical coherence tomography; nerve fibre; STRATUS OCT; HISTOLOGIC CORRELATION; GLAUCOMA DETECTION; REPRODUCIBILITY; DAMAGE; DISK;
D O I
10.1111/j.1442-9071.2010.02438.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
P>Background: To evaluate the agreement and repeatability between operator-dependent centring and automatic centring retinal nerve fibre layer thickness measurement patterns (RNFL3.45 and NHM4) of RTvue optical coherence tomography in normal and glaucomatous eyes. Design: Cross-sectional observational study. Participants: A total of 153 eyes from 149 normal subjects and subjects with glaucoma were analysed. Methods: The retinal nerve fibre layer thickness was measured using RNFL3.45 and NHM4 three times on the same day to determine the repeatability and agreement between the two scan patterns. Main Outcome Measures: Student's paired t-testing, intra-class correlation coefficient, coefficient of variation, test-retest viability, Pearson's correlation coefficient and Bland-Altman analysis of retinal nerve fibre layer thickness measurements. Results: The difference between RNFL3.45 and NHM4 measurements was statistically significant by paired t-testing (P = 0.003) only in severe glaucoma group. The Pearson's correlation test showed a high degree of correlation of the mean retinal nerve fibre layer thickness (r = 0.949). Bland-Altman plots showed that the differences between RNFL3.45 and NHM4 were smaller at thicker retinal nerve fibre layer values, but larger at thinner retinal nerve fibre layer values. The intra-class correlation coefficient for RNFL3.45 (and lower 95% confidence interval) in normal and glaucomatous eyes was 0.990 (0.985 confidence interval) and 0.997 (0.995 confidence interval), respectively. The intra-class correlation coefficient for NHM4 in normal and glaucomatous eyes was 0.989 (0.983 confidence interval) and 0.995 (0.992 confidence interval), respectively. The test-retest variability for RNFL3.45 and NHM4 ranged from 3.98 to 9.75 mu m in normal eyes, and from 4.10 to 13.34 mu m in glaucomatous eyes. Conclusion: Measurements of retinal nerve fibre layer thickness by RNFL3.45 and NHM4 are in good agreement.
引用
收藏
页码:222 / 229
页数:8
相关论文
共 19 条
  • [1] Optic disk and nerve fiber layer Imaging to detect glaucoma
    Badala, Federico
    Nouri-Mahdavi, Kouros
    Raoof, Duna A.
    Leeprechanon, Narakorn
    Law, Simon K.
    Caprioli, Joseph
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 144 (05) : 724 - 732
  • [2] Enhanced Glaucoma Staging System (GSS 2) for classifying functional damage in glaucoma
    Brusini, P
    Filacorda, S
    [J]. JOURNAL OF GLAUCOMA, 2006, 15 (01) : 40 - 46
  • [3] Reproducibility of peripapillary retinal nerve fiber thickness measurements with stratus OCT in glaucomatous eyes
    Budenz, Donald L.
    Fredette, Marie-Josee
    Feuer, William J.
    Anderson, Douglas R.
    [J]. OPHTHALMOLOGY, 2008, 115 (04) : 661 - 666
  • [4] Histologic correlation of in vivo optical coherence tomography images of the human retina
    Chen, TC
    Cense, B
    Miller, JW
    Rubin, PAD
    Deschler, DG
    Gragoudas, ES
    de Boer, JF
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (06) : 1165 - 1168
  • [5] Spectral domain optical coherence tomography - Ultra-high speed, ultra-high resolution ophthalmic imaging
    Chen, TC
    Cense, B
    Pierce, MC
    Nassif, N
    Park, BH
    Yun, SH
    White, BR
    Bouma, BE
    Tearney, GJ
    de Boer, JF
    [J]. ARCHIVES OF OPHTHALMOLOGY, 2005, 123 (12) : 1715 - 1720
  • [6] Optical coherence tomography scan circle location and mean retinal nerve fiber layer measurement variability
    Gabriele, Michelle L.
    Ishikawa, Hiroshi
    Wollstein, Gadi
    Bilonick, Richard A.
    Townsend, Kelly A.
    Kagemann, Larry
    Wojtkowski, Maciej
    Srinivasan, Vivek J.
    Fujimoto, James G.
    Duker, Jay S.
    Schuman, Joel S.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (06) : 2315 - 2321
  • [7] Histologic correlation of pig retina radial stratification with ultrahigh-resolution optical coherence tomography
    Gloesmann, M
    Hermann, B
    Schubert, C
    Sattmann, H
    Ahnelt, PK
    Drexler, W
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (04) : 1696 - 1703
  • [8] Glaucoma detection by Stratus OCT
    Hougaard, Jesper Leth
    Heijl, Anders
    Bengtsson, Boel
    [J]. JOURNAL OF GLAUCOMA, 2007, 16 (03) : 302 - 306
  • [9] Glaucoma detection using different Stratus optical coherence tomography protocols
    Hougaard, Jesper Leth
    Heijl, Anders
    Bengtsson, Boel
    [J]. ACTA OPHTHALMOLOGICA SCANDINAVICA, 2007, 85 (03): : 251 - 256
  • [10] HOYT WF, 1973, INVEST OPHTH VISUAL, V12, P814